Many biological structures of interest are beyond the diffraction limit of conventional microscopes and their visualization requires application of super-resolution techniques. Such techniques have found remarkable success in surpassing the diffraction limit to achieve sub-diffraction limited resolution; however, they are predominantly limited to fluorescent samples. Here, we introduce a non-fluorescent analogue to structured illumination microscopy, termed structured oblique illumination microscopy (SOIM), where we use simultaneous oblique illuminations of the sample to multiplex high spatial-frequency content into the frequency support of the system. We introduce a theoretical framework describing how to demodulate this multiplexed information to reconstruct an image with a spatial-frequency support exceeding that of the system's classical diffraction limit. This approach allows enhanced-resolution imaging of non-fluorescent samples. Experimental confirmation of the approach is obtained in a reflection test target with moderate numerical aperture.
Many biological structures of interest are beyond the diffraction limit of conventional microscopes and their visualization requires application of super-resolution techniques. Such techniques have found remarkable success in surpassing the diffraction limit to achieve sub-diffraction limited resolution; however, they are predominantly limited to fluorescent samples. Here, we introduce a non-fluorescent analogue to structured illumination microscopy, termed structured oblique illumination microscopy (SOIM), where we use simultaneous oblique illuminations of the sample to multiplex high spatial-frequency content into the frequency support of the system. We introduce a theoretical framework describing how to demodulate this multiplexed information to reconstruct an image with a spatial-frequency support exceeding that of the system's classical diffraction limit. This approach allows enhanced-resolution imaging of non-fluorescent samples. Experimental confirmation of the approach is obtained in a reflection test target with moderate numerical aperture.
Entities:
Keywords:
(030.0030) Coherence and statistical optics; (100.6640) Superresolution; (180.0180) Microscopy
Microscopy is critical in the biological sciences for its ability to visualize biological
samples at the cellular level. There are many subdivisions under this umbrella of general
microscopy, and each are tailored towards specific visualization, design, and contrast
requirements. Examples that have found widespread use include dark-field, phase-contrast,
holographic, and fluorescent microscopies [1]. However, a
critical factor that physically limits the optical resolution of microscopy in general is
diffraction [2]. Unfortunately, many biologically relevant
structures are at sizes beyond the spatial frequency support of this diffraction limit and are
thus physically unobservable using conventional optical techniques. This has prompted many
attempts to surpass this limit to achieve super-resolution.These efforts have found tremendous success in fluorescent imaging, where properties of
fluorophores are creatively utilized to visualize the sample beyond the diffraction limit. One
broad category of such super-resolution techniques relies on single molecule detection, and uses
techniques, such as photoswitching, to excite individual molecules at different times.
Individual emitters can thus be localized with subdiffraction resolution and composite images
incorporating information from the entire acquisition can be reconstructed. Examples of such
techniques include photoactivated localization microscopy (PALM) and stochastic optical
reconstruction microscopy (STORM) [3-6]. Another main category of super-resolution techniques uses patterned
excitation to spatially modulate the excitation of the sample. Two such techniques that have
garnered much attention, stimulated emission depletion (STED) and ground-state depletion (GSD)
microscopy, modulate between the transition states of the fluorophores to optically narrow the
excitation point-spread-function (PSF) [7,8].A particularly robust and efficient solution for far-field super-resolution that also relies
on patterned excitation is structured illumination microscopy (SIM). Conventional SIM typically
illuminates the sample with a sinusoidal pattern. In the linear case, the resulting emission is
a product of the sample structure and the excitation pattern [1-10]. In frequency space, this emission pattern
contains sample frequency components that have been aliased into the system’s passband from
beyond the diffraction limit. Acquiring multiple images with shifted illumination patterns
allows demodulation of the high frequency content from the diffraction-limited content, and
after appropriate shifting in frequency space, an image with subdiffraction resolution can be
reconstructed [11]. Rotating the sinusoidal illumination
pattern allows isotropic filling of frequency space, and the reconstructed image is thus
super-resolved in all orientations. With linear structured illumination, the maximum
super-resolution gain is approximately a factor of two. Nonlinearities in the
excitation/emission of particular fluorophores can be exploited to achieve even greater gains,
and super-resolution by factors of four or five have been reported using nonlinear processes
such as saturation and photoswitching [12-14].The super-resolution techniques introduced above require fluorescence and are thus ill-suited
for samples that are highly scattering but are either not autofluorescent or difficult to
fluorescently tag. A particularly exciting extension to conventional SIM that we explore here is
its application to non-fluorescent, coherently scattering samples. The idea of coherent SIM has
been introduced in [15-19]. Here, however, we provide a more extensive theoretical framework, include
quantitative experimental findings that support the theory, and demonstrate enhanced resolution
of an ex-vivo histological sample. We also draw a parallel between coherent SIM and oblique
illumination microscopy. We show that coherent SIM simultaneously illuminates the sample with
multiple oblique beams, each of which results in a shifted region of the sample’s frequency
spectrum being diffracted into the system aperture and multiplexed, enabling the reconstruction
of an enhanced-resolution image. Given the obliquities of the individual illuminations, we note
that each resulting region of multiplexed frequency content is accounted for by Abbe’s
diffraction theorem. Thus, coherent SIM is not a super-resolution technique in the conventional
sense for it does not gain sample information from beyond Abbe’s diffraction limit; however,
similar to conventional SIM, it allows reconstruction of an image equivalent to what would be
obtained with orthogonal illumination but with an enhanced detection passband.
