A pilot study comparing two polymethylpentene extracorporeal membrane oxygenators.

J Rambaud1, J Guilbert, I Guellec, S Renolleau.   

Abstract

OBJECTIVE: We compared two polymethylpentene oxygenators being used in our unit: the Maquet Quadrox-iD paediatric and the Medos Hilite 800LT. STUDY
DESIGN: A mono-centric, prospective pilot study was conducted on ten consecutive newborn patients who had been admitted to our hospital service for extracorporeal circulation (EC) treatment. We examined the rate of oxygen transfer, the CO2 removal capacity and the average sweep gas flow required to produce this result. We also assessed the disturbances of haemostasis, the need for labile blood products and the membrane oxygenator lifetime and cost of use.
CONCLUSIONS: According to our study, it seems to us that Medos Hilite 800LT membrane oxygenators demonstrate greater oxygen transfer and CO2 removal capacity than Maquet Quadrox-iD paediatric membrane oxygenators, at a similar cost. These results lead us to conclude that it is reasonable to continue using Medos Hilite 800LT membrane oxygenators. A broader comparison study would be necessary in order to support these initial results.

Entities:  

Mesh:

Year:  2012        PMID: 22918934     DOI: 10.1177/0267659112457970

Source DB:  PubMed          Journal:  Perfusion        ISSN: 0267-6591            Impact factor:   1.972


  4 in total

1.  Calcium homeostasis disorder during and after neonatal extracorporeal membrane oxygenation.

Authors:  Jerome Rambaud; Isabelle Guellec; Julia Guilbert; Pierre-Louis Léger; Sylvain Renolleau
Journal:  Indian J Crit Care Med       Date:  2015-09

2.  Venoarterial extracorporeal membrane oxygenation support for neonatal and pediatric refractory septic shock.

Authors:  Jerome Rambaud; Isabelle Guellec; Pierre-Louis Léger; Sylvain Renolleau; Julia Guilbert
Journal:  Indian J Crit Care Med       Date:  2015-10

3.  Microfluidic cell sorting: Towards improved biocompatibility of extracorporeal lung assist devices.

Authors:  Christian Bleilevens; Jonas Lölsberg; Arne Cinar; Maren Knoben; Oliver Grottke; Rolf Rossaint; Matthias Wessling
Journal:  Sci Rep       Date:  2018-05-23       Impact factor: 4.379

4.  Comparison of Circular and Parallel-Plated Membrane Lungs for Extracorporeal Carbon Dioxide Elimination.

Authors:  Leonie S Schwärzel; Anna M Jungmann; Nicole Schmoll; Stefan Caspari; Frederik Seiler; Ralf M Muellenbach; Moritz Bewarder; Quoc Thai Dinh; Robert Bals; Philipp M Lepper; Albert J Omlor
Journal:  Membranes (Basel)       Date:  2021-05-27
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.