Proton therapy dose distribution comparison between Monte Carlo and a treatment planning system for pediatric patients with ependymoma.

Yingcui Jia1, Chris Beltran, Daniel J Indelicato, Stella Flampouri, Zuofeng Li, Thomas E Merchant.   

Abstract

PURPOSE: Compare dose distributions for pediatric patients with ependymoma calculated using a Monte Carlo (MC) system and a clinical treatment planning system (TPS).
METHODS: Plans from ten pediatric patients with ependymoma treated using double scatter proton therapy were exported from the TPS and calculated in our MC system. A field by field comparison of the distal edge (80% and 20%), distal fall off (80% to 20%), field width (50% to 50%), and penumbra (80% to 20%) were examined. In addition, the target dose for the full plan was compared.
RESULTS: For the 32 fields from the 10 patients, the average differences of distal edge at 80% and 20% on central axis between MC and TPS are -1.9 ± 1.7 mm (p < 0.001) and -0.6 ± 2.3 mm (p = 0.13), respectively. Excluding the fields that ranged out in bone or an air cavity, the 80% difference was -0.9 ± 1.7 mm (p = 0.09). The negative value indicates that MC was on average shallower than TPS. The average difference of the 63 field widths of the 10 patients is -0.7 ± 1.0 mm (p < 0.001), negative indicating on average the MC had a smaller field width. On average, the difference in the penumbra was 2.3 ± 2.1 mm (p < 0.001). The average of the mean clinical target volume dose differences is -1.8% (p = 0.001), negative indicating a lower dose for MC.
CONCLUSIONS: Overall, the MC system and TPS gave similar results for field width, the 20% distal edge, and the target coverage. For the 80% distal edge and lateral penumbra, there was slight disagreement; however, the difference was less than 2 mm and occurred primarily in highly heterogeneous areas. These differences highlight that the TPS dose calculation cannot be automatically regarded as correct.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22894399     DOI: 10.1118/1.4736413

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  Is an analytical dose engine sufficient for intensity modulated proton therapy in lung cancer?

Authors:  Suliana Teoh; Francesca Fiorini; Ben George; Katherine A Vallis; Frank Van den Heuvel
Journal:  Br J Radiol       Date:  2019-11-20       Impact factor: 3.629

2.  Technical Note: Defining cyclotron-based clinical scanning proton machines in a FLUKA Monte Carlo system.

Authors:  Francesca Fiorini; Niek Schreuder; Frank Van den Heuvel
Journal:  Med Phys       Date:  2017-12-22       Impact factor: 4.071

3.  Highly efficient and sensitive patient-specific quality assurance for spot-scanned proton therapy.

Authors:  J E Johnson; C Beltran; H Wan Chan Tseung; D W Mundy; J J Kruse; T J Whitaker; M G Herman; K M Furutani
Journal:  PLoS One       Date:  2019-02-14       Impact factor: 3.240

4.  A fast Monte Carlo code for proton transport in radiation therapy based on MCNPX.

Authors:  Keyvan Jabbari; Jan Seuntjens
Journal:  J Med Phys       Date:  2014-07

5.  A simplified methodology to produce Monte Carlo dose distributions in proton therapy.

Authors:  Chris Beltran; Yingcui Jia; Roelf Slopsema; Daniel Yeung; Zuofeng Li
Journal:  J Appl Clin Med Phys       Date:  2014-07-08       Impact factor: 2.102

6.  Influence of internal fixation systems on radiation therapy for spinal tumor.

Authors:  Jingfeng Li; Lei Yan; Jianping Wang; Lin Cai; Dongcai Hu
Journal:  J Appl Clin Med Phys       Date:  2015-07-08       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.