| Literature DB >> 22959014 |
Hui Xu1, Karine Gorgy, Chantal Gondran, Alan Le Goff, Nicolas Spinelli, Christian Lopez, Eric Defrancq, Serge Cosnier.
Abstract
A label-free and highly sensitive impedimetric aptasensor was developed based on electropolymerized film for the determination of thrombin. The first step is the electrogeneration of a poly(pyrrole-nitrilotriacetic acid) (poly(pyrrole-NTA)) film onto the surface of electrodes followed by complexation of Cu(2+) ions. Then, the histidine labeled thrombin aptamer was immobilized onto the electrode through coordination of the histidine groups on the NTA-Cu(2+) complex. The aptamer sensor was applied for the detection and quantification of thrombin via impedimetric detection without a labeling step. A linear quantification of thrombin was obtained in the range 4.7×10(-12)-5.0×10(-10) mol L(-1) with a sensitivity of 2838 Ω/log unit (R(2)=0.9984). The impedance modulus at 0.3 Hz as a function of thrombin concentration was used to elaborate a similar linear relationship from 4.7×10(-12) to 5×10(-10) mol L(-1). In addition, aptamer-poly(pyrrole-NTA) electrodes incubated for 40 min in aqueous solutions of bovine serum albumin (BSA), lysozyme and IgG (5×10(-7) mol L(-1)) did not exhibit non-specific adsorption of proteins. Moreover, it has been demonstrated that the selective sensor can be regenerated several times with a good reproducibility.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22959014 DOI: 10.1016/j.bios.2012.07.044
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618