S Schütz1, J Egger, S Kühl, A Filippi, J Th Lambrecht.
Abstract
This study concerns intraosseous temperature changes during the use of piezosurgical inserts. On six fresh pig jaws heated to body temperature (36°C), osteotomies and osteoplasties were performed in vitro with the Piezosurgery(®) 3 device (Mectron, Carasco, Italy) and various inserts. The intraosseous temperature increases were measured at a depth of 3mm and at a distance of 1mm from the working site using nickel-chromium/nickel temperature sensors. 20°C Ringer's solution was used for cooling in an initial test series and 10°C Ringer's in a second series. The processed bone was examined using digital volume tomography images to determine the ratio of cortical to cancellous bone thickness. Mean temperature increases of 4.4-10.9°C were found; maximum temperature peaks were over 47°C for an average of only 8.5 s. The type of piezosurgical insert had a marked influence on intraosseous temperature generation (p=0.026); the thickness of the cortical bone and the temperature of the coolant did not. Coolant temperature had an influence on the bone cooling time (p=0.013). The results show that correct use of the piezosurgery device does not give rise to prolonged temperature increases over 47°C and hence does not cause any irreversible thermal damage in the bone.Entities:
Mesh:
Year: 2012 PMID: 22906407 DOI: 10.1016/j.ijom.2012.06.007
Source DB: PubMed Journal: Int J Oral Maxillofac Surg ISSN: 0901-5027 Impact factor: 2.789