2. Theory
2.1. Mathematical framework
We first note that a sample under spatially coherent illumination will scatter coherently,
such that scatterings from different sample locations will have fixed phase relationships with
each other. Therefore, for spatially coherent illumination, the image field is a linear
transform of the sample field. This is in contrast with conventional SIM, where the linear
relationship is between the image and object intensities [9,11] because of the spatial incoherence of the
detected fluorescence. Thus, disregarding magnification and operating in the paraxial regime,
the image intensity measured by a detector in a coherent illumination/detection system is given
by the nonlinear relationwhere ⊗ is the convolution operator, r is
the spatial position vector, is the image intensity distribution, is the object reflectance/transmittance function, and
is the illumination field distribution to be imaged onto the
object. We assume that the illumination and detection paths of the system share a limiting
circular aperture, defining the coherent point-spread-function (PSF) . After Fourier transforming, the image spatial frequency
distribution can be written aswhere is the
spatial-frequency vector, is the normalized frequency transfer function corresponding to
the system PSF and Y(),
X(), and
I() are the frequency distributions of
the detected image intensity, object reflectance/transmittance, and illumination field,
respectively.In the case of orthogonal plane wave illumination, and the resulting image frequency distribution is given byFrom coherent diffraction theory, it is known that the transfer function
is a real function given by the system’s pupil function [20]. Thus, from Eq.
(3), it is clear that, under orthogonal illumination, defines the passband of the system’s diffraction limit by sharply
rejecting field frequencies with magnitude beyond some cutoff but allowing all other frequencies to pass through unchanged. We
introduce a framework below to use the same physical aperture and still reconstruct an image
equivalent to one obtained with orthogonal illumination but an enhanced passband.For the sake of simplicity, however, we first consider orthogonal illumination through the
enhanced passband, defined by the extended transfer function,where and are two mutually orthogonal frequency vectors of equal magnitude
such that and . It is clear that this extended transfer has a larger frequency
support than and will thus lead to an image more highly resolved than
. We write this corresponding “enhanced-resolution” image as
and use Eq. (4) to
expand,where ⋆ is the correlation operator. By grouping the terms
together within the square brackets, Eq. (5) can
be directly written as a sum of nine “enhanced-resolution” components,
,We next show that these enhanced-resolution components can be
reconstructed using patterned illumination of the object through the original
aperture, and thus can be reconstructed without the extended
aperture.We first consider a 2D structured illumination pattern through the original aperture, given
byFourier transforming, substituting into Eq. (2), and mathematically expanding, we write the
corresponding intensity image ,We define the terms grouped together within the square
brackets so that Eq. (8) can be directly written
as a linear combination of nine “patterned-excitation” components ,where refer generally to the phase coefficients in Eq. (8). We mathematically show in the Appendix that
these component terms are directly related to and can be exactly
written in terms of the enhanced-resolution component terms introduced in Eq.
(6). Using the relations presented in the Appendix, we rewrite Eq. (8) as a linear combination of the
enhanced-resolution components,The enhanced-resolution components can be linearly solved
for by taking at least 9 acquisitions with independent phase shifts . Statistical or signal processing techniques, similar to ones
introduced for conventional SIM in [21], may be applied
at this point to optimize SNR or reduce the required number of raw acquisitions. After
appropriately calculating and shifting these enhanced-resolution components in frequency-space
and summing, can be reconstructed via Eq. (6).In Fig. 1
below, we show an extended transfer function and a simulated coherent SIM reconstruction
of a 1951 USAF test chart. We note that the amplitude of the extended transfer function shown
in Fig. 1(a) below is color-coded in field, not
intensity. As was explained above, because of the coherent imaging of the system, the image and
object fields are linearly related, and thus the concept of a linear transfer function is only
applicable in the field regime. In this regime, the simulated diffraction limit was set to
and the illumination frequency set to . The effects of enhanced-resolution can be clearly seen when
comparing the modulation of the Group −1 El 4 set of bars in the intensity image simulated in
Figs. 1(e)–1(g).
