Independent verification of gantry angle for pre-treatment VMAT QA using EPID.

Justus Adamson1, Qiuwen Wu.   

Abstract

We propose a method to incorporate independent verification of gantry angle for electronic portal imaging device (EPID)-based pre-treatment quality assurance (QA) of clinical volumetric modulated arc therapy (VMAT) plans. Gantry angle is measured using projections in the EPID of a custom phantom placed on the couch and the treatment plan is modified so as to be incident on the phantom with a portion of the beam that is collimated in the clinical plan. For our implementation, collimator and couch angles were set to zero and the inferior jaw and two most inferior multi-leaf collimator pairs were opened for the entire QA delivery. A phantom containing five gold coils was used to measure the gantry rotation through which each portal image was acquired. We performed the EPID QA for ten clinical plans and evaluated accuracy of gantry angle measurement, scatter incident on the imager due to the phantom, inter-image pixel linearity and inter- and intra-image noise. The gantry angle could be measured to within 0.0 ± 0.3° for static gantry and 0.2 ± 0.2° for arc acquisitions. Scatter due to the presence of the phantom was negligible. The procedure was shown to be feasible and adds gantry angle to the treatment planning parameters that can be verified by EPID-based pre-treatment VMAT QA.

Entities:  

Mesh:

Year:  2012        PMID: 23010739     DOI: 10.1088/0031-9155/57/20/6587

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  Comparisons of volumetric modulated arc therapy (VMAT) quality assurance (QA) systems: sensitivity analysis to machine errors.

Authors:  Bin Liang; Bo Liu; Fugen Zhou; Fang-Fang Yin; Qiuwen Wu
Journal:  Radiat Oncol       Date:  2016-11-07       Impact factor: 3.481

2.  A novel and independent method for time-resolved gantry angle quality assurance for VMAT.

Authors:  Todsaporn Fuangrod; Peter B Greer; Benjamin J Zwan; Michael P Barnes; Joerg Lehmann
Journal:  J Appl Clin Med Phys       Date:  2017-07-13       Impact factor: 2.102

3.  An investigation of gantry angle data accuracy for cine-mode EPID images acquired during arc IMRT.

Authors:  Peter M McCowan; Daniel W Rickey; Pejman Rowshanfarzad; Peter B Greer; William Ansbacher; Boyd M McCurdy
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102