Literature DB >> 23025452

In vitro tests for aerosol deposition. III: effect of inhaler insertion angle on aerosol deposition.

Renish R Delvadia1, P Worth Longest, Michael Hindle, Peter R Byron.   

Abstract

BACKGROUND: Inhaler orientation with respect to a patient's mouth may be an important variable determining the efficiency of aerosol lung delivery. The effect of insertion angle on regional deposition was evaluated for a series of inhalers using concurrent in vitro and computational fluid dynamics (CFD) analysis.
METHODS: Geometrically realistic physical mouth-throat (MT) and upper tracheobronchial (TB) models were constructed to connect different inhalers at a series of insertion angles relative to the horizontal plane of the model. These models were used to assess albuterol sulfate deposition from the Novolizer(®) dry powder inhaler (DPI), Proventil(®) HFA pressurized metered dose inhaler (MDI), and Respimat(®) Soft Mist™ Inhaler (SMI) following the actuation of a single dose. Drug deposition from Novolizer DPI was studied for Salbulin(®) and an experimental "drug only" formulation. Albuterol sulfate was recovered and quantified from the device and the MT and TB regions.
RESULTS: Significant differences in MT and total lung dose (TLD) of albuterol sulfate deposition were not observed for Salbulin Novolizer DPI and Respimat SMI inserted at different angles. In contrast, drug-only Novolizer DPI and Proventil HFA MDI showed a significant difference in MT and TLD deposition using different insertion angles. For drug-only Novolizer DPI and Proventil HFA MDI, the lowest and the highest MT depositions were observed at +10° and -20°, respectively; for Respimat SMI and Salbulin Novolizer DPI, these angles were -10° and +10°, and +20° and -20°, respectively. CFD simulations were in agreement with the experimental results and illustrated shifts in local particle deposition associated with changes in insertion angle.
CONCLUSION: The effect of inhaler orientation at the inhaler-mouth interface on MT aerosol deposition appeared to be dependent on velocity, aerosol size, and formulation. These findings not only demonstrate the need for patient education on correct inhaler orientation, but provide important new methods for those designing new inhalers.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23025452     DOI: 10.1089/jamp.2012.0989

Source DB:  PubMed          Journal:  J Aerosol Med Pulm Drug Deliv        ISSN: 1941-2711            Impact factor:   2.849


  10 in total

1.  Effect of Inhalation Flow Rate on Mass-Based Plume Geometry of Commercially Available Suspension pMDIs.

Authors:  Daniel F Moraga-Espinoza; Eli Eshaghian; Albert Shaver; Hugh D C Smyth
Journal:  AAPS J       Date:  2018-07-12       Impact factor: 4.009

2.  The use of condensational growth methods for efficient drug delivery to the lungs during noninvasive ventilation high flow therapy.

Authors:  Laleh Golshahi; Geng Tian; Mandana Azimi; Yoen-Ju Son; Ross Walenga; P Worth Longest; Michael Hindle
Journal:  Pharm Res       Date:  2013-06-26       Impact factor: 4.200

3.  Validating Whole-Airway CFD Predictions of DPI Aerosol Deposition at Multiple Flow Rates.

Authors:  P Worth Longest; Geng Tian; Navvab Khajeh-Hosseini-Dalasm; Michael Hindle
Journal:  J Aerosol Med Pulm Drug Deliv       Date:  2016-04-15       Impact factor: 2.849

Review 4.  Use of computational fluid dynamics deposition modeling in respiratory drug delivery.

Authors:  P Worth Longest; Karl Bass; Rabijit Dutta; Vijaya Rani; Morgan L Thomas; Ahmad El-Achwah; Michael Hindle
Journal:  Expert Opin Drug Deliv       Date:  2018-12-10       Impact factor: 6.648

5.  Validating CFD Predictions of Pharmaceutical Aerosol Deposition with In Vivo Data.

Authors:  Geng Tian; Michael Hindle; Sau Lee; P Worth Longest
Journal:  Pharm Res       Date:  2015-05-06       Impact factor: 4.200

6.  Characterizing the Effects of Nasal Prong Interfaces on Aerosol Deposition in a Preterm Infant Nasal Model.

Authors:  Karl Bass; Mohammad A M Momin; Connor Howe; Ghali Aladwani; Sarah Strickler; Arun V Kolanjiyil; Michael Hindle; Robert M DiBlasi; Worth Longest
Journal:  AAPS PharmSciTech       Date:  2022-04-19       Impact factor: 3.246

7.  Recommendations for Simulating Microparticle Deposition at Conditions Similar to the Upper Airways with Two-Equation Turbulence Models.

Authors:  Karl Bass; P Worth Longest
Journal:  J Aerosol Sci       Date:  2018-02-21       Impact factor: 3.433

8.  In Vitro Tests for Aerosol Deposition. IV: Simulating Variations in Human Breath Profiles for Realistic DPI Testing.

Authors:  Renishkumar R Delvadia; Xiangyin Wei; P Worth Longest; Jurgen Venitz; Peter R Byron
Journal:  J Aerosol Med Pulm Drug Deliv       Date:  2015-10-08       Impact factor: 2.849

9.  High-Efficiency Dry Powder Aerosol Delivery to Children: Review and Application of New Technologies.

Authors:  Karl Bass; Dale Farkas; Amr Hassan; Serena Bonasera; Michael Hindle; P Worth Longest
Journal:  J Aerosol Sci       Date:  2020-10-14       Impact factor: 3.433

10.  Instant velocity and consistency of emitted cloud change by the different levels of canister filling with Metered Dose Inhalers (MDIs), but not with Soft Mist Inhalers (SMIs): a bench study.

Authors:  Roberto W Dal Negro; Pietro Longo; Orestepaolo Villanis Ziani; Luca Bonadiman; Paola Turco
Journal:  Multidiscip Respir Med       Date:  2017-05-29
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.