| Literature DB >> 22960003 |
Hamdi Torun1, Ofer Finkler, F Levent Degertekin.
Abstract
The authors describe a method for biomolecular force clamp measurements using atomic force microscope (AFM) cantilevers and micromachined membrane-based electrostatic actuators. The actuators comprise of Parylene membranes with embedded side actuation electrodes and are fabricated on a silicon substrate. The devices have a displacement range of 1.8 μm with 200 V actuation voltage, and displacement uncertainty is 0.8 nm, including the noise and drift. The settling time, limited by the particular amplifier is 5 ms, with an inherent range down to 20 μs. A force clamp measurement setup using these actuators in a feedback loop has been used to measure bond life-times between human IgG and anti-human IgG molecules to demonstrate the feasibility of this method for biological experiments. The experimental findings are compared with a molecular pulling experiment and the results are found to be in good agreement.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22960003 PMCID: PMC3515702 DOI: 10.1016/j.ultramic.2012.07.023
Source DB: PubMed Journal: Ultramicroscopy ISSN: 0304-3991 Impact factor: 2.689