Adequate modeling of light propagation in the head is important to predict the sensitivity of NIRS signal and the spatial sensitivity profile of source-detector pairs. The 3D realistic head models of which the geometry is based upon the anatomical images acquired by magnetic resonance imaging and x-ray computed tomography are constructed to investigate the influence of the frontal sinus on the NIRS signal and spatial sensitivity. Light propagation in the head is strongly affected by the presence of the frontal sinus. The light tends to propagate around the frontal sinus. The influence of the frontal sinus on the sensitivity of the NIRS signal to the brain activation is not consistent and depends on the depth of the frontal sinus, the optical properties of the superficial tissues and the relative position between the source-detector pair and the frontal sinus. The frontal sinus located in the shallow region of the skull tends to reduce the sensitivity of the NIRS signal while the deep frontal sinus can increase the sensitivity of the NIRS signal.
Adequate modeling of light propagation in the head is important to predict the sensitivity of NIRS signal and the spatial sensitivity profile of source-detector pairs. The 3D realistic head models of which the geometry is based upon the anatomical images acquired by magnetic resonance imaging and x-ray computed tomography are constructed to investigate the influence of the frontal sinus on the NIRS signal and spatial sensitivity. Light propagation in the head is strongly affected by the presence of the frontal sinus. The light tends to propagate around the frontal sinus. The influence of the frontal sinus on the sensitivity of the NIRS signal to the brain activation is not consistent and depends on the depth of the frontal sinus, the optical properties of the superficial tissues and the relative position between the source-detector pair and the frontal sinus. The frontal sinus located in the shallow region of the skull tends to reduce the sensitivity of the NIRS signal while the deep frontal sinus can increase the sensitivity of the NIRS signal.
Entities:
Keywords:
(170.2655) Functional monitoring and imaging; (170.3660) Light propagation in tissues; (170.5280) Photon migration
Near-infrared spectroscopy (NIRS) and optical imaging have been used to
non-invasively measure cerebral oxygenation and hemodynamic changes associated with
brain activation [1,2]. They have been widely applied to a variety of brain function
studies [3-7] and clinical measurements [8-13]. In NIRS
measurements, the source and detector probes are attached to the scalp to detect the
change in optical density caused by the hemodynamic change in the brain.
Consequently, the light needs to pass through the superficial tissues such as,
scalp, skull both before and after passing through the brain tissue. It is important
to obtain the sensitivity of the NIRS signal to the absorption change in the brain
and the volume of tissue sampled with a particular source-detector pair, however,
light is strongly scattered by the tissues, thus the path of light propagation in
the head is ambiguous. Since light intensity in the head cannot be non-invasively
measured by experiments, light propagation in the head has been predicted by
simulations.Light propagation in the head is affected by the heterogeneity in optical properties
of the tissues. Especially, cerebrospinal fluid (CSF) that has both low scattering
and absorption coefficients has a strong effect on light propagation in the brain
[14-16]. The influence of a CSF layer cannot be ignored in any
modeling of light propagation in the head. The boundaries between various tissues in
the head have complex geometry. The brain surface is folded with sulci filled with
the CSF. The geometry of the tissue boundaries is also important for adequate
modeling of light propagation in the head. Recently, the realistic head models, the
structures of which are based upon magnetic resonance (MR) head images has been used
to predict light propagation in the head for NIRS and optical imaging [17-20]. These head models basically consisted of five types of tissues that
imitate the scalp, skull, CSF, gray matter and white matter. A more sophisticated
model, for instance, which includes the extra-cerebral vasculature, has been
constructed to investigate the influence of the large pial vessels, such as sagittal
sinus, on the diffuse optical imaging [21].The frontal sinus (FS) in the skull is an optically void region. The analysis of
light propagation in a simplified head model with a void region mimicking the FS
indicated that light propagation in the brain is affected by the FS [22]. However, the FS tends to be ignored in the
realistic head models based on the MR images because there is very little contrast
between the FS and skull regions in the MR image to obtain precise geometry of the
FS.In this study, the 3D realistic head model including a FS is constructed in order to
investigate the influence of a FS on the NIRS measurements. The FS can be clearly
observed in images acquired by x-ray computed tomography (CT). The anatomical
geometry of the head model is based upon MR and x-ray CT images. Three additional
models including the artificially modified FS are also constructed and light
propagation in these head models is compared to discuss the influence of the FS on
the NIRS signal and the spatial sensitivity profile.
