Effect of acceleration forces during transport through a pneumatic tube system on ROTEM® analysis.

Gerhard Amann1, Christine Zehntner, Florian Marti, Giuseppe Colucci.   

Abstract

BACKGROUND: ROTEM® is considered a helpful point-of-care device to monitor blood coagulation in emergency situations. Centrally performed analysis is desirable but rapid transport of blood samples is an important prerequisite. The effect of acceleration forces on sample transport through a pneumatic tube system on ROTEM® should be tested at each institution to exclude a pre-analytical influence. The aims of the present work were: (i) to investigate the effect of pneumatic tube transport on ROTEM® parameters; (ii) to compare blood sample transport via pneumatic tube vs. manual transportation; and (iii) to determine the effect of acceleration forces on ROTEM® parameters.
METHODS: This is a single centre study with 20 healthy volunteers. Five whole blood samples were transferred to the central haematology laboratory by either normal transport or pneumatic delivery with different speed and acceleration forces. EXTEM, INTEM, FIBTEM and APTEM were analysed in parallel with two ROTEM® devices and compared. Acceleration forces were measured during transport with two different instruments.
RESULTS: Increment of transport time, speed and distance resulted in an augmentation of acceleration forces and peaks. All results of the ROTEM® analysis after manual transport or pneumatic delivery were within normal range. However, increase in acceleration forces resulted in minimally but statistically significant changes in multiple ROTEM® parameters. The higher the acceleration forces, the more ROTEM® parameters are influenced.
CONCLUSIONS: Acceleration forces during transport through a pneumatic tube system have an influence on ROTEM® parameters. Prior to transfer blood samples via pneumatic tube system these influences should be tested to exclude clinically relevant blood coagulation activation in vitro.

Entities:  

Mesh:

Year:  2012        PMID: 22868797     DOI: 10.1515/cclm-2011-0800

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  6 in total

1.  Biological samples transportation by drones: ready for prime time?

Authors:  Giuseppe Lippi; Camilla Mattiuzzi
Journal:  Ann Transl Med       Date:  2016-03

2.  The effects of pneumatic tube transport on fresh and stored platelets in additive solution.

Authors:  Per Sandgren; Stella Larsson; Poon Wai-San; Beatrice Aspevall-Diedrich
Journal:  Blood Transfus       Date:  2013-12-03       Impact factor: 3.443

3.  Effect of Pneumatic Tubing System Transport on Platelet Apheresis Units.

Authors:  Jevgenia Zilberman-Rudenko; Frank Z Zhao; Stephanie E Reitsma; Annachiara Mitrugno; Jiaqing Pang; Joseph J Shatzel; Beth Rick; Christina Tyrrell; Wohaib Hasan; Owen J T McCarty; Martin A Schreiber
Journal:  Cardiovasc Eng Technol       Date:  2018-05-21       Impact factor: 2.495

4.  Prediction of Post-Weaning Fibrinogen Status during Cardiopulmonary Bypass: An Observational Study in 110 Patients.

Authors:  Gabor Erdoes; Germaine Gerster; Giuseppe Colucci; Heiko Kaiser; Lorenzo Alberio; Balthasar Eberle
Journal:  PLoS One       Date:  2015-05-26       Impact factor: 3.240

5.  Pneumatic tube system transport does not alter platelet function in optical and whole blood aggregometry, prothrombin time, activated partial thromboplastin time, platelet count and fibrinogen in patients on anti-platelet drug therapy.

Authors:  Dietmar Enko; Harald Mangge; Andreas Münch; Tobias Niedrist; Elisabeth Mahla; Helfried Metzler; Florian Prüller
Journal:  Biochem Med (Zagreb)       Date:  2017-02-15       Impact factor: 2.313

6.  Triple apheresis platelet concentrate quality after pneumatic tube system, conveyor box, and courier transport: An observational study.

Authors:  Lena Reichert; Stefan Wallner; Ralph Burkhardt; Robert Offner; Norbert Ahrens; Viola Hähnel
Journal:  Health Sci Rep       Date:  2022-04-07
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.