Michael B Townsend1, Adam MacNeil, Mary G Reynolds, Christine M Hughes, Victoria A Olson, Inger K Damon, Kevin L Karem.
Abstract
The commercially available Orthopox BioThreat® Alert assay for orthopoxvirus (OPV) detection is piloted. This antibody-based lateral-flow assay labels and captures OPV viral agents to detect their presence. Serial dilutions of cultured Vaccinia virus (VACV) and Monkeypox virus (MPXV) were used to evaluate the sensitivity of the Tetracore assay by visual and quantitative determinations; specificity was assessed using a small but diverse set of diagnostically relevant blinded samples from viral lesions submitted for routine OPV diagnostic testing. The BioThreat® Alert assay reproducibly detected samples at concentrations of 10(7)pfu/ml for VACV and MPXV and positively identified samples containing 10(6)pfu/ml in 4 of 7 independent experiments. The assay correctly identified 9 of 11 OPV clinical samples and had only one false positive when testing 11 non-OPV samples. Results suggest applicability for use of the BioThreat® Alert assay as a rapid screening assay and point of care diagnosis for suspect human monkeypox cases.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22981983 PMCID: PMC9534008 DOI: 10.1016/j.jviromet.2012.08.023
Source DB: PubMed Journal: J Virol Methods ISSN: 0166-0934 Impact factor: 2.623
Fig. 1Tetracore Orthopox cassette image and strip layout showing application port, the sample test line for the orthopox detection, and a control line to indicate the test was successfully completed.
Visual and quantitative results using laboratory preparations of orthopoxviruses.
| Vaccinia virus | Monkeypox virus | |||||||
|---|---|---|---|---|---|---|---|---|
| Visual ID | Quantitative results (test 1 and 2 only) | Visual ID | ||||||
| Test 1 | Test 2 | Test 3 | Test 4 | Test 5 | Peak area | Test 1 | Test 2 | |
| 108 pfu/ml | Y | Y | Y | Y | Y | 2248 ± 446 | Y | Y |
| 107 pfu/ml | Y | Y | Y | Y | Y | 2346 ± 85 | Y | Y |
| 106 pfu/ml | Y | N | N | N | Y | 165 ± 69 | Y | Y |
| 105 pfu/ml | N | N | N | N | N | 18 ± 6 | N | N |
| 104 pfu/ml | N | N | N | N | N | 32 ± 9 | N | N |
| Crude VZV | N | N | – | – | – | 26 | – | – |
| HLF lysate | N | N | – | – | – | 22 | – | – |
| BSC-40 lysate | N | N | – | – | – | 26 | – | – |
Identified positive (Y) or negative (N) by two independent users 15 min after sample application.
Peak area measurements are defined in Section 2.3.1.
Average and standard deviation measurements, no cutoff was defined by the ESE reader.
Fig. 2Sensitivity of BioThreat Alert reader scanned strips to (a) lab grown vaccinia virus and (b) negative controls. The manufacturer cutoff for the reader was 0.01. Samples were applied in duplicate to test strips and measured at approximately 7.5 min intervals. Not all samples had measurements taken at each interval.
Blind study results for Orthopox diagnostic remainder clinical specimens (tested in duplicate).
| Sample | Visual ID | Line intensity | Sample type | Days post rash onset | Virus | CT | PFU/ml | Correct? |
|---|---|---|---|---|---|---|---|---|
| 1 | + | Med | Vesicle/skin | 3 | Vaccinia | 22 | 4.6E + 03 | ✓ |
| + | Med | ✓ | ||||||
| 2 | + | Med | Swab | Unknown | Vaccinia | Nd | 6.1E + 05 | ✓ |
| + | Med | ✓ | ||||||
| 3 | −/+ | Faint | Swab | 11 | Vaccinia | 21 | 4.9E + 04 | × |
| − | – | × | ||||||
| 4 | + | Weak | Swab | 12 | Vaccinia | 19 | 1.2E + 05 | ✓ |
| + | Weak | ✓ | ||||||
| 5 | + | Faint | Vesicle/skin | 3 | Vaccinia | 22 | Nd | ✓ |
| + | Weak | ✓ | ||||||
| 6 | + | Weak | Scab | Unknown | Monkeypox | 15 | Nd | ✓ |
| + | Med | ✓ | ||||||
| 7 | + | Faint | Swab | Unknown | Monkeypox | 18 | Nd | ✓ |
| + | Faint | ✓ | ||||||
| 8 | − | – | Swab | Unknown | Monkeypox | 22 | 1.0E + 01 | × |
| − | – | × | ||||||
| 9 | + | Med | Scab | Unknown | Monkeypox | 16 | Nd | ✓ |
| + | Med | ✓ | ||||||
| 10 | + | Faint | Swab | Unknown | Monkeypox | 17 | 2.8E + 03 | ✓ |
| + | Faint | ✓ | ||||||
| 11 | + | Weak | Swab | Unknown | Monkeypox | 15 | 1.0E + 01 | ✓ |
| + | Faint | ✓ | ||||||
Sample became positive after scheduled 15 min read time.
Not determined.
Re-extracted samples.