2. Method of simulation
2.1. Construction of the realistic head models
The geometry of the realistic head models is based on the anatomical images
obtained by MR imaging and x-ray CT. The original 3D MR images contain 166
planes with 1.0 mm intervals and each plane consists of 256 × 256 pixels
with square sides of 1.0 mm. The original 3D CT images contain 381 planes with
0.5 mm intervals and each plane consists of 512 × 512 pixels with square
sides of 0.41 mm. Figures 1(a)
and 1(b) show the T1-weighted MR
image and CT image in the same plane of a slice, respectively. In the
T1-weighted MR images, the contrast among the grey matter, white matter and CSF
is sufficient enough to identify these boundaries. It is difficult to find the
boundaries between the skull and the adjacent regions because bone and water
(CSF) produce a low signal while the bone marrow and fat in the scalp produce a
high signal. Since the air produces a low signal in the T1-weighted MR images,
the region of the FS cannot be recognized in the images. In the CT images, the
boundaries between the skull and adjacent regions are easily distinguishable and
the FS is clearly observed in the frontal bone.
Fig. 1
(a1) An axial slice of MR image. (a2) A sagittal slice of MR image.
(b1) An axial slice of CT image. (b2) A sagittal slice of CT image.
The MR and CT images are in the same plane. White lines in (a1) and
(a2) indicate the plane of the sagittal and axial slices,
respectively.
(a1) An axial slice of MR image. (a2) A sagittal slice of MR image.
(b1) An axial slice of CT image. (b2) A sagittal slice of CT image.
The MR and CT images are in the same plane. White lines in (a1) and
(a2) indicate the plane of the sagittal and axial slices,
respectively.The head models consist of 6 regions, which include the air, scalp, skull, CSF
(subarachnoid space), gray matter and white matter. Since the spatial resolution
of the MR and CT images are different, the CT images were realigned by SPM8
software [23] to coregistrate images of
the MR and CT. The MR and CT images were binarized by the conventional region
growing algorithm [24] for the 3D
grayscale images. The outer surface of the whole head, gray matter and white
matter were identified from the MR images while the skull and air in the FS were
identified from the CT images to determine the boundaries of each type of
tissues. An axial slice of the head model segmented from the MR and CT images
and the 3D segmented geometry of the head model are shown in Figs. 2(a)
and 2(b), respectively. The outer
surfaces of the whole head, FS and brain are represented in the 3D segmented
geometry. The head models consist of 256 × 256 × 106 cubic voxels
with side 1 mm. Each voxel is specified with its optical properties according to
the results of the tissue segmentation. The optical properties of each type of
tissues in the head models at the 830 nm wavelength were chosen from the
reported data [16,25-27]
(parameter set (I)). Actually, a wide range of different values in the optical
properties has been reported for the same types of tissues. Light propagation in
the head models with two additional parameter sets was predicted in order to
investigate the dependence of the influence of the FS on the optical properties
of the head. Extreme values were chosen based on the literature [21,28,29] as parameter set (II).
The scattering coefficient of all the types of tissues in parameter set (II) is
lower than those in parameter set (I). In parameter set (III), the optical
properties of the scalp and skull are the same as those in parameter set (II)
while the optical properties of the other parts of the head are the same as
those in parameter set (I). The optical properties of each type of tissues in
the head models are listed in Table 1
.
Fig. 2
(a) An axial slice of the segmented image of the realistic head
model. (b) 3D segmented geometry of the realistic head model. The
outer surfaces of the whole head, FS and brain are only
represented.
Table 1
Optical properties of each type of tissues in the head
model
Parameter
set
Scalp
Skull
FS
CSF
Gray matter
White matter
Transport
scattering coefficient (mm−1)
I
1.84
1.48
0.01
0.24
2.12
8.82
II
0.66
0.86
0.01
0.01
0.7
0.7
III
0.66
0.86
0.01
0.24
2.12
8.82
Absorption
coefficient (mm−1)
I
0.025
0.025
0.001
0.004
0.039
0.017
II
0.019
0.014
0.001
0.003
0.020
0.080
III
0.019
0.014
0.001
0.004
0.039
0.017
(a) An axial slice of the segmented image of the realistic head
model. (b) 3D segmented geometry of the realistic head model. The
outer surfaces of the whole head, FS and brain are only
represented.
