Literature DB >> 22956615

A model to measure fluid outflow in rabbit capsules post glaucoma implant surgery.

Dan Q Nguyen1, Craig M Ross, Yu Qin Li, Surinder Pandav, Bruce Gardiner, David Smith, Alicia C How, Jonathan G Crowston, Michael A Coote.   

Abstract

PURPOSE: Prior models of glaucoma filtration surgery assess bleb morphology, which does not always reflect function. Our aim is to establish a model that directly measures tissue hydraulic conductivity of postsurgical outflow in rabbit bleb capsules following experimental glaucoma filtration surgery.
METHODS: Nine rabbits underwent insertion of a single-plate pediatric Molteno implant into the anterior chamber of their left eye. Right eyes were used as controls. The rabbits were then allocated to one of two groups. Group one had outflow measurements performed at 1 week after surgery (n = 5), and group two had measurements performed at 4 weeks (n = 4). Measurements were performed by cannulating the drainage tube ostium in situ with a needle attached to a pressure transducer and a fluid column at 15 mm Hg. The drop in the fluid column was measured every minute for 5 minutes. For the control eyes (n = 6), the anterior chamber of the unoperated fellow eye was cannulated. Animals were euthanized with the implant and its surrounding capsule dissected and fixed in 4% paraformaldehyde, and embedded in paraffin before 6-μm sections were cut for histologic staining.
RESULTS: By 7 days after surgery, tube outflow was 0.117 ± 0.036 μL/min/mm Hg at 15 mm Hg (mean ± SEM), whereas at 28 days, it was 0.009 ± 0.003 μL/min/mm Hg. Control eyes had an outflow of 0.136 ± 0.007 μL/min/mm Hg (P = 0.004, one-way ANOVA). Hematoxylin and eosin staining demonstrated a thinner and looser arrangement of collagenous tissue in the capsules at 1 week compared with that at 4 weeks, which had thicker and more densely arranged collagen.
CONCLUSIONS: We describe a new model to directly measure hydraulic conductivity in a rabbit glaucoma surgery implant model. The principal physiologic endpoint of glaucoma surgery can be reliably quantified and consistently measured with this model. At 28 days post glaucoma filtration surgery, a rabbit bleb capsule has significantly reduced tissue hydraulic conductivity, in line with loss of implant outflow facility, and increased thickness and density of fibrous encapsulation.

Entities:  

Mesh:

Year:  2012        PMID: 22956615     DOI: 10.1167/iovs.12-10438

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  6 in total

1.  Porosity of Bleb Capsule declines rapidly with Fluid Challenge.

Authors:  Surinder S Pandav; Craig M Ross; Faisal Thattaruthody; Ritambhra Nada; Nirbhai Singh; Natasha Gautam; Stephen Beirne; Gordon G Wallace; Mark B Sherwood; Jonathan G Crowston; Michael Coote
Journal:  J Curr Glaucoma Pract       Date:  2016-10-29

2.  Blebs, Barriers, and Bagpipes: Why is It so Hard?

Authors:  Michael Coote
Journal:  J Curr Glaucoma Pract       Date:  2016-10-29

3.  The Inflammatory and Foreign Body Reaction of Polymethyl Methacrylate Glaucoma Drainage Device in the Rabbit Eye.

Authors:  Virna D Oktariana Asrory; Ratna Sitompul; Widya Artini; Sri Estuningsih; Deni Noviana; William H Morgan
Journal:  Transl Vis Sci Technol       Date:  2020-02-18       Impact factor: 3.283

4.  Comparison of two prototypes of a magnetically adjustable glaucoma implant in rabbits.

Authors:  Birthe Schwerk; Lisa Harder; Claudia Windhövel; Marion Hewicker-Trautwein; Anna Wagner; Jan-Peter Bach; Lena Carolin Voigt; Ulf Hinze; Boris Chichkov; Heinz Haferkamp; Holger Lubatschowski; Stephan Nikolic; Ingo Nolte
Journal:  PLoS One       Date:  2019-04-11       Impact factor: 3.240

5.  Novel approaches to model effects of subconjunctival blebs on flow pressure to improve clinical grading systems after glaucoma drainage surgery.

Authors:  Yann Bouremel; Richard M H Lee; Ian Eames; Steve Brocchini; Peng Tee Khaw
Journal:  PLoS One       Date:  2019-10-24       Impact factor: 3.240

6.  A model for designing intraocular pressure-regulating glaucoma implants.

Authors:  Inês C F Pereira; Hans M Wyss; Leonard Pinchuk; Henny J M Beckers; Jaap M J den Toonder
Journal:  PLoS One       Date:  2022-09-02       Impact factor: 3.752