AIM: Our manuscript proposes recommendations to standardise implantable cardioverter-defibrillator (ICD) programming practice in New Zealand (NZ) and provides a review of the literature behind such an initiative.
BACKGROUND: Shocks have traditionally been the mainstay of ICD therapy for ventricular tachyarrhythmia. There is substantial evidence to suggest shocks may cause psychological harm, are often painful and may increase hospitalisation for heart failure. Shock therapy may be reduced by avoiding unnecessary therapy of non-sustained arrhythmia, utilising anti-tachycardia pacing to terminate ventricular tachycardia and by using algorithms to identify non-ventricular rhythms with rates overlapping therapy zones. This manuscript outlines evidence-based strategies to safely reduce unnecessary shocks from ICDs and includes manufacturer-specific parameters for ease of practical use.
CONCLUSION: These recommendations aim to optimise programming of ICD devices in NZ. Developed from a substantial evidence base they are intended to align national programming practice in the hope that we achieve fewer hospitalisations, improved quality of life and possibly achieve greater survival for our ICD-treated patients.
Copyright © 2012 Australian and New Zealand Society of Cardiac and Thoracic Surgeons (ANZSCTS) and the Cardiac Society of Australia and New Zealand (CSANZ). Published by Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22926144     DOI: 10.1016/j.hlc.2012.07.017

Source DB:  PubMed          Journal:  Heart Lung Circ        ISSN: 1443-9506            Impact factor:   2.975


  3 in total

1.  Analysis of ECG Measures of Cardiac Repolarization in Relation to Arrhythmic Events in an Implantable Cardioverter Defibrillator Population.

Authors:  Bijia Shi; Scott Harding; Peter Larsen
Journal:  Ann Noninvasive Electrocardiol       Date:  2016-06-13       Impact factor: 1.468

2.  Inappropriate Detection of a Supraventricular Tachycardia as Dual Tachycardia by the PR Logic™ Algorithm.

Authors:  Ajit Thachil; Sridevi Chennapragada; Narasimhan Calambur
Journal:  Indian Pacing Electrophysiol J       Date:  2014-05-25

3.  2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing.

Authors:  Bruce L Wilkoff; Laurent Fauchier; Martin K Stiles; Carlos A Morillo; Sana M Al-Khatib; Jesœs Almendral; Luis Aguinaga; Ronald D Berger; Alejandro Cuesta; James P Daubert; Sergio Dubner; Kenneth A Ellenbogen; N A Mark Estes; Guilherme Fenelon; Fermin C Garcia; Maurizio Gasparini; David E Haines; Jeff S Healey; Jodie L Hurtwitz; Roberto Keegan; Christof Kolb; Karl-Heinz Kuck; Germanas Marinskis; Martino Martinelli; Mark McGuire; Luis G Molina; Ken Okumura; Alessandro Proclemer; Andrea M Russo; Jagmeet P Singh; Charles D Swerdlow; Wee Siong Teo; William Uribe; Sami Viskin; Chun-Chieh Wang; Shu Zhang
Journal:  J Arrhythm       Date:  2016-02-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.