Reference hearing threshold levels for chirp signals delivered by an ER-3A insert earphone.

Kristian Gøtsche-Rasmussen1, Torben Poulsen, Claus Elberling.   

Abstract

OBJECTIVE: To establish reference hearing threshold levels for chirps and frequency-specific chirps.
DESIGN: Hearing thresholds were determined monaurally for broad-band chirps and octave-band chirps using the Etymotic Research, ER-3A insert earphone. The chirps were presented using two repetition rates, 20 and 90 stimuli/s, and with alternating polarity in blocks of one second duration. The test procedure and test conditions were in accordance with the recommendations given in ISO 389-9 (2009) . The ascending method ( ISO 8253-1, 2010 ) was applied using a step size of 5 dB. The chirps were played back from a Tucker Davies Technologies System II, and a Matlab program controlled the test setup. The results are specified in dB peak-to-peak equivalent threshold sound pressure levels (dB peETSPL). STUDY SAMPLE: The test group consisted of 25 otologically-normal young adults (age 18-25 years).
RESULTS: The results are in good agreement with the results from another investigation of hearing thresholds using the same chirp stimuli, and the values for the octave-band chirps are in line with the standardized reference values for corresponding tone bursts ( ISO 389-6, 2007 ).
CONCLUSIONS: The results of the present investigation are relevant for the international standard on short duration signals, ISO 389-6 (2007) .

Mesh:

Year:  2012        PMID: 22909083     DOI: 10.3109/14992027.2012.705901

Source DB:  PubMed          Journal:  Int J Audiol        ISSN: 1499-2027            Impact factor:   2.117


  1 in total

1.  LS CE-Chirp® vs. Click in the neuroaudiological diagnosis by ABR.

Authors:  Michelle Cargnelutti; Pedro Luis Cóser; Eliara Pinto Vieira Biaggio
Journal:  Braz J Otorhinolaryngol       Date:  2016-05-31
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.