2.2. Modification of the FS
Three additional models were constructed to investigate the influence of the FS
on the sensitivity of the NIRS signal. In the additional models, the geometry of
the FS was only modified from that in the realistic head model shown in Fig. 2 (the model with the whole FS). Figure 3(a)
shows an axial slice of the frontal area of the model with the whole FS.
In the model without the FS, the FS is replaced by the skull as shown in Fig. 3(b). In the model with the shallow
FS, half of the volume in the deeper region of the FS was replaced by the skull
and that in the shallow region was replaced by the skull in the model with the
deep FS as shown in Figs. 3(c) and 3(d), respectively.
Fig. 3
An axial slice of the head models for evaluation of the influence of
the FS on light propagation in the head. (a) The model with the
whole FS, (b) the model without the FS, (c) the model with the
shallow FS, (d) the model with the deep FS.
An axial slice of the head models for evaluation of the influence of
the FS on light propagation in the head. (a) The model with the
whole FS, (b) the model without the FS, (c) the model with the
shallow FS, (d) the model with the deep FS.
2.3. Analysis of light propagation in the head models
Light propagation in the head models was calculated by use of a Monte Carlo
method [30-32]. Refractive index mismatch was considered only at the
interface between the scalp and air. The refractive indices of air and tissue
were 1.0 and 1.4, respectively. A fraction of the photon weight absorbed at each
interaction site was accumulated in each voxel in time-gates of width 50 ps to
calculate the time-resolved photon fluence. When a photon packet reached the
detector position, its ultimate weight was accumulated to calculate the detected
intensity. The path length of the detected photon packet in each voxel and each
type of tissues weighted by the ultimate weight was accumulated to calculate the
spatial sensitivity profile and partial optical path length in each type of
tissues. The partial optical path length can be used as an index of the
sensitivity of the NIRS signal to global absorption change in a particular type
of tissues while the spatial sensitivity profile indicates the volume of tissue
sampled by the source-detector pair of the NIRS instrument. The positions of the
sources and detector are shown in Fig.
2(b). The distance between each source-detector pair is 30 mm.
3. Results
3.1. Influence of the FS on light propagation in the head models
The time-resolved photon fluence maps in the head models with the whole FS and
without the FS shown in Figs. 3(a) and
3(b), respectively, were predicted in
order to evaluate the influence of the FS on light propagation in the head. A
pulse of light was irradiated at source position A in Fig. 2(b) and the photon fluence maps during every 50 ps
were recorded up to 2.5 ns. A frame from a movie (Media 1) of the time
resolved photon fluence map in the head models with parameter set (I) is shown
in Fig. 4
. When the photon fluence reaches the FS at about 250 ps, the photon
fluence around the FS rapidly increases. This causes the distortion of the
photon fluence map and the distribution of the photon fluence is broadened in
the direction of the FS. The penetration of the photon fluence tends to be
confined to the gray matter and the photon fluence laterally spreads in the gray
matter in both models. This phenomenon is caused by the low-scattering and
low-absorption characteristics of the CSF.
Fig. 4
Frame from a movie (Media
1) of light propagation in the
head model with the whole FS and without the FS. The red region
indicates the FS.
Frame from a movie (Media
1) of light propagation in the
head model with the whole FS and without the FS. The red region
indicates the FS.
3.2. Influence of the FS on the NIRS signal and spatial sensitivity
profile
Light propagation in the four head models with and without the FS was predicted
in order to evaluate the influence of the FS on the NIRS signal. The source at
position A or B and detector were placed on the forehead of the model as shown
in Fig. 2(b). In the case of source A,
the FS existed inside the distribution of the light path connecting the source
and detector, while the FS was almost outside of the light path in the case of
source B.The spatial sensitivity profiles for the source at position A and detector in the
four head models with parameter set (I) are shown in Fig. 5
. The spatial sensitivity profile is distorted by the FS and the
distortion depends on the depth of the FS. In the models with the FS, the
spatial sensitivity in the scalp around the midpoint between the source and
detector is significantly reduced and the sensitivity around the FS increased.
