Vagus nerve stimulator stability and interference on radiation oncology x-ray beams.

Michael S Gossman1, Amruta Ketkar, Arthur K Liu, Bryan Olin.   

Abstract

Five different models of Cyberonics, Inc. vagus nerve stimulation (VNS) therapy pulse generators were investigated for their stability under radiation and their ability to change the absorbed dose from incident radiation. X-ray beams of 6 MV and 18 MV were used to quantify these results up to clinical doses of 68-78 Gy delivered in a single fraction. In the first part, the effect on electronic stimulation signaling of each pulse generator was monitored during and immediately afterwards with computer interrogation. In the second part, the effects of having the pulse generators scatter or attenuate the x-ray beam was also characterized from dose calculations on a treatment planning system as well as from actual radiation measurements. Some device models were found to be susceptible to radiation interference when placed directly in the beam of high energy therapeutic x-ray radiation. While some models exhibited no effect at all, others showed an apparent loss of stimulation output immediately after radiation was experienced. Still, other models were observed to have a cumulative dose effect with a reduced output signal, followed by battery depletion above 49 Gy. Absorbed dose changes on computer underestimated attenuation by nearly half for both energies amongst all pulse generators, although the computer did depict the proper shape of the changed distribution of dose around the device. Measured attenuation ranged from 7.0% to 11.0% at 6 MV and 4.2% to 5.2% at 18 MV for x-rays. Processes of back-scatter and side-scatter were deemed negligible although recorded. Identical results from 6 MV and 18 MV x-ray beams conclude no neutron effect was induced for the 18 MV beam. As there were documented effects identified in this research regarding pulse generation, it emphasizes the importance of caution when considering radiation therapy on patients with implanted VNS devices with observed malfunctions consequential.

Entities:  

Mesh:

Year:  2012        PMID: 23032351     DOI: 10.1088/0031-9155/57/20/N365

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  1 in total

Review 1.  A Review and Analysis of Managing Commonly Seen Implanted Devices for Patients Undergoing Radiation Therapy.

Authors:  Maria F Chan; Claire Young; Daphna Gelblum; Chengyu Shi; Carolanne Rincon; Elizabeth Hipp; Jingdong Li; Dongxu Wang
Journal:  Adv Radiat Oncol       Date:  2021-08-06
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.