A moving-actuator type electromechanical total artificial heart--Part II: Circular type and animal experiment.

B G Min1, H C Kim, J W Choi, G H Ryu, K P Seo, J R Rho, H Ahn, S W Kim, P D Diegel, D B Olsen.   

Abstract

A new type of electromechanical total artificial heart (TAH) based on circular rolling-cylinder mechanism was developed to overcome critical problems in motor-driven artificial hearts such as large size and difficulties in fitting the heart to atrial remnants and arterial vessels. Its performance and reliability were evaluated in mock circulation and in an animal implant experiment. The total weight and volume of the pump is 650 g and 600 mL, respectively. This new pump was implanted in a calf for total heart replacement and 96 h of survival was achieved. The whole system, including pump, controller, and control algorithm performed well enough to improve the prospect of eventual clinical application of our TAH system.

Entities:  

Mesh:

Year:  1990        PMID: 2289793     DOI: 10.1109/10.64462

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  2 in total

1.  Fluid dynamic study of the moving actuator type total artificial heart by image-processing technique.

Authors:  J Kim; J Kim; J Lee; D Jung; B Min; J Kim
Journal:  Med Biol Eng Comput       Date:  1994-05       Impact factor: 2.602

Review 2.  The ongoing quest for the first total artificial heart as destination therapy.

Authors:  Annemijn Vis; Maziar Arfaee; Husain Khambati; Mark S Slaughter; Jan F Gummert; Johannes T B Overvelde; Jolanda Kluin
Journal:  Nat Rev Cardiol       Date:  2022-06-06       Impact factor: 32.419

  2 in total