Particle release from haemodialysers.

N A Hoenich1, J Thompson, J McCabe, D R Appleton.   

Abstract

Release of particles from eight makes of hollow-fibre and one make of flatplate dialyser have been studied, and the relationship between rinsing volume and particles recovered established. In a supplementary study the effect on the number of particles released was assessed when striking the dialyser header during priming. Particle size distribution indicated that the majority of the particles recovered were less than 5 microns in diameter. A separate analysis of the particles in the 5-30 micron size range showed two different patterns of particle release. In the dialysers containing ethylene vinyl acetate (EVAL) or Cellulate the number of particles recovered were within the 95% range of the particles in the rinsing fluid and did not alter with increasing rinsing volume. In the dialysers containing Cuprophan and Hemophan the initial particle numbers were higher but fell rapidly up to a 450 ml rinse volume; increasing the rinse volume to 1050 ml did not alter particle recovery. Analysis of variance failed to differentiate between individual dialysers. Striking the header during priming resulted in a transient statistically nonsignificant increase in the number of particles recovered.

Entities:  

Mesh:

Substances:

Year:  1990        PMID: 2289833

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  1 in total

1.  Overexposure to Bisphenol A and Its Chlorinated Derivatives of Patients with End-Stage Renal Disease during Online Hemodiafiltration.

Authors:  Astrid Bacle; Antoine Dupuis; Mohamed Belmouaz; Marc Bauwens; Guillaume Cambien; Nicolas Venisse; Pascale Pierre-Eugene; Sophie Potin; Virginie Migeot; Sarah Ayraud-Thevenot
Journal:  Biomolecules       Date:  2019-08-22
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.