Fig. 1
Numerical simulation showing the extended transfer function and its enhanced-resolution
reconstruction ability. (a) The extended transfer function given by Eq. (4) and the intensity of the associated
structured illumination field given by Eq.
(7). The transfer function’s axes are given in multiples of . The green dashed circle outlines the frequency support of the
original diffraction limit. (b,c,d) True, orthogonally illuminated, and enhanced-resolution
images, respectively, of a sample USAF test chart. (e,f,g) Magnified view of Group −1 El 4
set of bars at 0.71 lpmm from (b,c,d), respectively. Note the enhanced-resolution
capabilities shown in (d,g).
Numerical simulation showing the extended transfer function and its enhanced-resolution
reconstruction ability. (a) The extended transfer function given by Eq. (4) and the intensity of the associated
structured illumination field given by Eq.
(7). The transfer function’s axes are given in multiples of . The green dashed circle outlines the frequency support of the
original diffraction limit. (b,c,d) True, orthogonally illuminated, and enhanced-resolution
images, respectively, of a sample USAF test chart. (e,f,g) Magnified view of Group −1 El 4
set of bars at 0.71 lpmm from (b,c,d), respectively. Note the enhanced-resolution
capabilities shown in (d,g).We now emphasize a fundamental concept that directly arises from this framework. We note that
the last four enhanced-resolution components in Eq. (6) contain
cross-correlation terms involving orthogonal frequency shifts of . These terms require the orthogonal shifts to be simultaneously
aliased into the system’s detection passband, and imply that cannot be reconstructed by simply illuminating the sample with a
rotating sinusoidal pattern. This is in contrast with conventional SIM, where 2D frequency
space is typically filled by illuminating the sample with a rotating sinusoid. This key
difference arises due to the coherent imaging of the sample, where the relationship between the
object transmittance and detected image intensity is intrinsically nonlinear. Thus, the
illumination pattern itself must contain the same frequency components around which
enhanced-resolution is required. In the example above, we used two orthogonal frequency
components in the illumination to fill frequency space. It naturally follows that, to have a
more isotropic filling of frequency space, more frequency components at different rotation
angles could be used in the illumination pattern. They would, however, all need to be present
simultaneously
2.2. Comparison to oblique illumination microscopy
We now note an interesting observation that allows a more physical understanding of coherent
SIM. Generating the structured illumination pattern given in Eq. (7) requires the interference of four separate, coherent beams at the
sample plane, but not orthogonal to it. Because scattering, and
not fluorescence, is being detected, this setup is in essence oblique
illumination microscopy, albeit with multiple illumination beams. For each individual
illumination beam, a diffraction limited region of the sample’s spatial-frequency spectrum is
transmitted through the optical system; however, each region has a unique position in the
sample’s complete spectrum, depending on the obliquity of its corresponding illumination beam,
and may include high spatial frequencies not present in other regions [15]. Thus, due to the multiple illumination beam superposition at the sample,
each raw image contains multiplexed regions of spatial frequency content that
are individually diffraction-limited and have identical frequency support, but are
shifted over different regions of the sample spectrum. This is in contrast to the
true super-resolution provided by conventional SIM, where the reconstructed high spatial
frequencies are unattainable regardless of the obliquity of any single illumination beam.
However, though the coherent SIM reconstructed image does not contain any “super-resolved” frequencies, it has an
enhanced frequency support with up to twice the extent of the diffraction-limited support of
any single-shot image. Therefore, analogous to conventional SIM, a coherent SIM reconstructed
image is equivalent to an image obtained with orthogonal plane wave illumination with an
enhanced detection passband.
3. System design
To experimentally test our framework for coherent structured illumination enhanced-resolution,
we used a pixel-addressable spatial light modulator (Digital Light Processing (DLP), Texas
Instruments) to gain custom control over the structured pattern imaged onto the sample. A
schematic of our optical system is shown in Fig. 2
. We note here that for the sake of simple theoretical verification, a low numerical
aperture lens was used to image the sample. The resulting modest diffraction limit was used to
avoid optical aberrations and other experimental non-paraxial imperfections. The spatial light
modulator (DLP) was aligned so that the plane of the chip was orthogonal to the optical axis.
