Literature DB >> 22974168

Intentionally retained pressure-regulating balloon in artificial urinary sphincter revision.

Tobias S Köhler1, Aaron Benson, Loren Ost, Steven K Wilson, William O Brant.   

Abstract

INTRODUCTION: The traditional surgical approach for revision of a malfunctioning artificial urinary sphincter (AUS) includes removal and replacement of all device components, identical to that employed in the setting of an infected or eroded AUS. AIMS: To describe outcomes of our technique in which we intentionally leave behind the original pressure-regulating balloon (PRB) at time of AUS revision in a clinically non-infected setting.
METHODS: We retrospectively reviewed our combined institutional series of 35 patients who underwent 36 AUS revisions in which the original pressure-regulating balloon was left undisturbed. We removed and replaced the defective cuff and pump through a single peno-scrotal incision for most patients requiring revision of a non-infected AUS. The new PRB was then placed on the opposite side through this single incision. MAIN OUTCOME MEASURES: Assessment of outcomes, complication, and infection rate of this surgical series.
RESULTS: All of the patients had the original pressure-regulating balloon placed through an inguinal counter-incision. Mean follow-up time was 14 months (2-33 months). Overall complication rate for the revision series was 11%. No infections or complications occurred secondary to the retained PRB.
CONCLUSION: Follow-up of our series provides evidence that retention of the original PRB at the time of non-infected AUS revision is safe. Potential advantages include elimination of a counter incision and technically challenging exploration. By eliminating these aspects, the potentially resultant decreased operative time may help counter the theoretical yet unproven risk of infection from leaving the old PRB in place.
© 2012 International Society for Sexual Medicine.

Entities:  

Keywords:  Artificial Urinary Sphincter; Erosion; Fluid Reservoir of Inflatable Penile Implant; Infection; Pressure-Regulating Balloon

Mesh:

Year:  2012        PMID: 22974168     DOI: 10.1111/j.1743-6109.2012.02914.x

Source DB:  PubMed          Journal:  J Sex Med        ISSN: 1743-6095            Impact factor:   3.802


  2 in total

Review 1.  Reservoir Placement Considerations During Inflatable Penile Prosthesis Surgery.

Authors:  Steven J Hudak
Journal:  Curr Urol Rep       Date:  2019-01-28       Impact factor: 3.092

2.  Infrequent Reservoir-Related Complications of Urologic Prosthetics: A Case Series and Literature Review.

Authors:  Tao Cui; Ryan Terlecki; Majid Mirzazadeh
Journal:  Sex Med       Date:  2015-10-19       Impact factor: 2.491