Peak expiratory flow rate in a sample of normal saudi males at riyadh, saudi arabia.

A A Al-Taweel1, K A Kalantan, H A Ghani.   

Abstract

OBJECTIVE: To find out the normal peak expiratory flow rate for adult Saudi males and to compare our standards with British standards.
METHODS: A cross-sectional study was carried out in five primary health care centers representing Riyadh city in the period between 15(th) November through December 1993. Six hundred and eighty Saudi men who satisfied stringent criteria of normality were included in the study.
RESULTS: The mean and standard deviation of the subjects' age and height were 28.4 ± 13 and 167.6 ± 6.4 respectively. Linear regression analysis was performed through step-wise procedure to determine the form of regression of peak expiratory flow on age and height. Regression curves were obtained and it was found that peak expiratory flow rate did not begin to decline until about the age of 25 years; and as height increased the peak expiratory flow rate increased in a linear relationship.
CONCLUSION: It was demonstrated that our study group had lower peak expiratory flow rate compared with British people. These findings will serve as a basis for preparing flow rate values for our population.

Entities:  

Keywords:  Bronchial asthma; Peak expiratory flow rate; Saudi Arabia

Year:  1999        PMID: 23008593      PMCID: PMC3437070     

Source DB:  PubMed          Journal:  J Family Community Med        ISSN: 1319-1683


INTRODUCTION

Bronchial asthma is one of the most important and highly prevalent problems in Saudi Arabia. The prevalence of asthma in Saudi Arabia ranges between 10-17% (10% in Riyadh, 13% in Jeddah and Qassim and 17% in Abha); which is higher than the range reported in Western countries.12 Pulmonary function tests are important in the detection and evaluation of pulmonary dysfunction. Peak expiratory flow rate (PEFR) is the simplest estimate for lung functions. It indicates the severity of airflow limitation.34 Peak expiratory flow rate is measured by peak expiratory flow meter (PEFM) which is a simple and relatively cheap device (cost ranges from 25-40 US dollars). It has a great diagnostic and prognostic value in patients with hyperactive air-way disease.5–7 It is widely used in general practice in the UK and in family practice in the USA with proved importance in evaluating and monitoring patients with bronchial asthma.6 Although PEFM is a simple, practical and effective tool for use in the primary health care (PHC) setting, few of PHC practitioners in Saudi Arabia seem to use it.8 Many factors can affect the reading of peak expiratory flow rate (PEFR). The age, sex and height are important variables upon which peak expiratory flow depends.9 Other factors include the diurnal variation10 and ethnic differences.11 The degree to which both the instructor and the patient understand the technique of using the flow meter is important and can greatly affect the result. Widely used standard peak flow rates have not been established for our population and the PEFR tables in use are based on studies in the Western countries. Not many local studies on the PEFR of Saudis in PHC clinics have been done.1213 The objective of the current study was to identify the normal peak expiratory flow rate for a sample of Saudi males and to compare the findings with British standards.

MATERIAL AND METHODS

Riyadh (the capital of Saudi Arabia) is divided into five main sectors for the provision of the Health Services (i.e., North, South, East, West and Central). A number was given to each center in all sectors, and five primary health care centers were randomly selected to represent different geographical areas. Within each center, patients were enrolled randomly in a systematic way to ensure a good non-biased presentation. The sample size for the centers was nearly the same to make the allocation equal. Subjects included in the study were Saudi males aged 15 years and above, non-smokers who satisfied stringent criteria of normality.10 These were as follows 1) had never smoked; 2) were not subject to recurrent or persistent expectoration; 3) were not subject to wheezing nor had had an episode of acute bronchitis; 4) had not had asthma or recurrent bronchitis during their childhood; 5) had not had any serious respiratory disease. Since it was difficult for the subjects to recall one attack of acute bronchitis, this criterion was not used to exclude any. The number of subjects included in this study was 680. Each center was provided with mini-Wright peak flow meter (product of Ciement Clarke International Ltd, England), with disposable mouth pieces. All the mini-Wright peak expiratory flow meters were new and had been calibrated to make sure that they had similar readings. Only males were included in the study to ensure accessibility so that we could deal directly with the subjects and supervise more closely. It was agreed that all the tests be done in the afternoon sessions (4 -8 p.m.) to avoid the effect of the diurnal variation of peak expiratory flow10 and get more volunteers. The Director of each health center allocated one of the male nurses to work with the investigators. To ensure a unified correct procedure, all the nurses were instructed and trained by the assigned investigator on the technique of using the device. A group of subjects were instructed to do the test in front of both the nurse and the investigator to ensure that they perform it correctly and to use this as demonstration for the nurse. The instructions adopted14 for using the peak flow meter were as follows: 1) Every subject was asked to do the test three times. 2) Take a full deep breath in. 3) Hold the peak flow meter horizontally and close the lips tightly round the mouth piece. 4) Blow out as hard and as fast as you can in a short sharp blow. 5) Wait one minute between each blow. Take the best of 3 readings. This was recorded on the table along with subject's age and height. Data were analyzed using Stat Pac Gold statistical analysis package. Linear regression step-wise analysis was performed. The mean was expressed as mean ± 1 standard deviation.

