Influence of central corneal thickness on measured intraocular pressure differentials: Nidek RKT-7700, Topcon CT-80 NCTs and Goldmann Tonometer.

Kelechi C Ogbuehi1, Sven Mucke, Uchechukwu L Osuagwu.   

Abstract

PURPOSE: We sought to compare the intraocular pressure (IOP) measured by RKT-7700 and CT-80 noncontact tonometers (NCTs) with that measured by Goldmann applanation tonometry (GAT). We also examined the influence of central corneal thickness (CCT) on the agreement between both NCTs and GAT in a sub-population of healthy, young normals.
METHODS: Triplicate IOP and CCT measurements were obtained twice from one randomized eye of 49 subjects (28 males and 21 females) aged 22.2 ± 1.3 (mean ± S.D.) years. Goldmann tonometry was performed subsequent to assessment with the RKT-7700 and CT-80, to negate the 'ocular massage effect', followed by ultrasound pachymetry. The results from each method were compared and assessed for repeatability and between-observer reproducibility. Analysis was performed to determine the correlation between the differences in IOP measurements and corneal thicknesses.
RESULTS: The mean differences (±S.D.) in sessions 1 and 2 respectively between RKT-7700 and GAT (2.6 ± 2.0 and 2.7 ± 1.4 mmHg), between CT-80 and GAT (2.8 ± 2.0 and 3.2 ± 1.3 mmHg) were statistically significant (p < 0.0001). The repeatability coefficients in sessions 1 and 2 respectively were; ±1.2 and ±1.0 mmHg (GAT), ±1.3  and ±1.6 mmHg (CT-80), ±2.3  and ±1.7 mmHg (RKT-7700) and inter-observer reproducibility was; ±1.9 (RKT-7700), ±2.3 (GAT) and ±2.6 mmHg (CT-80). Between the corneal thickness and the differences in GAT-measured and NCT-measured IOPs, there was a significant negative correlation and a trend for a larger difference in thicker corneas.
CONCLUSION: Neither the RKT-7700 nor the CT-80 can be used interchangeably with the Goldmann tonometer, although all three tonometers give repeatable measurements of IOP, in this sub-population. IOP differences between GAT and the NCTs significantly correlated with CCT, with the possibility of even greater differences in thicker corneas. Ophthalmic & Physiological Optics
© 2012 The College of Optometrists.

Entities:  

Mesh:

Year:  2012        PMID: 23009324     DOI: 10.1111/j.1475-1313.2012.00945.x

Source DB:  PubMed          Journal:  Ophthalmic Physiol Opt        ISSN: 0275-5408            Impact factor:   3.117


  2 in total

1.  Comparison of Two Tonometers in the Evaluation of 24-Hour Intraocular Pressure and Mean Ocular Perfusion Pressure in Patients with Thyroid-Associated Ophthalmopathy.

Authors:  Weijie Liu; Yao Zhou; Xiaozhou Hu; Haochen Jin; Jie Ye; Mingna Xu; Zihui Liu; Wencan Wu; Yunhai Tu
Journal:  J Ophthalmol       Date:  2022-04-22       Impact factor: 1.974

2.  Repeatability and reproducibility of measurements using a NT-530P noncontact tono/pachymeter and correlation of central corneal thickness with intraocular pressure.

Authors:  Fusako Fujimura; Kazutaka Kamiya; Kazuko Fujiwara; Nobuyuki Shoji; Kimiya Shimizu
Journal:  Biomed Res Int       Date:  2013-10-08       Impact factor: 3.411

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.