Literature DB >> 23034689

Late complications of hydrogel scleral buckle implants and a technique for effective removal.

Ayo-Lynn Richards1, Dale R Meyer.   

Abstract

PURPOSE: The aim of this article is to describe complications of hydrogel (MIRAgel) scleral buckle implants that may appear many years after implantation, and to describe an effective technique to facilitate safe removal of these implants, which become extremely friable with time.
METHODS: Four case reports of patients with complications of hydrogel orbital implants and a technique for removal of the implants are described with review of the literature. The technique involves a conjunctival incision into each involved quadrant, followed by use of blunt instrumentation for "push" and a Frasier suction handpiece for "pull" to allow segmental removal. This technique limits the tendency of the implant to fragment, which has been a problem described with previous techniques.
RESULTS: Cases typically present simulating an orbital tumor or infection, with the diagnosis not immediately apparent, as was the case in 3 of 4 cases in our series. The technique we used allowed for expeditious removal and prompt recovery in all 4 patients.
CONCLUSIONS: Hydrogel scleral buckle implants can produce late complications that may be difficult to diagnose and manage. The "push-pull" technique described in this article facilitates effective removal, and may have implications for other types of orbital hydrogel implants as well.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23034689     DOI: 10.1097/IOP.0b013e3182696515

Source DB:  PubMed          Journal:  Ophthalmic Plast Reconstr Surg        ISSN: 0740-9303            Impact factor:   1.746


  3 in total

1.  Globe Loss From Intraocular Invasion of MIRAgel Scleral Buckle Components.

Authors:  Tal J Rubinstein; Maria M Choudhary; Yasha S Modi; Justis P Ehlers; Julian D Perry
Journal:  Ophthalmic Plast Reconstr Surg       Date:  2016 Sep-Oct       Impact factor: 1.746

2.  Management of Delayed Complications of Hydrogel Scleral Buckles.

Authors:  Hsin-Yu Yang; Wei-Kuang Yu; Chieh-Chih Tsai
Journal:  J Pers Med       Date:  2022-04-14

3.  Removal of MIRAgel scleral buckle implants using Yankauer suction catheter with adjusted diameter.

Authors:  Shintaro Nakao; Takashi Nishida; Koh-Hei Sonoda
Journal:  Am J Ophthalmol Case Rep       Date:  2022-03-07
  3 in total