Literature DB >> 22907918

Transcatheter closure of multiple interatrial communications.

Johannes Masseli1, Stefan Bertog, Lukas Stanczak, Kira Blankenbach, Nicolas Majunke, Isabel Reiffenstein, Kristina Renkhoff, Katharina Lehn, Nina Wunderlich, Horst Sievert.   

Abstract

OBJECTIVES: We sought to examine acute and midterm results of closure of multiple interatrial communications with staged device deployment and to review the relevant literature.
BACKGROUND: Information about percutaneous methods of closure for multiple defects is limited.
METHODS: We treated 148 patients with multiple defects. Of these, 88 had a relevant left to right shunt ("LRS"), 52 had a presumed paradoxical embolism ("PPE"), five had both (LRS and PPE), and one patient, respectively, had migraine, decompression sickness, and a right to left shunt. After implantation of the first device, closure of additional septal defects was attempted only if indicated clinically.
RESULTS: Ninety-four patients received a single device and 53 more than one. In four patients, surgical defect closure followed. At the end of follow-up (FU; mean 4.5 ± 3.4 years), complete closure of all defects occurred in 67.6% (62.1% for LRS, 76.5% for PPE). Clinical success (small or trivial residual shunt) was achieved in 86.9% (83.9% for LRS, 90.2% for PPE). Complications included pericardial effusions in 2.7%, recurrent thromboembolic events in 4.8%, and new onset of atrial fibrillation in 10.1%. In a significant number of patients with multiple defects, after single device implantation, the likelihood of complete closure increased with FU time (26% complete closure at 1 month vs. 78% at 24 months).
CONCLUSION: Percutaneous closure of multiple interatrial communications is feasible and safe. Importantly, many residual defects close without further intervention at FU. Therefore, staged device delivery is an alternative to simultaneous device implantation, possibly requiring fewer and smaller second devices.
Copyright © 2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2012        PMID: 22907918     DOI: 10.1002/ccd.24329

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  2 in total

1.  Transcatheter closure of multiple secundum atrial septal defects using multiple occluder devices: A comparative experience between pediatric and adult patients.

Authors:  Pramod Sagar; Kothandam Sivakumar; Ramyashri Chandrasekaran; Sreeja Pavithran; Puthiyedath Thejaswi; Rajendran Monica
Journal:  Ann Pediatr Cardiol       Date:  2022-08-19

2.  Comparison of Transcatheter Atrial Septal Defect Closure in Children, Adolescents and adults: Differences, Challenges and Short-, Mid- and Long-Term Results.

Authors:  Turkay Saritas; Ilker Kemal Yucel; Ibrahim Halil Demir; Fadli Demir; Abdullah Erdem; Ahmet Celebi
Journal:  Korean Circ J       Date:  2016-10-20       Impact factor: 3.243

  2 in total