Validation of a novel respiratory rate monitor based on exhaled humidity.

M Niesters1, R Mahajan, E Olofsen, M Boom, S Garcia Del Valle, L Aarts, A Dahan.   

Abstract

BACKGROUND: Postoperative monitoring of ventilation is largely restricted to the measurement of haemoglobin-oxygen saturation and respiratory rate (RR) derived from the ECG. measurement is inadequate when used with supplemental oxygen and ECG-derived RR is subject to artifacts. A new monitor measures RR by quantifying the humidity of exhaled air (respiR8(®)).
METHODS: The accuracy of the system was tested using a breathing simulator. In healthy volunteers, the respiR8(®) monitor was compared with two other methods of measuring RR: capnometry and counting of thoracic breathing movements. The ability of the monitor to track changes in RR resulting from the infusion of 2.5 μg kg(-1) fentanyl was assessed and compared with RR measured from a validated flow measurement system. The RR in 50 postoperative patients measured with the respiR8(®) was compared with that derived from the ECG. RR values were compared by population-based Bland-Altman analyses.
RESULTS: The respiR8(®) monitor was accurate in the range required in clinical practice. There was a close agreement between RR from respiR8(®), capnometry, and manual counting of respiratory movements without bias (limits of agreement ±1 bpm). The respiR8(®) monitor was well able to accurately track RR changes from fentanyl. In postoperative patients, RR from respiR8(®) and ECG had a bias of 1.7 (5.7) bpm due to greater RR values observed from the ECG due to artifacts.
CONCLUSIONS: The respiR8(®) gives an accurate measurement of RR and is useful in postoperative care.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22907341     DOI: 10.1093/bja/aes275

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  5 in total

1.  Breathing is good.

Authors:  Michael Ramsay
Journal:  J Clin Monit Comput       Date:  2015-04       Impact factor: 2.502

2.  Continuous Positive Airway Pressure Mitigates Opioid-induced Worsening of Sleep-disordered Breathing Early after Bariatric Surgery.

Authors:  Sebastian Zaremba; Christina H Shin; Matthew M Hutter; Sanjana A Malviya; Stephanie D Grabitz; Teresa MacDonald; Daniel Diaz-Gil; Satya Krishna Ramachandran; Dean Hess; Atul Malhotra; Matthias Eikermann
Journal:  Anesthesiology       Date:  2016-07       Impact factor: 7.892

3.  A Low-Cost Breath Analyzer Module in Domiciliary Non-Invasive Mechanical Ventilation for Remote COPD Patient Monitoring.

Authors:  Antonio Vincenzo Radogna; Pietro Aleardo Siciliano; Saverio Sabina; Eugenio Sabato; Simonetta Capone
Journal:  Sensors (Basel)       Date:  2020-01-24       Impact factor: 3.576

4.  Ultrafast response humidity sensor using supramolecular nanofibre and its application in monitoring breath humidity and flow.

Authors:  Umesha Mogera; Abhay A Sagade; Subi J George; Giridhar U Kulkarni
Journal:  Sci Rep       Date:  2014-02-17       Impact factor: 4.379

5.  Frequent respiratory events in postoperative patients aged 60 years and above.

Authors:  Suzanne Jl Broens; Xuan He; Rachel Evley; Erik Olofsen; Marieke Niesters; Ravi P Mahajan; Albert Dahan; Monique van Velzen
Journal:  Ther Clin Risk Manag       Date:  2017-08-26       Impact factor: 2.423

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.