The DLP was programmed with a periodic 2D grid pattern and was illuminated with a Gaussian
single mode, spatially coherent, and collimated 405 nm laser beam, resulting in 2 sets of
diffraction orders at orthogonal orientations. The illumination angle onto the DLP chip was
aligned such that the diffraction orders were centered around the optical axis. These
diffraction orders (blue) were directed through a relay of two 4-f systems, and then recombined
to coherently interfere at the sample. A mask was used at the Fourier plane of the first 4-f
system to allow through only the +/− 1 orders from both orthogonal diffraction sets to achieve
the 2D sinusoidal illumination as given by Eq.
(7). The coherent scattering from the sample (green) was imaged by the second 4-f system
and a beam-splitter onto the detector (Pixelink, Ottawa, ON). An adjustable iris was used in the
Fourier plane of the second 4-f system to act as the common limiting aperture for the
illumination and detection arms of the system. We note that due to the diffractive nature of the
DLP, the physical positions of the diffraction orders in the Fourier planes are dependent on
wavelength. Thus, the mask and iris would require realignment for different illumination
wavelengths.
4.1. Structured Oblique Illumination Imaging of Calibration Chart
To experimentally verify the theory, we used a 1951 USAF test target as our coherently
scattering sample. By adjusting the limiting aperture, we effectively tune the numerical
aperture of the last lens before the sample (L4). By adjusting the limiting aperture, we
manually set the diffraction limited resolution to be . The illumination frequencies were pushed to the edge of the
system’s passband to effectively double the system resolution after coherent SIM
reconstruction.We compared the resolution achieved between the orthogonally-illuminated (BF) and
enhanced-resolution (SI) image of the Group 5 Element 4 set of bars of
spacing. Figures 3(a)
and 3(b) show the BF and SI images, respectively.
The dashed lines indicate the regions (averaged over 5 cross-sectional cuts) over which the
horizontal and vertical cross-sectional profiles are shown in Figs. 3(c) and 3(d), respectively. The
resolution bars, which lie beyond the BF diffraction limit, were clearly resolved after
reconstruction. Qualitatively, these experimental results match well with theoretical
predictions.
Fig. 3
Experimental data showing (a) an orthogonally-illuminated (BF) image and (b) an
enhanced-resolution (SI) reconstruction. (c) Horizontal cross-sectional profiles taken from
(a),(b). (d) Vertical cross-sectional profiles taken from (a),(b). (e) Intensity modulations
vs bar freq compared between BF and SI.
Experimental data showing (a) an orthogonally-illuminated (BF) image and (b) an
enhanced-resolution (SI) reconstruction. (c) Horizontal cross-sectional profiles taken from
(a),(b). (d) Vertical cross-sectional profiles taken from (a),(b). (e) Intensity modulations
vs bar freq compared between BF and SI.For a more quantitative analysis, we provide experimental verification for the extended
transfer function given by Eq. (4) and shown in
Fig. 1(a). To do so, we used the same procedure to
resolve other group elements on the test chart and measured their corresponding intensity
modulations, defined asHere, are the maxima and minima, respectively, of the imaged test bars.
In an incoherent imaging system, where the object and image intensity are linearly related,
this intensity modulation measured as a function of the spatial frequency will directly map out
the system’s transfer function [22]. In our coherent
imaging system, where the object and image field are linearly related and
assuming an object with positive and real transmittance, we define an analogous equation to map
out a coherent system’s amplitude transfer function,where we make use of the fact that the field amplitude is
the square-root of the intensity. It can be mathematically shown that .Given that the test chart was oriented to contain only horizontal and vertical frequencies
(frequencies only along the red dashed line in Fig.
1(a)), we can see from Fig. 1(a) that
for passed frequencies. Thus, the detected intensity modulation
is theoretically . In Fig. 3(e) below, we
plot the experimental intensity modulations observed in the BF and SI images with respect to
horizontal bar frequencies. The analogous plot for experimental modulations for vertical bar
frequencies is not shown. In both cases, excellent agreement with the theoretical expectations
was observed.
4.2. Structured Oblique Illumination Microscopy of Polystyrene Beads and Histological
Sample
To demonstrate the utility of this framework on more relevant samples, we image 20 micron
polystyrene beads and an ex-vivo histological sample of a mouse bone joint.
