Vassiliki Vlachopoulou1, Georgia Malatara, Harry Delis, Dimitrios Kardamakis, George Panayiotakis.
Abstract
The purpose of this study was to measure the peripheral dose which is the absorbed dose in organs located outside the treatment volume such as the thyroid gland and the breast in patients undergoing radiotherapy, utilising the MOSFET dosemeters, as well as to estimate the probability of secondary cancer. The thyroid gland doses, expressed as a percentage of the prescribed dose (%TD), were measured to be 2.0±0.3 %, in whole brain irradiation, 10.0±8.0 % in mediastinum treatment and 8.0±2.0 and 2.0±0.8 % in breast treatment, with and without the supraclavicular irradiation, respectively, with a corresponding risk of 0.2, 2.0, 1.0 and 0.3 %. The dose to the breast was 7.0±2.0 %, in the mediastinum treatment, and 4.0±1.0 and 2.0±0.8 %, in the breast treatment, with and without supraclavicular irradiation, respectively, with a corresponding risk of 4.0, 2.0 and 1.0 %. Although the results indicate that the risk is not negligible, its significance should be considered in conjunction with the existing pathology and age of the patients.Entities:
Mesh:
Year: 2012 PMID: 22923288 DOI: 10.1093/rpd/ncs134
Source DB: PubMed Journal: Radiat Prot Dosimetry ISSN: 0144-8420 Impact factor: 0.972