| Literature DB >> 22889600 |
Christophe Devos1, Frank David, Pat Sandra.
Abstract
According to recent directives of the European Union (EU), limits of quantification (LOQ) for the determination of tributyltin (TBT) in surface waters should be ca. 60 pg/L (ppq). This put very stringent requirements on analytical methodologies; definitely when they have to be applied in a routine environment. Stir bar sorptive extraction (SBSE), followed by thermal desorption (TD) and capillary gas chromatography-triple quadrupole mass spectrometry (GC-MS/MS) can provide accurate and precise data at the 2 ng/L level (ppt). For lower concentrations, matrix and reagent interferences together with contamination may provide too high TBT values. A two-dimensional heart-cut GC method was developed to fractionate TBT from interferences. The GC-GC-MS/MS method shows excellent linearity in the range 50 pg/L-4 ng/L, good repeatability (RSD<20% at 200 pg/L), and a limit of detection of 11 pg/L. The method performance is demonstrated with representative samples i.e. harbor water and waste water samples.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22889600 DOI: 10.1016/j.chroma.2012.07.072
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759