The histological sample was excised and fixed onto a cover slide. No staining, freezing, or
other preparation measures were used. Because the samples themselves had relatively little
inherent scattering, the signal was mainly from the specular scattering off the coverslip from
around the sample. We expect this issue to not be a problem for biological samples known to be
highly scattering. The images from orthogonal-illumination (BF) and coherent SIM imaging (SI)
are compared in Fig. 4
below.
Fig. 4
Top: (a) Experimental orthogonally-illuminated (BF) image and (b) enhanced-resolution (SI)
reconstruction of 20 µm polystyrene beads. (c) and (d) represent enlarged regions of (a) and
(b). (e) shows a comparison of the cross-sectional intensity profiles between the BF and SI
images at the locations marked in yellow in images (c) and (d) Bottom: Same as above, but
for a histological sample of a mouse joint.
Top: (a) Experimental orthogonally-illuminated (BF) image and (b) enhanced-resolution (SI)
reconstruction of 20 µm polystyrene beads. (c) and (d) represent enlarged regions of (a) and
(b). (e) shows a comparison of the cross-sectional intensity profiles between the BF and SI
images at the locations marked in yellow in images (c) and (d) Bottom: Same as above, but
for a histological sample of a mouse joint.Looking at the close-up of the region in the samples with higher structural intricacy, it is
clear from comparing Figs. 4(c) and 4(d) and Fig. 4(c′) and 4(d′) that the enhanced-resolution images (SI) contains
significantly more highly resolved structural information than the orthogonally-illuminated
images (BF). Figures 4(e) and 4(e′) compare the BF and SI cross-sectional intensity profiles from the
locations marked in yellow in Figs. 4(c), 4(d), 4(c′), and 4(d′). As shown, the enhanced-resolution images shows clear
intensity modulations, in contrast with the orthogonally-illuminated images, for structures
spaced 23.5 µm and 19.1 µm apart (Fig. 4(e) and 4(e′), respectively),
which is beyond the diffraction limit set by orthogonal plane wave illumination through the
limiting aperture.
4. Discussion
We have shown experimental results that validate the theoretical framework proposed to obtain
enhanced-resolution through structured oblique illumination microscopy (SOIM) of coherently
scattering samples. We applied this framework to ex-vivo samples and achieved
promising enhanced-resolution imaging of the sample. It is worthwhile now to compare some of the
implications of this framework with that of conventional SIM theory. In our treatment here, the
illumination and detection arms of the system share the same limiting aperture. Thus, the
illumination frequencies are limited by the detection bandwidth. Maximal enhanced-resolution
gain is therefore reached when the illumination frequencies are pushed to the edge of the
detection passband, resulting in an effective doubling of the orthogonally-illuminated system’s
passband. This is analogous to conventional linear fluorescent SIM.A key difference, however, arises when considering the non-linear nature of this coherent
illumination enhanced-resolution reconstruction. This nonlinearity manifests in four of the
enhanced-resolution components in Eq. (6)
containing cross-correlation terms between orthogonal frequency shifts of
. Reconstructing these components with patterned illumination
through the normal aperture requires orthogonal frequency shifts in the illumination to be
simultaneously interfered at the sample and aliased into the system’s detectable bandwidth. This
can only be achieved if the illumination pattern itself has the frequency components around
which enhanced-resolution is desired. Attempting to achieve enhanced-resolution by filling out
2D frequency space with a rotating 1D sinusoidal illumination, as is typical of conventional
SIM, will not allow reconstruction of these cross-correlation enhanced-resolution components,
and will thus yield reconstructed images with artifacts. By extension, to isotropically fill out
frequency space with more than two orthogonally oriented sinusoids as we have presented above,
more orientations can be implemented into the illumination at the cost of a more mathematically
complex reconstruction procedure. The number of orientations that can be simultaneously embedded
into the illumination pattern, however, is ultimately limited by the resolution of the spatial
light modulator.
5. Conclusion
We have described a theoretical framework that extends the concept for obtaining far-field
enhanced resolution via structured illumination to include non-fluorescent, coherently
scattering samples. We applied this framework to enhance resolution of varying spatial frequency
patterns on a grid test target and have demonstrated excellent quantitative agreement between
the observed and theoretical modulations of the spatial patterns. We also used this framework to
image polystyrene beads and an ex-vivo histological sample and were able to
visualize the samples with enhanced resolution. These successful implementations of coherent SIM
show potential to apply this framework to enhance resolution of more biologically relevant
samples in the future. Particularly because proper fluorescent tagging may be difficult or
impossible for many biologically relevant scattering samples, enhanced resolution from
scattering may be especially useful.