Heather M Tassone1, Vivek S Tayal, Anthony J Weekes, Craymon L Garner, James H Norton.
Abstract
BACKGROUND: Ultrasound (US) vascular guidance is traditionally performed in transverse (T) and longitudinal (L) axes, each with drawbacks. We hypothesized that the introduction of a novel oblique (O) approach would improve the success of US-guided peripheral venous access. We examined emergency physician (EP) performance using the O approach in a gel US phantom.Entities:
Year: 2012 PMID: 22871109 PMCID: PMC3439390 DOI: 10.1186/2036-7902-4-14
Source DB: PubMed Journal: Crit Ultrasound J ISSN: 2036-3176
Figure 1The three approaches to vessel aspiration. Out-of-plane Transverse Approach, In-plane Longitudinal Approach, Out-of-plane Oblique Approach.
Figure 2Sonographic view of each approach. Out-of-plane Transverse Approach, In-plane Longitudinal Approach, Out-of-plane Oblique Approach.
Mean, median, minimum, and maximum aspiration times and number of attempts overall
| Time | | | | | |
| Time T (s) | 9.47 | 10.15 | 5.15 | 2.30 | 36.90 |
| Time L (s) | 9.74 | 14.44 | 5.25 | 1.00 | 63.10 |
| Time O (s) | 12.50 | 11.08 | 8.15 | 2.10 | 45.20 |
| Attempt | | | | | |
| Attempt T | 1.75 | 1.51 | 1.00 | 1.00 | 6.00 |
| Attempt L | 1.71 | 2.16 | 1.00 | 1.00 | 10.00 |
| Attempt O | 2.33 | 1.83 | 2.00 | 1.00 | 9.00 |
The p value for times when comparing T, L, and O is 0.572. When comparing attempts, the p value for Cochran’s Q was 0.046. The p values for the Wilcoxon signed ranked tests or the adjustment Bonferroni correction were not significant for the approaches when compared two at a time.
Mean aspiration time and mean number of attempts by experience level
| Time T (s) | 10.43 | 16.98 | 6.12 | 4.33 | 0.131 |
| Time L (s) | 15.45 | 14.67 | 4.47 | 4.37 | 0.371 |
| Time O (s) | 17.50 | 15.45 | 6.22 | 10.83 | 0.007 |
| Attempt T | 2.17 | 2.67 | 1.17 | 1.00 | 0.131 |
| Attempt L | 2.17 | 2.50 | 1.00 | 1.17 | 0.530 |
| Attempt O | 3.67 | 2.50 | 1.33 | 1.83 | 0.132 |