The depth profile of the spatial sensitivity along the line in Fig. 5 is shown in Fig. 6
. The whole FS and shallow FS reduce the penetration depth of the spatial
sensitivity profile. The deep FS increases the penetration depth of the spatial
sensitivity profile, however, the penetration depth in the gray matter is almost
the same as that in the model without the FS. The spatial sensitivity profiles
on the brain surface in the four models with parameter set (I) are shown in
Fig. 7
. The dashed line indicates the position of the FS. A hypothetical
position of the FS is also shown in the model without the FS for reference. The
spatial sensitivity profile on the brain in the model without the FS shown in
Fig. 7(b) is almost symmetrically
distributed with respect to the line connecting the source and detector while
that in the model with the whole FS shown in Fig. 7(a) is slightly shifted toward the lower side and the peak of
the sensitivity is reduced. The spatial sensitivity is narrowed in the model
with the shallow FS while the spatial sensitivity is broadened toward the lower
side in the model with the deep FS. These results indicate that the increase in
photon fluence around the FS affects the spatial sensitivity profile of the NIRS
measurements.
Fig. 5
The influence of the FS on the spatial sensitivity profile for the
source at position A and detector. (a) The model with the whole FS,
(b) the model without the FS, (c) the model with the shallow FS, (d)
the model with the deep FS. The line in the figures represents the
position of the depth profile shown in Fig. 6.
Fig. 6
The depth profile of the spatial sensitivity profile in the four head
models. The position of the profile is shown in Fig. 5.
Fig. 7
The influence of the frontal sinus on the spatial sensitivity profile
on the brain surface. (a) The model with the whole FS, (b) the model
without the FS, (c) the model with the shallow FS, (d) the model
with the deep FS. Dashed line indicates the position of the FS.
The influence of the FS on the spatial sensitivity profile for the
source at position A and detector. (a) The model with the whole FS,
(b) the model without the FS, (c) the model with the shallow FS, (d)
the model with the deep FS. The line in the figures represents the
position of the depth profile shown in Fig. 6.The depth profile of the spatial sensitivity profile in the four head
models. The position of the profile is shown in Fig. 5.The influence of the frontal sinus on the spatial sensitivity profile
on the brain surface. (a) The model with the whole FS, (b) the model
without the FS, (c) the model with the shallow FS, (d) the model
with the deep FS. Dashed line indicates the position of the FS.The tendency of the influence of the FS can be quantitatively evaluated by the
intensity and optical path length of the detected light shown in Table 2
. The partial optical path length in the scalp is reduced while
that in the skull including the FS is increased in all the models with the FS.
This indicates that the light which propagates through the FS tends to be
detected by the detector. Since light propagates in straight line without
absorption in the FS, the detected intensity is increased and the total optical
path length is decreased by the presence of the FS. It should be noted that the
partial optical path length in the FS is unexpectedly long although no
scattering can occur in the FS. This indicates that many photons scattered in
the skull passes through the FS without absorption. These photons increase the
photon fluence around the FS. The influence of the FS on the partial optical
path length in the gray matter depends on the depth of the FS. The deep FS
increases the partial optical path length in the gray matter while the shallow
FS decreases the partial optical path length in the gray matter.
Table 2
Intensity and path length of detected light for the head models
with different types of FS
Type
of frontal sinus
Detected intensity
Total
path length (mm)
Partial optical path length (mm)
Scalp
Skull
FS
CSF
Gray matter
White matter
(a) Whole FS
3.4 ×
10−8
166
88.4
67.7
37.6
7.81
2.06
0.253
(b) Without
FS
7.5 ×
10−9
230
115
101
0
11.3
2.71
0.397
(c) Shallow
FS
2.7 ×
10−8
179
91.6
80.7
29.0
5.23
1.35
0.163
(d) Deep FS
1.3 ×
10−8
210
100
94.7
16.6
12.4
3.01
0.404
Figure 8
shows the ratio of the partial optical path length in the gray matter for
the head models with the FS to that for the head model without the FS. The
results for the models with the three different parameter sets are compared in
order to investigate the dependence of the influence of the FS on the optical
properties of the head. In the head models with all the parameter sets, the deep
FS increases and the shallow FS decreases the partial optical path length in the
gray matter. However, in the case of the whole FS, the partial optical path
length in the gray matter for the models with parameter sets (II) and (III) are
increased by the influence of the FS while that in the model with parameter set
(I) is decreased. The tendency of the influence of the FS in the head model with
parameter set (III) is almost the same as that in the head model with parameter
set (II). This result indicates that the influence of the FS on the NIRS signal
depends on not only the depth of the FS but also the optical properties of the
superficial tissues such as the scalp and skull.
