Literature DB >> 22882572

Feasibility of early discharge after implantable cardioverter-defibrillator procedures.

Indrajit Choudhuri1, Dipan Desai, Jon Walburg, Phyllis August, Seth I Keller, Ranjit Suri.   

Abstract

INTRODUCTION: Registry data demonstrate considerably low complication rates after implantable cardioverter-defibrillator (ICD) procedures for primary prevention of sudden death. Yet standard of care includes postimplant overnight in-hospital observation that may levy substantial unnecessary financial burden on health care systems. In appropriate patients, discharge soon after implant could translate into significant cost savings, if such practice does not result in complications. We applied a simple clinical algorithm to assess feasibility of discharge on the same day of ICD implantation in patients at low risk for procedural complications.
METHODS: We prospectively randomized primary prevention ICD candidates at low risk for complications (not pacing-dependent or requiring bridging heparin anticoagulation) to next-day discharge with overnight in-hospital observation, or same-day discharge with remote monitoring for 24 hours after ICD implant. Implants were performed via cephalic vein access, and randomization occurred after 4-hours clinical observation and device interrogation. All patients were followed for a minimum of 6 weeks to assess acute procedural complications.
RESULTS: 71 patients comprised the study cohort (mean age 62, 79% male) after 3 were excluded. The most common indication for ICD implant was ischemic cardiomyopathy with ejection fraction ≤35%. Device data obtained through 24-hour remote monitoring was comparable to 4-hour postimplant parameters in same-day discharge patients. No acute complications occurred in same-day discharge patients; 1 next-day discharge patient developed pneumothorax.
CONCLUSION: ICD implantation with same-day discharge is reasonable in patients at low risk for complications. Remote monitoring can be useful in indicating lead-parameter stability during the immediate postoperative period.
© 2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2012        PMID: 22882572     DOI: 10.1111/j.1540-8167.2012.02367.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  5 in total

1.  Organizational model and reactions to alerts in remote monitoring of cardiac implantable electronic devices: A survey from the Home Monitoring Expert Alliance project.

Authors:  Gabriele Zanotto; Antonio D'Onofrio; Paolo Della Bella; Francesco Solimene; Ennio C Pisanò; Saverio Iacopino; Cristina Dondina; Daniele Giacopelli; Alessio Gargaro; Renato P Ricci
Journal:  Clin Cardiol       Date:  2018-12-15       Impact factor: 2.882

Review 2.  Insertion of miniaturized cardiac monitors outside the catheter operating room: experience and practical advice.

Authors:  Jan Steffel; David Jay Wright; Harald Schäfer; Tabinda Rashid-Fadel; Thorsten Lewalter
Journal:  Europace       Date:  2017-10-01       Impact factor: 5.214

3.  Safety of same-day discharge versus overnight stay strategy following cardiac device implantations: a high-volume single-centre experience.

Authors:  Stefanos Archontakis; Evangelos K Oikonomou; Konstantinos Sideris; Ageliki Laina; Dimitra Tirovola; Dimitra Paraskevopoulou; Panagiotis Kostakis; Ioannis Doundoulakis; Petros Arsenos; Ioannis Ntalakouras; Emmanouil Charitakis; Konstantinos Gatzoulis; Konstantinos Tsioufis; Skevos Sideris
Journal:  J Interv Card Electrophysiol       Date:  2022-09-05       Impact factor: 1.759

Review 4.  Interactive telemedicine: effects on professional practice and health care outcomes.

Authors:  Gerd Flodgren; Antoine Rachas; Andrew J Farmer; Marco Inzitari; Sasha Shepperd
Journal:  Cochrane Database Syst Rev       Date:  2015-09-07

5.  Novel Method to Flag Cardiac Implantable Device Infections by Integrating Text Mining With Structured Data in the Veterans Health Administration's Electronic Medical Record.

Authors:  Hillary J Mull; Kelly L Stolzmann; Marlena H Shin; Emily Kalver; Marin L Schweizer; Westyn Branch-Elliman
Journal:  JAMA Netw Open       Date:  2020-09-01
  5 in total