Sara Watt1, Niels Agerlin, Bertil Romner.
Abstract
BACKGROUND: A new adjustable valve, the Codman CertasTM valve for treatment of hydrocephalus was introduced into clinical practice in January 2011. It has 8 different settings with an opening pressure varying from 36 to over 400 mm H2O at a flow rate of 20 mL/h. The 8th setting is designed to provide a "virtual off" function. The objective of this report is to describe the initial clinical experience with the CertasTM valve and evaluate clinical usage with the main focus on the portable adjustment device - Therapeutic Management System (TMS), the "virtual off" setting and compatibility with magnetic resonance imaging (MRI).Entities:
Year: 2012 PMID: 22995221 PMCID: PMC3490860 DOI: 10.1186/2045-8118-9-21
Source DB: PubMed Journal: Fluids Barriers CNS ISSN: 2045-8118
Figure 1Therapeutic management system for the Certasvalve. Locater tool (A), indicator tool (B), and adjustment tool (C)
Etiology of hydrocephalus and the number of patients who had adjustments of the valve setting
| Normal pressure hydrocephalus | 14 | 7 |
| Secondary to subarachnoid hemorrhage | 12 | 7 |
| Congenital | 4 | 3 |
| Secondary to traumatic brain injury | 3 | 3 |
| Secondary to intraventricular hemorrhage | 2 | 1 |
| Secondary to meningitis | 2 | 1 |
| Giant aneurism obstruction | 2 | 0 |
| Tumor obstruction | 1 | 2 |
| Syrinx obstruction | 1 | 1 |
| Aqueductal stenosis | 1 | 0 |
| Total | 42 | 25 |
Reasons for performing valve adjustments and number of adjustments
| Optimising clinical status due to clinical symptoms of: | |
| Normal pressure hydrocephalus | 10 |
| Over drainage | 7 |
| Under drainage | 10 |
| CT- results showing large ventricles | 9 |
| Hygromas | 5 |
| Diminished hygromas | 5 |
| Non-specific headache | 17 |
| Dizziness | 8 |
| Total | 71 |
Relevant individual features of adjustable valves
| Yes | Yes | Yes | Yes | Yes | |
| Yes | Yes | No | No | No | |
| 30-200 | 15-135 | 30-200 | 0-200 | 36-400 | |
| 18 | 5 | 5* | 20 | 8 | |
| No | No | No | No | Yes |
*Available in 4 different setting intervals with 5 possible settings.