Fig. 8
The influence of the optical properties of the head model on the
partial optical path length in the gray matter. The partial optical
path length in the gray matter for the head models with the FS is
normalized by that for the head model without the FS. Each parameter
set is listed in Table
1.
The influence of the optical properties of the head model on the
partial optical path length in the gray matter. The partial optical
path length in the gray matter for the head models with the FS is
normalized by that for the head model without the FS. Each parameter
set is listed in Table
1.Figures 9(a)
and 9(b) show the spatial
sensitivity profiles for the source at position B and detector in the models
(parameter set (I)) with the FS shown in Fig.
3(a) and without the FS shown in Fig.
3(b), respectively. The spatial sensitivity profile in the skull is
significantly distorted toward the FS as shown in Fig. 9(a). Consequently, the spatial sensitivity profile
on the brain surface is broadened toward the direction of the FS. The influence
of the FS on the sensitivity of the NIRS signal depends on the relative position
of the source-detector pair and the FS.
Fig. 9
The influence of the FS on the spatial sensitivity profile for the
source at position B and detector. (a) The model with the whole FS.
The FS is almost outside of the distribution of the light path
connecting the source and detector. (b) The model without the
FS.
The influence of the FS on the spatial sensitivity profile for the
source at position B and detector. (a) The model with the whole FS.
The FS is almost outside of the distribution of the light path
connecting the source and detector. (b) The model without the
FS.
4. Discussion
Light propagation in tissue is known to be strongly affected by the presence of voids
and low-scattering regions. The light tends to propagate around the void region, and
consequently, the spatial sensitivity around the void region is increased. In the
brain activation measurements by NIRS, the low-scattering CSF affects light
propagation in the head and increases the partial optical path length, i.e., the
sensitivity of the NIRS signal, of the gray matter neighboring the CSF [14-16]. The FS also increases the spatial sensitivity around the FS, and
consequently, decreases that apart from the FS. Since the brain is not next to the
FS, the influence of the FS on the sensitivity of the NIRS signal is not consistent
and is more complicated than that of the CSF. The influence of the FS on the
sensitivity of the NIRS signal depends on the depth of the FS and the optical
properties of the scalp and skull. The shallow FS decreases the sensitivity of the
NIRS signal while the deep FS increases the sensitivity. The depth at which this
inversion based on the influence of the FS occurred is changed by the optical
properties of the scalp and skull. The optical properties of the tissues depend on
the wavelength and the wide range of different values for the same type of tissues
is obtained from the literature. The influence of the FS also depends on the
relative position of the source-detector pair and FS. When the FS exists almost
outside of the distribution of the light path connecting the source and the
detector, the spatial sensitivity profile tends to be broadened. Furthermore, the
realistic head models in this study do not consider the presence of the
extra-cerebral vasculature, such as sagittal sinus. The influence of the sagittal
sinus on the NIRS signal has been reported [21] and the sagittal sinus is located behind the middle of the FS.
Careful consideration is needed to estimate the influence of the FS on the
sensitivity of the NIRS signal.The FS tends to locate below the eyebrows, and in this case, the influence of the FS
on the NIRS measurements of the brain activation in the frontal area may be
minimized by the optimal probe arrangement. However, the depth and size of the FS is
different between individuals. It is necessary to analyze light propagation in the
3D head model, the anatomical structure of which is based on the structural image of
the individual if you need to obtain details of the influence of the FS on the NIRS
measurements.
5. Conclusions
The 3D realistic head models of which the structure is based on the MR and x-ray CT
images are constructed to investigate the influence of a FS on the brain function
measurements by near-infrared spectroscopy and imaging. Light propagation in the
head model with the FS of which the geometry was based on the CT images was compared
to that in the head models with the modified FS to investigate the influence of the
FS on the NIRS signal. Light propagation in the head is strongly affected by the
presence of the FS. The distribution of the photon fluence is broadened in the
direction of the FS, consequently, the spatial sensitivity around the FS tends to be
increased. Since the influence of the FS on light propagation in the brain depends
on the depth of the FS and the optical properties of the scalp and skull, the
influence of the FS on the sensitivity of the NIRS signal to the brain activation is
not consistent across the source-detector pairs. The shallow FS tends to reduce the
sensitivity of the NIRS signal, however, the deep FS can increase the sensitivity of
the NIRS signal.