First clinical evaluation of a new long-term subconjunctival glucose sensor.

Achim Josef Müller1, Monika Knuth, Katharina Sibylle Nikolaus, Peter Herbrechtsmeier.   

Abstract

BACKGROUND: To evaluate the feasibility of an implantable subconjunctival glucose monitoring system (SGMS) for glucose monitoring in humans, we investigated the in vivo performance of the sensor in a clinical trial with five patients.
METHODS: The new SGMS consists of an implantable ocular mini implant (OMI) and a hand-held fluorescence photometer. The implantable subconjunctival glucose sensor is composed of a fluorescence resonance energy transfer system based on Concanavalin A chemistry, embedded in a nelfilcon polymer hydrogel disk. Blood glucose changes in humans were induced by oral glucose intake and insulin injections.
RESULTS: The in vivo response of the new SGMS was tested in a first human clinical study with five diabetes patients. The OMI was well tolerated in the eyes of the patients. The SGMS exhibited high correlation coefficients (>0.88) with blood glucose changes and a good stability of the sensor response to glucose for the study period of 2 weeks. Lag times were in the range of 5-10 min. A total of 98% of all data pairs was in the clinical acceptable ranges A and B of the consensus error grid.
CONCLUSIONS: For the first time, the possibility to measure glucose in vivo in the subconjunctival interstitial fluid for a period of 2 weeks was demonstrated in a human clinical trial.
© 2012 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22920814      PMCID: PMC3440159          DOI: 10.1177/193229681200600419

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  30 in total

1.  Intraocular lens glucose sensor.

Authors:  W F March; K Ochsner; J Horna
Journal:  Diabetes Technol Ther       Date:  2000       Impact factor: 6.118

2.  Tear glucose dynamics in diabetes mellitus.

Authors:  Jennifer D Lane; David M Krumholz; Robert A Sack; Carol Morris
Journal:  Curr Eye Res       Date:  2006-11       Impact factor: 2.424

3.  Clinical trial of a noninvasive contact lens glucose sensor.

Authors:  Wayne F March; Achim Mueller; Peter Herbrechtsmeier
Journal:  Diabetes Technol Ther       Date:  2004-12       Impact factor: 6.118

Review 4.  Continuous glucose monitoring: roadmap for 21st century diabetes therapy.

Authors:  David C Klonoff
Journal:  Diabetes Care       Date:  2005-05       Impact factor: 19.112

5.  Identification of Hemagglutinin of Jack Bean with Concanavalin A.

Authors:  J B Sumner; S F Howell
Journal:  J Bacteriol       Date:  1936-08       Impact factor: 3.490

6.  A fluorescence-based glucose biosensor using concanavalin A and dextran encapsulated in a poly(ethylene glycol) hydrogel.

Authors:  R J Russell; M V Pishko; C C Gefrides; M J McShane; G L Coté
Journal:  Anal Chem       Date:  1999-08-01       Impact factor: 6.986

7.  A fluorescence affinity hollow fiber sensor for continuous transdermal glucose monitoring.

Authors:  R Ballerstadt; J S Schultz
Journal:  Anal Chem       Date:  2000-09-01       Impact factor: 6.986

8.  In vivo performance evaluation of a transdermal near- infrared fluorescence resonance energy transfer affinity sensor for continuous glucose monitoring.

Authors:  Ralph Ballerstadt; Colton Evans; Ashok Gowda; Roger McNichols
Journal:  Diabetes Technol Ther       Date:  2006-06       Impact factor: 6.118

Review 9.  Fluorescence-based glucose sensors.

Authors:  John C Pickup; Faeiza Hussain; Nicholas D Evans; Olaf J Rolinski; David J S Birch
Journal:  Biosens Bioelectron       Date:  2004-11-21       Impact factor: 10.618

10.  Characterizing dual wavelength polarimetry through the eye for monitoring glucose.

Authors:  Bilal H Malik; Gerard L Coté
Journal:  Biomed Opt Express       Date:  2010-10-28       Impact factor: 3.732

View more
  5 in total

1.  Blood glucose self-monitoring with a long-term subconjunctival glucose sensor.

Authors:  Achim Josef Müller; Monika Knuth; Katharina Sibylle Nikolaus; Roland Krivánek; Frank Küster; Christoph Hasslacher; Gerd Uwe Auffarth
Journal:  J Diabetes Sci Technol       Date:  2013-01-01

2.  First clinical evaluation of a new percutaneous optical fiber glucose sensor for continuous glucose monitoring in diabetes.

Authors:  Achim Josef Müller; Monika Knuth; Katharina Sibylle Nikolaus; Roland Krivánek; Frank Küster; Christoph Hasslacher
Journal:  J Diabetes Sci Technol       Date:  2013-01-01

Review 3.  In-Vivo Microsystems: A Review.

Authors:  Paddy French
Journal:  Sensors (Basel)       Date:  2020-09-01       Impact factor: 3.576

4.  In vitro evaluation of fluorescence glucose biosensor response.

Authors:  Mamdouh Aloraefy; T Joshua Pfefer; Jessica C Ramella-Roman; Kim E Sapsford
Journal:  Sensors (Basel)       Date:  2014-07-08       Impact factor: 3.576

Review 5.  Ocular Complications of Diabetes and Therapeutic Approaches.

Authors:  Victoria J Vieira-Potter; Dimitrios Karamichos; Darren J Lee
Journal:  Biomed Res Int       Date:  2016-03-28       Impact factor: 3.411

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.