Literature DB >> 22867962

Catheter fracture and embolization associated with arm implantation of the cook vital port.

Brent Burbridge1, Grant Stoneham, Peter Szkup, Rob Otani, Stefan Kriegler.   

Abstract

PURPOSE: To determine the frequency of arm port catheter fracture and embolization related to the Cook Vital Port Mini Titanium.
MATERIALS AND METHODS: A retrospective audit of our Cerner Radiology Information System was performed between June 1, 2006, and June 30, 2011, to determine the number of Cook arm venous ports implanted and the frequency of foreign body retrievals related to catheter fracture for these arm ports.
RESULTS: A total of 691 arm implantations of the Cook Vital Port during the 5-year time frame were analysed. Eleven of these patients (1.6%) required intravenous foreign body retrieval in the interventional radiology suite related to catheter fracture and embolization. Three of these fractured catheters were retrieved from the peripheral venous system upstream of the pulmonary circulation, whereas 8 embolized to the pulmonary arteries. All were successfully extracted with an intravenous snare by interventional radiology.
CONCLUSION: We discovered a 1.6% frequency of catheter fracture and embolization associated with arm implantation of the Cook Vital Port. All the catheters fractured at the vein entry site and did not detach from the port housing. The cause for catheter fracture and embolization is uncertain. Pulmonary embolization of the fractured catheters puts the patients at risk for possible further complications. No patients had ancillary complications related to catheter embolization or to catheter extraction procedures. Further investigation is required in an attempt to determine the circumstances that may result in catheter fracture and embolization related to this venous access device.
Copyright © 2013 Canadian Association of Radiologists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Arm ports; Catheter fracture; Embolization; Interventional radiology; Vascular foreign body extraction

Mesh:

Year:  2012        PMID: 22867962     DOI: 10.1016/j.carj.2012.03.001

Source DB:  PubMed          Journal:  Can Assoc Radiol J        ISSN: 0846-5371            Impact factor:   2.248


  4 in total

1.  CT Guided Removal of Iatrogenic Foreign Body: A Broken Intravenous Cannula.

Authors:  Amanpreet Singh; Anupama Kaur; Majhail Singh; Surinder Kaur
Journal:  J Clin Diagn Res       Date:  2015-09-01

2.  Retrieval of embolized tip of port catheter from branch of right pulmonary artery using a macro snare catheter.

Authors:  Mehmet Bostan; Murtaza Emre Durakoğlugil; Omer Satiroğlu; Başar Erdivanli; Gülnihal Tufan
Journal:  Interv Med Appl Sci       Date:  2014-06-04

3.  Short-term and long-term outcome of radiological-guided insertion of central venous access port devices implanted at the forearm: a retrospective monocenter analysis in 1704 patients.

Authors:  Moritz Wildgruber; Sebastian Borgmeyer; Bernhard Haller; Heike Jansen; Jochen Gaa; Marion Kiechle; Reinhard Meier; Johannes Ettl; Hermann Berger
Journal:  Eur Radiol       Date:  2014-09-20       Impact factor: 5.315

4.  Transcatheter Removal of Embolized Port Catheters from the Hearts of Two Children.

Authors:  Osman Baspinar; Ayse Sulu; Derya Aydin Sahin
Journal:  Case Rep Cardiol       Date:  2015-04-29
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.