RESULTS

The age distribution of a total of 680 subjects included in this study is shown in Table 1. The mean and standard deviation of their age and height were 28.4 ± 13 and 167.6 ± 6.4 respectively.
Table 1

Age distribution of subjects in series

Age distribution of subjects in series Linear regression analysis was performed through step-wise procedure to determine the form of regression of peak expiratory flow on age and height (Table 2). When the data was plotted it became evident that the relation between the variables was curvilinear (Figure 1). It showed that as age increased the PEFR decreased (b = -0.5, T = -2.5, p-value 0.012). However, it showed a positive association between height of the study population and PEFR (b=4.3, T=11.7, p-value < 0.0001). From the multiple regression equation R2 = 21% and F-ratio + 46%, p-value = <0.0001, i.e., about 21% variability of the PEFR can be determined by the factors in the equation.
Table 2

Regression Equations

Figure 1

Peak expiratory flow in normal males

Regression Equations Peak expiratory flow in normal males The regression curves were drawn for six heights, and standardized to 167 cm (Figure 1). The peak expiratory flow began to decline around the age of 25 (Figure 2). The height was plotted versus the peak expiratory flow and showed that there was a relatively linear relationship (Figure 3).
Figure 2

Comparison of PEF in normal subjects between British and Saudi study

Figure 3

Peak expiratory flow rate in normal Saudi males

Comparison of PEF in normal subjects between British and Saudi study Peak expiratory flow rate in normal Saudi males

DISCUSSION

We found that as the age increased beyond 25 years the PEFR significantly decreased. In calculating regression equations for peak expiratory flow on age and height, it was assumed that there was a linear relation between the fall of peak expiratory flow and advancing age. A similar assumption was made by most investigators who have published normal values of peak expiratory flow including Gregg and Nunn.10 In this study, it was clearly shown that a linear fall of peak expiratory flow occurred only after the age at which maximal peak expiratory flow was attained and if adolescents were included in the series the shape of regression became curvilinear. This finding is supported by a study done by Graff-lonnevig et al for Saudi children aged 6-16 years.12 Our findings suggest that peak expiratory flow began to decline at the age of 25 years while the decline occurred at the age of 35 years in British reports.10 Since our series contained too few subjects over 65 years of age (1.9%) the regression curves had a high peak expiratory flow rate for that particular age group. Studies of peak expiratory flow rate in the elderly have shown that the rate in elderly had a lower standard than younger age groups.1516 The comparison of the peak expiratory flow rates in this study with that of Gregg and Nunn10 demonstrated clearly that peak expiratory flow rates for Saudi males were lower than those of British males. This is supported by other local studies for other respiratory functions.13 Graff-Lonneving et al found that peak expiratory flow rate measured in Saudi children by means of the mini-Wright peak flow meter were significantly lower than in children from Europe and North America.12 However, a significant positive association was found between the height of the study population and the PEFR. The multiple regression equation R2 = 21% and fraction = 46% with p-value <0.0001, indicates a significant dependency of the PEFR of the study population on their age and height. One limitation of the current study was the non-inclusion of female subjects since certain cultural difficulties in performing the measurements were envisaged. The other limitation was that the computed equation was reliable in predicting only 21%, for the whole range of age groups. However, this study has shown that Saudi Arabian male adults living in Riyadh have lower peak expiratory flow rate than adults of different racial origins living in UK. There is a need for baseline normal values for peak expiratory flow rate for both children and adults in Saudi Arabia, and the data obtained from this study can be used as a base-line for future studies for the preparation of standard charts for our population. Further studies without the limitations of this study i.e., to include females and a wide range of patient characteristics are required.
  12 in total

1.  Guidelines for management of asthma in adults: I--Chronic persistent asthma. Statement by the British Thoracic Society, Research Unit of the Royal College of Physicians of London, King's Fund Centre, National Asthma Campaign.

Authors: 
Journal:  BMJ       Date:  1990-09-29

2.  Peak expiratory flow rate in an elderly population.

Authors:  N R Cook; D A Evans; P A Scherr; F E Speizer; S Vedal; L G Branch; J C Huntley; C H Hennekens; J O Taylor
Journal:  Am J Epidemiol       Date:  1989-07       Impact factor: 4.897

Review 3.  ABC of sleep disorders. Nocturnal asthma.

Authors:  J R Catterall; C M Shapiro
Journal:  BMJ       Date:  1993-05-01

4.  Circadian rhythms in peak expiratory flow rate in workers exposed to cotton dust.

Authors:  F F Cinkotai; T C Sharpe; A C Gibbs
Journal:  Thorax       Date:  1984-10       Impact factor: 9.139

5.  The Mini-Wright Peak Flow Meter for evaluating airway obstruction in a family practice.

Authors:  D N Katz
Journal:  J Fam Pract       Date:  1983-07       Impact factor: 0.493

6.  Asthma in Saudi Arabia: is the system appropriate for optimal primary care?

Authors:  T A Khoja; L A al-Ansary
Journal:  J Public Health Manag Pract       Date:  1998-05

7.  Evaluation of peak flow and symptoms only self management plans for control of asthma in general practice.

Authors:  I Charlton; G Charlton; J Broomfield; M A Mullee
Journal:  BMJ       Date:  1990-12-15

8.  Peak expiratory flow in normal subjects.

Authors:  I Gregg; A J Nunn
Journal:  Br Med J       Date:  1973-08-04

9.  Evaluation of home-monitoring of asthmatic children with the mini-Wright peak flow meter.

Authors:  K Battu; C Collins-Williams; C Zaleskey
Journal:  J Asthma       Date:  1982       Impact factor: 2.515

10.  Death from asthma in two regions of England. British Thoracic Association.

Authors: 
Journal:  Br Med J (Clin Res Ed)       Date:  1982-10-30
View more
  2 in total

1.  Issues in peak expiratory flow rate measurements: analysis and interpretation of results.

Authors:  K M Al-Dawood
Journal:  J Family Community Med       Date:  2000-09

2.  Predictive Regression Equations of Flowmetric and Spirometric Peak Expiratory Flow in Healthy Moroccan Adults.

Authors:  Khalid Bouti; Iliass Maouni; Jouda Benamor; Jamal Eddine Bourkadi
Journal:  Int Sch Res Notices       Date:  2017-03-30
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.