The effects of electrode misplacement on clinicians' interpretation of the standard 12-lead electrocardiogram.

R R Bond1, D D Finlay, C D Nugent, C Breen, D Guldenring, M J Daly.   

Abstract

INTRODUCTION: This study investigates how a particular incorrect electrode configuration affects the 12-lead Electrocardiogram (ECG).
METHODS: A correct and an incorrect 12-lead ECG were extracted from a 192-lead BSPM. This was done for 232 BSPMs yielding 464 12-lead ECGs. The particular incorrect ECG involved displacing electrodes V1 and V2 in the second intercostal space whilst also offsetting the remaining electrodes. These ECGs were examined in two stages: (a) analysis of the effects of electrode misplacement on signal morphology and (b) analysis of how often the incorrect electrode configuration changed the diagnosis of two clinicians in a random sample of 75 patients.
RESULTS: According to the Root Mean Square Error (RMSE) of the difference between PQRST intervals in the correct and incorrect ECGs, lead V2 is the most affected lead (mean: 185 μV ± 82 μV), followed by lead V4 (mean: 114 μV ± 59 μV) and lead V1 (mean: 100 μV ± 47 μV). It was found that if the incorrect electrode configuration is applied, there is a 17% to a 24% chance the diagnostic interpretation will be different. Quantified using Similarity Coefficient (SC) leads V1 and V2 were found to be more alike when misplaced in the second intercostal space. The average SC between these leads when correctly placed was 0.08 (± 0.65), however when incorrectly placed, the average SC was 0.43 (± 0.3).
CONCLUSION: There is a reasonable chance this particular incorrect electrode configuration will change the diagnosis of the 12-lead ECG. This highlights the importance of developing algorithms to detect electrode misplacement along with better education regarding ECG acquisition.
Copyright © 2012 European Federation of Internal Medicine. Published by Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22939805     DOI: 10.1016/j.ejim.2012.03.011

Source DB:  PubMed          Journal:  Eur J Intern Med        ISSN: 0953-6205            Impact factor:   4.487


  8 in total

1.  Significance of a negative sinus P wave in lead V2 of the clinical electrocardiogram.

Authors:  Rex N MacAlpin
Journal:  Ann Noninvasive Electrocardiol       Date:  2017-02-18       Impact factor: 1.468

2.  Assessment of cardiac variables using a new electrocardiography lead system in horses.

Authors:  Worakij Cherdchutham; Kanoklada Koomgun; Suchanan Singtoniwet; Napattra Wongsutthawart; Napass Nontakanun; Wipasitnee Wanmad; Soontaree Petchdee
Journal:  Vet World       Date:  2020-06-29

3.  The Reconstruction of a 12-Lead Electrocardiogram from a Reduced Lead Set Using a Focus Time-Delay Neural Network.

Authors:  Gerard H Smith; Dawie J Van den Heever; Wayne Swart
Journal:  Acta Cardiol Sin       Date:  2021-01       Impact factor: 2.672

4.  Chest-lead ST-J amplitudes using arm electrodes as reference instead of the Wilson central terminal in smartphone ECG applications: Influence on ST-elevation myocardial infarction criteria fulfillment.

Authors:  Thomas Lindow; Henrik Engblom; Ardavan Khoshnood; Ulf Ekelund; Marcus Carlsson; Olle Pahlm
Journal:  Ann Noninvasive Electrocardiol       Date:  2018-05-07       Impact factor: 1.468

5.  The effect of precordial lead displacement on ECG morphology.

Authors:  Michał Kania; Hervé Rix; Małgorzata Fereniec; Heriberto Zavala-Fernandez; Dariusz Janusek; Tomasz Mroczka; Günter Stix; Roman Maniewski
Journal:  Med Biol Eng Comput       Date:  2013-10-19       Impact factor: 2.602

6.  Perspectives of human verification via binary QRS template matching of single-lead and 12-lead electrocardiogram.

Authors:  Vessela Krasteva; Irena Jekova; Ramun Schmid
Journal:  PLoS One       Date:  2018-05-17       Impact factor: 3.240

7.  False Alarm Reduction in Self-Care by Personalized Automatic Detection of ECG Electrode Cable Interchanges.

Authors:  Jocelyne Fayn; Paul Rubel
Journal:  Int J Telemed Appl       Date:  2020-01-27

8.  Accuracy of ECG chest electrode placements by paramedics: an observational study.

Authors:  Pete Gregory; Tim Kilner; Stephen Lodge; Suzy Paget
Journal:  Br Paramed J       Date:  2021-05-01
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.