Literature DB >> 22955052

Delivery of titrated oxygen via a self-inflating resuscitation bag.

Paul Young1, Janine Pilcher, Mitesh Patel, Laird Cameron, Irene Braithwaite, Mark Weatherall, Richard Beasley.   

Abstract

AIM: To investigate whether titration of inspired oxygen can be achieved through adjustment of oxygen flow into a self-inflating resuscitation bag with a reservoir of a type used in standard ambulance practice.
METHODS: In a series of bench experiments, oxygen was delivered via a flow metre to a 1500 ml self-inflating resuscitation bag with a 2500 ml reservoir bag and connected to a test lung. The oxygen concentration delivered to the test lung by manual inflation of the resuscitation bag was measured using an anaesthetic machine while the delivered tidal volume was measured using a respirometer. The delivered oxygen concentration was measured at flows of 0.5, 2, 6, 12 and 15 l min(-1) for tidal volumes of 300, 600, and 900 ml with bag inflation rates of 10, 20 and 30 min(-1).
RESULTS: A wide range of delivered oxygen concentrations ranging between 24% and 99.5% were achieved with different oxygen flows, tidal volumes, and inflation rates. Overall, the mean delivered oxygen concentration increased significantly with each of the increments of oxygen flow tested (p<0.001 for all comparisons).
CONCLUSIONS: Effective titration of oxygen delivery can be achieved using adjustment of oxygen flow with a standard self-inflating resuscitation bag and reservoir.
Copyright © 2012 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22955052     DOI: 10.1016/j.resuscitation.2012.08.330

Source DB:  PubMed          Journal:  Resuscitation        ISSN: 0300-9572            Impact factor:   5.262


  4 in total

Review 1.  Towards an integrated framework for air quality monitoring and exposure estimation-a review.

Authors:  Savina Singla; Divya Bansal; Archan Misra; Gaurav Raheja
Journal:  Environ Monit Assess       Date:  2018-08-31       Impact factor: 2.513

2.  European Resuscitation Council and European Society of Intensive Care Medicine guidelines 2021: post-resuscitation care.

Authors:  Jerry P Nolan; Claudio Sandroni; Bernd W Böttiger; Alain Cariou; Tobias Cronberg; Hans Friberg; Cornelia Genbrugge; Kirstie Haywood; Gisela Lilja; Véronique R M Moulaert; Nikolaos Nikolaou; Theresa Mariero Olasveengen; Markus B Skrifvars; Fabio Taccone; Jasmeet Soar
Journal:  Intensive Care Med       Date:  2021-03-25       Impact factor: 17.440

3.  Potential Maneuvers for Providing Optimal Tidal Volume Using the One-Handed EC Technique.

Authors:  Dongchoon Uhm; Ajung Kim
Journal:  Healthcare (Basel)       Date:  2022-07-23

4.  Tidal volume according to the 4-point sealing forces of a bag-valve-mask: an adult respiratory arrest simulator-based prospective, descriptive study.

Authors:  Dongchoon Uhm; Ajung Kim
Journal:  BMC Emerg Med       Date:  2021-05-01
  4 in total