Efficacy of tuned waveforms based on different membrane time constants on defibrillation thresholds: primary results from the POWER trial.

Shephal K Doshi1, Michael D Pittaro, Russell Reeves, Ker Boyce, John P Payne, Mark W Kroll, Robert J Graumann, Ashish L Oza, Jesus E Val-Mejias.   

Abstract

BACKGROUND: The efficacy of tuned defibrillation waveforms versus the nominal fixed-tilt waveform has been previously studied. However, the optimal membrane time constant for tuning was not known. The POWER (Pulsewidth Optimized Waveform Evaluation tRial) trial was designed to determine the optimal membrane time constant for programming "tuned" biphasic waveforms.
METHODS: This acute, multicenter study included 121 implantable cardioverter-defibrillator/cardiac resynchronization therapy defibrillator patients who were randomized at implant to any two of the three membrane time constant waveforms (2.5, 3.5, and 4.5 ms). Fixed pulse widths were programmed using the measured high voltage shock impedance. The defibrillation threshold (DFT) estimates were obtained using a hybrid protocol starting with an upper limit of vulnerability estimate followed by a step-up/step-down ventricular fibrillation induction process.
RESULTS: DFT voltage was significantly lower using 3.5- and 4.5-ms waveforms as compared to the 2.5-ms waveform (P = 0.004 and 0.035, respectively). DFT voltage with both 3.5- and 4.5-ms waveforms was ≤ that obtained with the 2.5-ms waveform in 78.5% of the cases. The mean difference in DFT voltage using the 3.5-ms waveform and the 4.5-ms waveform was not significant (P = 0.4). However, the 3.5-ms waveform gave a lower DFT than the 4.5-ms waveform in 19 patients although the reverse was true in only nine (P = 0.02 not significant for multiple comparisons).
CONCLUSIONS: The use of a 3.5- or 4.5-ms time constant-based waveforms had lower DFTs when compared to the 2.5-ms waveform. This study suggests that the first defibrillation attempt at implantation should be with 3.5- or 4.5-ms time constant-based waveforms. The 3.5-ms-based waveform trended toward the best choice. ©2012, The Authors. Journal compilation ©2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2012        PMID: 22906140     DOI: 10.1111/j.1540-8159.2012.03500.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

1.  Does defibrillation threshold increase as left ventricular ejection fraction decreases?

Authors:  Jesus E Val-Mejias; Ashish Oza
Journal:  Europace       Date:  2010-01-03       Impact factor: 5.214