A clinical comparison of the performance of four blood glucose reagent strips.

R D Cheeley1, S M Joyce.   

Abstract

The widespread use of visually read blood glucose reagent strips for initiation of emergent treatment of hypoglycemia and hyperglycemia has produced concern over the accuracy of this method. This study evaluated the accuracy of Chemstrip bG (Bio-Dynamics, Boehringer Mannheim, Indianapolis, IN), Dextrostix (Ames, Miles Laboratories, Elkhart, IN), Glucostix (Ames), and Visidex II (Ames) as compared with hospital laboratory values in an emergency department (ED) setting. Blood samples from 96 ED patients were tested for glucose concentration by each of the four strips and by the hospital laboratory. Each strip was evaluated for sensitivity, specificity, correlation coefficient (r), 95% confidence intervals, and kappa statistic (kappa, a measure of agreement between nonparametric data) using laboratory values as reference. In addition, six observers scored each strip for ease of interpretation using an ordinal scale of 1 (poor) to 4 (excellent). From the samples, no patients were hypoglycemic (less than or equal to 60 mg/dL), 83 were euglycemic (greater than 60 and less than 160 mg/dL), and 13 were hyperglycemic (greater than or equal to 160 mg/dL). Results suggest that over the range of glucose concentration sampled, there is good to excellent correlation with laboratory values for all strips except Dextrostix. The lower r value for Dextrostix is in part artifact due to limitation of its range of measurement to less than or equal to 250 mg/dL. Decreased accuracy for all strips in the hyperglycemic range may have been attributable to small sample size. Chemstrip bG and Visidex II were found to be subjectively easier to interpret.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Substances:

Year:  1990        PMID: 2293826     DOI: 10.1016/0735-6757(90)90286-9

Source DB:  PubMed          Journal:  Am J Emerg Med        ISSN: 0735-6757            Impact factor:   2.469


  2 in total

1.  Analysis of Paper-Based Colorimetric Assays With a Smartphone Spectrometer.

Authors:  Elizabeth V Woodburn; Kenneth D Long; Brian T Cunningham
Journal:  IEEE Sens J       Date:  2019-06-15       Impact factor: 3.301

2.  Sticky fingers.

Authors:  C Watts; J Stevenson
Journal:  J Accid Emerg Med       Date:  1995-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.