EPID-based verification of the MLC performance for dynamic IMRT and VMAT.

Pejman Rowshanfarzad1, Mahsheed Sabet, Michael P Barnes, Daryl J O'Connor, Peter B Greer.   

Abstract

PURPOSE: In advanced radiotherapy treatments such as intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT), verification of the performance of the multileaf collimator (MLC) is an essential part of the linac QA program. The purpose of this study is to use the existing measurement methods for geometric QA of the MLCs and extend them to more comprehensive evaluation techniques, and to develop dedicated robust algorithms to quantitatively investigate the MLC performance in a fast, accurate, and efficient manner.
METHODS: The behavior of leaves was investigated in the step-and-shoot mode by the analysis of integrated electronic portal imaging device (EPID) images acquired during picket fence tests at fixed gantry angles and arc delivery. The MLC was also studied in dynamic mode by the analysis of cine EPID images of a sliding gap pattern delivered in a variety of conditions including different leaf speeds, deliveries at fixed gantry angles or in arc mode, and changing the direction of leaf motion. The accuracy of the method was tested by detection of the intentionally inserted errors in the delivery patterns.
RESULTS: The algorithm developed for the picket fence analysis was able to find each individual leaf position, gap width, and leaf bank skewness in addition to the deviations from expected leaf positions with respect to the beam central axis with sub-pixel accuracy. For the three tested linacs over a period of 5 months, the maximum change in the gap width was 0.5 mm, the maximum deviation from the expected leaf positions was 0.1 mm and the MLC skewness was up to 0.2°. The algorithm developed for the sliding gap analysis could determine the velocity and acceleration∕deceleration of each individual leaf as well as the gap width. There was a slight decrease in the accuracy of leaf performance with increasing leaf speeds. The analysis results were presented through several graphs. The accuracy of the method was assessed as 0.01 mm for both the gap size and peak position determination.
CONCLUSIONS: This study provides fast, easy, and accurate test methods for routine QA of the MLC performance and helps in faster troubleshooting of MLC problems in both IMRT and VMAT treatments.

Entities:  

Mesh:

Year:  2012        PMID: 23039657     DOI: 10.1118/1.4752207

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  11 in total

1.  Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi-leaf collimator quality assurance.

Authors:  Simon K Goodall; Craig Norvill
Journal:  J Appl Clin Med Phys       Date:  2022-06-06       Impact factor: 2.243

2.  Evaluation of the TrueBeam machine performance check (MPC) beam constancy checks for flattened and flattening filter-free (FFF) photon beams.

Authors:  Michael P Barnes; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2016-11-30       Impact factor: 2.102

3.  A novel and independent method for time-resolved gantry angle quality assurance for VMAT.

Authors:  Todsaporn Fuangrod; Peter B Greer; Benjamin J Zwan; Michael P Barnes; Joerg Lehmann
Journal:  J Appl Clin Med Phys       Date:  2017-07-13       Impact factor: 2.102

4.  Evaluation of the truebeam machine performance check (MPC): mechanical and collimation checks.

Authors:  Michael P Barnes; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2017-04-17       Impact factor: 2.102

5.  Quality control of VMAT synchronization using portal imaging.

Authors:  James L Bedford; Honorata Chajecka-Szczygielska; Michael D R Thomas
Journal:  J Appl Clin Med Phys       Date:  2015-01-08       Impact factor: 2.102

6.  VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests.

Authors:  Michael P Barnes; Pejman Rowshanfarzad; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

7.  Efficient quality assurance method with automated data acquisition of a single phantom setup to determine radiation and imaging isocenter congruence.

Authors:  Hyejoo Kang; Rakesh Patel; John C Roeske
Journal:  J Appl Clin Med Phys       Date:  2019-09-19       Impact factor: 2.102

8.  Determination of Multileaf Collimator Positional Errors as a Function of Dose Rate, Speed, and Delivery Interruption for Volumetric-Modulated Arc Therapy Delivery.

Authors:  Soumya Roy; Biplab Sarkar; Anirudh Pradhan
Journal:  J Med Phys       Date:  2021-12-31

9.  An EPID-based system for gantry-resolved MLC quality assurance for VMAT.

Authors:  Benjamin J Zwan; Michael P Barnes; Todsaporn Fuangord; Cameron J Stanton; Daryl J O'Connor; Paul J Keall; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

10.  Quality assurance of geometric accuracy based on an electronic portal imaging device and log data analysis for Dynamic WaveArc irradiation.

Authors:  Hideaki Hirashima; Yuki Miyabe; Mitsuhiro Nakamura; Nobutaka Mukumoto; Takashi Mizowaki; Masahiro Hiraoka
Journal:  J Appl Clin Med Phys       Date:  2018-04-06       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.