| Literature DB >> 23049379 |
Angela Melgaço Ferreira1, Stela Brener, Ricardo Vilas Freire de Carvalho, Marcele Cunha Ribeiro do Valle, Helio Moraes-Souza.
Abstract
OBJECTIVE: This study aimed at identifying associations between the participation of transfusion services in immunohematology external quality control programs and their accuracy in immunohematology testing and adaptation to technical and legal operational procedures.Entities:
Keywords: Blood banks; Blood transfusion; Program evaluation; Quality assurance, health care; Quality control
Year: 2012 PMID: 23049379 PMCID: PMC3459603 DOI: 10.5581/1516-8484.20120009
Source DB: PubMed Journal: Rev Bras Hematol Hemoter ISSN: 1516-8484
Performance of the 219 State Transfusion Services that participated in external quality control programs and those that did not in mandatory pre-transfusion testing - Minas Gerais, 2007-2008
| Pre-transfusion tests | Participated in external quality control programs n (%) | Did not participate in external quality control programs n (%) | p-value | ||
| Correct | Incorrect | Correct | Incorrect | ||
| ABO classification | 276 (97.2) | 8 (2.8) | 560 (94.6) | 32 (5.4) | 0.085 |
| RhD classification | 281 (98.9) | 3(1.1) | 558 (94.2) | 34 (5.7) | 0.001 |
| Antibody screening | 120 (84.5) | 22 (15.5) | 201 (68.0) | 95 (32.0) | < 0.001 |
| Compatibility testing | 131 (92.2) | 11 (7.7) | 259 (85.7) | 37(14.2) | 0.136 |
* tests performed on samples from recipients and donors
** tests performed on samples from recipients
*** tests performed between samples from recipients and donors
Figure 1Percentage of false positive and false negative results in irregular red blood cell antibody screening performed by State Transfusion Services that participated in external quality control programs and those that did not
EQCP: Participated in external quality control programs; No EQCP: Did not participate in external quality control programs
Analysis of the parameters of compliance in accordance with current legislation of the 219 State Transfusion Services that had participated in External Quality control programs and those that had not - Minas Gerais, 2007-2008
| Parameter | Participated in external quality control programs | Did not participate in external quality control programs | p-value | ||
| n = 71 | n= 148 | ||||
| n | % | n | % | ||
| Materials | |||||
| Reagents with instructions for use | 52 | 73.2 | 89 | 60.1 | 0.058 |
| Non-manual preparation of' 'O' RBC reagents (for irregular RBC antibody screening) | 55 | 77.5 | 84 | 56.8 | 0.016 |
| Sera stored between 2ºC and 8ºC | 64 | 90.1 | 120 | 81.1 | 0.126 |
| Use of RhD serum and Rh control from same manufacturer | 68 | 95.8 | 126 | 85.1 | 0.020 |
| RBC suspensions in pre-transfusion tests | 63 | 88.7 | 124 | 83.8 | 0.332 |
| Equipment | |||||
| Specific refrigerator to store blood bags | 70 | 98.6 | 146 | 98.6 | 1.000 |
| Specific refrigerator with adequate temperature | 62 | 87.3 | 124 | 83.8 | 0.469 |
| Corrective maintenance of equipment | 58 | 81.7 | 108 | 73.0 | 0.159 |
| Preventive maintenance of equipment | 40 | 56.3 | 35 | 23.6 | 0.000 |
| Tests and Methods | |||||
| Transfusion records | 29 | 40.8 | 18 | 12.2 | 0.000 |
| ABO/RhD phenotyping of each new recipient sample | 63 | 88.7 | 105 | 70.9 | 0.004 |
| Research/Classification of irregular antibodies of recipients of packed red blood cells | 71 | 100.0 | 139 | 93.9 | 0.033 |
| Transportation | |||||
| Protocols of transportation for non-conforming procedures | 21 | 29.6 | 21 | 14.2 | 0.007 |
| Registration of arrival and departure of blood | 67 | 94.4 | 129 | 87.2 | 0.129 |
| SOPs/ Transportation of blood products | 21 | 29.6 | 20 | 13.5 | 0.004 |
| Transport of blood products with thermometer and ice | 42 | 60.9 | 103 | 69.6 | 0.204 |
| Working practice | |||||
| Up-to-date SOPs / pre-transfusion tests | 23 | 32.4 | 24 | 16.2 | 0.006 |
| Quality control | |||||
| Quality control of reagents | 7 | 9.9 | 5 | 3.4 | 0.060 |
| Sensitized RBCs/validation of antiglobulin tests | 16 | 22.5 | 11 | 7.4 | 0.002 |
| Human resources | |||||
| Clinical pathology technician | 40 | 56.3 | 52 | 35.1 | 0.003 |
PRC: Packed red cells; SOP: Standard operating procedure; RBC: Red blood cells
Parameters in the final logistic regression model analyzing State Transfusion Services that had participated in external quality control programs and those that had not
| Parameter | p-value | OR | 95% CI |
| Correct results of proficiency panel testing | 0.008 | 2.35 | (1.25 -4.42) |
| Transfusion records | 0.002 | 3.16 | (1.50 - 6.63) |
| Preventive maintenance of equipment | 0.002 | 2.81 | (1.46-5.41) |
95% CL - 95% confidence interval; OR - Odds ratio
Hosmer & Lemeshow - p-value = 0.914