A moving-actuator type electromechanical total artificial heart--Part I: Linear type and mock circulation experiments.

B G Min1, H C Kim, S H Lee, J W Kim, J T Kim, I Y Kim, S W Kim, P D Diegel, D B Olsen.   

Abstract

A new type of motor-driven total artificial heart system with a moving-actuator mechanism has been developed. The prototype system consists of a brushless dc motor inside of a rolling-cylinder, two arc-shaped pusher-plates and two polyurethane sacs. The moving-actuator type electromechanical pump has structural advantages of small size and light weight, as compared to other reported motor-driven pumps with fixed-actuator mechanisms. The results of the mock circulation tests are reported in this paper with a cardiac output of 9 L/min at an aortic pressure of 120 mmHg and a heart rate of 120 bpm. The fulfillment of the basic control requirements of the artificial heart was also confirmed, i.e., preload sensitive and afterload insensitive cardiac output response and balanced right and left ventricular outputs.

Entities:  

Mesh:

Year:  1990        PMID: 2289792     DOI: 10.1109/10.64461

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  2 in total

1.  Fluid dynamic study of the moving actuator type total artificial heart by image-processing technique.

Authors:  J Kim; J Kim; J Lee; D Jung; B Min; J Kim
Journal:  Med Biol Eng Comput       Date:  1994-05       Impact factor: 2.602

Review 2.  The ongoing quest for the first total artificial heart as destination therapy.

Authors:  Annemijn Vis; Maziar Arfaee; Husain Khambati; Mark S Slaughter; Jan F Gummert; Johannes T B Overvelde; Jolanda Kluin
Journal:  Nat Rev Cardiol       Date:  2022-06-06       Impact factor: 32.419

  2 in total