Literature DB >> 22883197

[Distinguishing central from obstructive sleep apnea with chest wall surface electrodes].

Hui-er Zhu1, Wei Wang, Yuan-ming Luo.   

Abstract

OBJECTIVE: To determine whether or not diaphragm electromyography recorded from chest wall surface electrodes (EMGsur) can be used to distinguish central from obstructive sleep apnea.
METHODS: Ten patients (age (44 ± 10) years, body mass index (25.9 ± 1.8) kg/m²) with suspected obstructive sleep apnea referred from Guangzhou Institute of Respiratory Disease were studied between January and September 2009. EMGsur and diaphragm electromyography from esophageal electrode (EMGeso) were recorded during conventional overnight full polysomnography. And chest-abdominal movement was measured with chest and abdominal bands.
RESULTS: High-quality EMGsur and EMGeso were recorded in all subjects except for one who could not tolerate a multipair esophageal electrode. Excellent correlation was found between EMGsur and EMGeso during sleep including apnea events (r = 0.81 ± 0.06, P < 0.05). The central sleep apnea events diagnosed by EMGeso were exactly the same as those diagnosed by EMGsur. However, the central sleep apnea events diagnosed by EMGsur were less than those diagnosed by conventional thoracic-abdominal bands (7 ± 11 vs 28 ± 31, P < 0.05).
CONCLUSION: EMGsur may be used to distinguish central from obstructive sleep apnea events.

Entities:  

Mesh:

Year:  2012        PMID: 22883197

Source DB:  PubMed          Journal:  Zhonghua Yi Xue Za Zhi        ISSN: 0376-2491


  1 in total

1.  Neural Respiratory Drive Measured Using Surface Electromyography of Diaphragm as a Physiological Biomarker to Predict Hospitalization of Acute Exacerbation of Chronic Obstructive Pulmonary Disease Patients.

Authors:  Dan-Dan Zhang; Gan Lu; Xuan-Feng Zhu; Ling-Ling Zhang; Jia Gao; Li-Cheng Shi; Jian-Hua Gu; Jian-Nan Liu
Journal:  Chin Med J (Engl)       Date:  2018-12-05       Impact factor: 2.628