Wear at the titanium-titanium and the titanium-zirconia implant-abutment interface: a comparative in vitro study.

Michael Stimmelmayr1, Daniel Edelhoff, Jan-Frederik Güth, Kurt Erdelt, Arndt Happe, Florian Beuer.   

Abstract

OBJECTIVE: The purpose of this study was to determine and measure the wear of the interface between titanium implants and one-piece zirconia abutments in comparison to titanium abutments.
METHODS: 6 implants were secured into epoxy resin blocks. The implant interface of these implants and 6 corresponding abutments (group Zr: three one-piece zirconia abutments; group Ti: three titanium abutments) were examined by a microscope and scanning electron micrograph (SEM). Also the implants and the abutments were scanned by 3D-Micro Computer Tomography (CT). The abutments were connected to the implants and cyclically loaded with 1,200,000 cycles at 100N in a two-axis fatigue testing machine. Afterwards, all specimens were unscrewed and the implants and abutments again were scanned by microscope, SEM and CT. The microscope and SEM images were compared, the CT data were superimposed and the wear was calculated by inspection software. The statistical analysis was carried out with an unpaired t-test.
RESULTS: Abutment fracture or screw loosening was not observed during cyclical loading. Comparing the microscope and SEM images more wear was observed on the implants connected to zirconia abutments. The maximum wear on the implant shoulder calculated by the inspection software was 10.2μm for group Zr, and 0.7μm for group Ti. The influence of the abutment material on the measured wear was statistically significant (p≤0.001; Levene-test). SIGNIFICANCE: Titanium implants showed higher wear at the implant interface following cyclic loading when connected to one-piece zirconia implant abutments compared to titanium abutments. The clinical relevance is not clear; hence damage of the internal implant connection could result in prosthetic failures up to the need of implant removal.
Copyright © 2012 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23021964     DOI: 10.1016/j.dental.2012.08.008

Source DB:  PubMed          Journal:  Dent Mater        ISSN: 0109-5641            Impact factor:   5.304


  17 in total

Review 1.  Silicon Nitride as a Biomedical Material: An Overview.

Authors:  Xiaoyu Du; Seunghun S Lee; Gurdial Blugan; Stephen J Ferguson
Journal:  Int J Mol Sci       Date:  2022-06-11       Impact factor: 6.208

2.  Experimental study on the factors affecting torque of beam-type implant torque wrenches.

Authors:  Hiroki Shiba; Yuji Sato; Junichi Furuya; Tokiko Osawa; Akio Isobe; Myu Hayashi; Noboru Kitagawa
Journal:  BMC Oral Health       Date:  2021-07-15       Impact factor: 2.757

3.  Microscopical and chemical surface characterization of CAD/CAM zircona abutments after different cleaning procedures. A qualitative analysis.

Authors:  Peter Gehrke; Astrid Tabellion; Carsten Fischer
Journal:  J Adv Prosthodont       Date:  2015-04-23       Impact factor: 1.904

Review 4.  The Role of Oral Cavity Biofilm on Metallic Biomaterial Surface Destruction-Corrosion and Friction Aspects.

Authors:  Joanna Mystkowska; Katarzyna Niemirowicz-Laskowska; Dawid Łysik; Grażyna Tokajuk; Jan R Dąbrowski; Robert Bucki
Journal:  Int J Mol Sci       Date:  2018-03-06       Impact factor: 5.923

5.  Sphere-plane methodology to evaluate the wear of titanium of dental implants: a research proposal.

Authors:  Teresa Almeida Mendes; João Caramês; Luís Pires Lopes; Amílcar Lopes Ramalho
Journal:  BMC Res Notes       Date:  2018-07-31

6.  Potential Causes of Titanium Particle and Ion Release in Implant Dentistry: A Systematic Review.

Authors:  Rafael Delgado-Ruiz; Georgios Romanos
Journal:  Int J Mol Sci       Date:  2018-11-13       Impact factor: 5.923

Review 7.  General review of titanium toxicity.

Authors:  Kyeong Tae Kim; Mi Young Eo; Truc Thi Hoang Nguyen; Soung Min Kim
Journal:  Int J Implant Dent       Date:  2019-03-11

8.  Tensile Bond Strength of Three Custom-made Tooth-Colored Implant Superstructures to Titanium Inserts.

Authors:  Mitra Zirak; Mahroo Vojdani; Amir Ali Reza Khaledi; Mitra Farzin
Journal:  J Dent (Shiraz)       Date:  2019-06

9.  Effects of Liner-Bonding of Implant-Supported Glass-Ceramic Crown to Zirconia Abutment on Bond Strength and Fracture Resistance.

Authors:  Yong-Seok Jang; Sang-Hoon Oh; Won-Suck Oh; Min-Ho Lee; Jung-Jin Lee; Tae-Sung Bae
Journal:  Materials (Basel)       Date:  2019-08-30       Impact factor: 3.623

10.  The Influence of Custom-Milled Framework Design for an Implant-Supported Full-Arch Fixed Dental Prosthesis: 3D-FEA Sudy.

Authors:  João Paulo Mendes Tribst; Amanda Maria de Oliveira Dal Piva; Roberto Lo Giudice; Alexandre Luiz Souto Borges; Marco Antonio Bottino; Ettore Epifania; Pietro Ausiello
Journal:  Int J Environ Res Public Health       Date:  2020-06-05       Impact factor: 3.390

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.