Literature DB >> 22913100

Wear of moderately cross-linked polyethylene in fixed-bearing total knee replacements.

Claire L Brockett1, Louise M Jennings, Catherine Hardaker, John Fisher.   

Abstract

Cross-linked polyethylene has been introduced into total joint replacement to improve wear resistance. Although the performance of highly cross-linked polyethylene is well documented clinically and experimentally for total hip replacements, the reduction in mechanical properties with increasing irradiation is of concern for application to total knee replacement. The aim of this study was to investigate the wear performance of a moderately cross-linked polyethylene material in a fixed-bearing total knee replacement. The study was conducted using two femoral geometries, a conventional cruciate-retaining femoral and a high-flexion femoral geometry. The femoral geometry appeared to have no effect on the wear of the knee replacement under standard gait conditions. A significant reduction in wear volume was measured with the moderately cross-linked polyethylene compared with the conventional polyethylene over a six-million-cycle wear study. This study indicates the use of a moderately cross-linked polyethylene in a fixed-bearing total knee replacement may provide a low wearing option for total knee replacement.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22913100     DOI: 10.1177/0954411912445265

Source DB:  PubMed          Journal:  Proc Inst Mech Eng H        ISSN: 0954-4119            Impact factor:   1.617


  9 in total

1.  The choice of the femoral center of rotation affects material loss in total knee replacement wear testing - A parametric finite element study of ISO 14243-3.

Authors:  Steven P Mell; Markus A Wimmer; Hannah J Lundberg
Journal:  J Biomech       Date:  2019-03-23       Impact factor: 2.712

Review 2.  Highly cross-linked polyethylene may not have an advantage in total knee arthroplasty.

Authors:  Vasileios I Sakellariou; Peter Sculco; Lazaros Poultsides; Timothy Wright; Thomas P Sculco
Journal:  HSS J       Date:  2013-08-10

3.  The influence of simulator input conditions on the wear of total knee replacements: An experimental and computational study.

Authors:  Claire L Brockett; Abdellatif Abdelgaied; Tony Haythornthwaite; Catherine Hardaker; John Fisher; Louise M Jennings
Journal:  Proc Inst Mech Eng H       Date:  2016-05       Impact factor: 1.617

4.  PEEK-OPTIMA as an alternative to cobalt chrome in the femoral component of total knee replacement: A preliminary study.

Authors:  Raelene M Cowie; Adam Briscoe; John Fisher; Louise M Jennings
Journal:  Proc Inst Mech Eng H       Date:  2016-09-27       Impact factor: 1.617

5.  A comparison between electromechanical and pneumatic-controlled knee simulators for the investigation of wear of total knee replacements.

Authors:  Abdellatif Abdelgaied; John Fisher; Louise M Jennings
Journal:  Proc Inst Mech Eng H       Date:  2017-03-25       Impact factor: 1.617

6.  Influence of conformity on the wear of total knee replacement: An experimental study.

Authors:  Claire L Brockett; Silvia Carbone; John Fisher; Louise M Jennings
Journal:  Proc Inst Mech Eng H       Date:  2017-12-17       Impact factor: 1.617

7.  Early failure of a modern moderately cross-linked polyethylene acetabular liner.

Authors:  Cynthia A Kahlenberg; Luke Menken; Amar S Ranawat; Jose A Rodriguez
Journal:  Arthroplast Today       Date:  2020-03-07

8.  The effect of insert conformity and material on total knee replacement wear.

Authors:  Abdellatif Abdelgaied; Claire L Brockett; Feng Liu; Louise M Jennings; Zhongmin Jin; John Fisher
Journal:  Proc Inst Mech Eng H       Date:  2013-12-02       Impact factor: 1.617

9.  A comprehensive combined experimental and computational framework for pre-clinical wear simulation of total knee replacements.

Authors:  A Abdelgaied; J Fisher; L M Jennings
Journal:  J Mech Behav Biomed Mater       Date:  2017-11-17
  9 in total