A S Thurmond1, B T Uchida, J Rösch.
Abstract
A device was developed for hysterosalpingography and fallopian tube recanalization. It differs from the previously used vacuum-cup device in that the central shaft slides and has an acorn-shaped tip. Optimal results were obtained in all 14 women (100%) who underwent catheterization with the new device; optimal results were achieved in only five of eight women (62%) who underwent catheterization with the fixed-shaft device during the same time period.Entities:
Mesh:
Year: 1990 PMID: 2296670 DOI: 10.1148/radiology.174.2.2296670