In vitro evaluation of cleaning efficacy of detergents recommended for use on dental instruments.

Gordon W G Smith1, June McNeil, Gordon Ramage, Andrew J Smith.   

Abstract

BACKGROUND: The cleaning stage of instrument decontamination processes is a critical control point, and removal of protein deposits is used as a marker of cleaning efficacy. An important factor is the choice of cleaning solution especially in the absence of any defined standards for detergent effectiveness.
METHODS: Following method validation, stainless steel tokens were inoculated with reconstituted citrated blood and added to a 24-multiwell plate and immersed in different cleaning solutions for 5 minutes, agitated at 25 (20°) tilts/min at 22°C and at the manufacturers' recommended temperatures. Desorbed protein was measured using the bicinchoninic acid assay.
RESULTS: From a starting concentration with a median of 3,700 μg of blood protein of all solutions tested, alkaline detergent (Haemo-sol) removed the largest proportion of protein (median, 2,070 μg), and surgical handwash removed the least protein (median, 0 μg). Reverse osmosis water demonstrated useful blood-removing properties with a median of 1,421 μg.
CONCLUSION: The cleaning system we utilized is a simple, inexpensive method to compare the cleaning efficacies of detergents and may be used as a first stage in benchmarking cleaning efficacy of detergents. Not all solutions used in cleaning dental instruments are efficacious at removal of blood.
Copyright © 2012 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22999772     DOI: 10.1016/j.ajic.2012.05.009

Source DB:  PubMed          Journal:  Am J Infect Control        ISSN: 0196-6553            Impact factor:   2.918


  1 in total

1.  Assessing the efficacy and cost of detergents used in a primary care automated washer disinfector.

Authors:  S Winter; G McDonagh; D Lappin; A J Smith
Journal:  Br Dent J       Date:  2018-08-24       Impact factor: 1.626

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.