Dose reduction by moving a region of interest (ROI) beam attenuator to follow a moving object of interest.

Ashish S Panse1, S N Swetadri Vasan, A Jain, D R Bednarek, S Rudin.   

Abstract

Region-of-interest (ROI) fluoroscopy takes advantage of the fact that most neurovascular interventional activity is performed in only a small portion of an x-ray imaging field of view (FOV). The ROI beam filter is an attenuating material that reduces patient dose in the area peripheral to the object of interest. This project explores a method of moving the beam-attenuator aperture with the object of interest such that it always remains in the ROI. In this study, the ROI attenuator, which reduces the dose by 80% in the peripheral region, is mounted on a linear stage placed near the x-ray tube. Fluoroscopy is performed using the Microangiographic Fluoroscope (MAF) which is a high-resolution, CCD-based x-ray detector. A stainless-steel stent is selected as the object of interest, and is moved across the FOV and localized using an object-detection algorithm available in the IMAQ Vision package of LabVIEW. The ROI is moved to follow the stent motion. The pixel intensities are equalized in both FOV regions and an adaptive temporal filter dependent on the motion of the object of interest is implemented inside the ROI. With a temporal filter weight of 5% for the current image in the peripheral region, the SNR measured is 47.8. The weights inside the ROI vary between 10% and 33% with a measured SNR of 57.9 and 35.3 when the object is stationary and moving, respectively. This method allows patient dose reduction as well as maintenance of superior image quality in the ROI while tracking the object.

Entities:  

Year:  2012        PMID: 22866212      PMCID: PMC3409666          DOI: 10.1117/12.911136

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  8 in total

1.  A filtering method for signal equalization in region-of-interest fluoroscopy.

Authors:  Normand Robert; Philip T Komljenovic; J A Rowlands
Journal:  Med Phys       Date:  2002-05       Impact factor: 4.071

2.  Region of interest fluoroscopy.

Authors:  S Rudin; D R Bednarek
Journal:  Med Phys       Date:  1992 Sep-Oct       Impact factor: 4.071

3.  Moving segments region of interest attenuator for x-ray fluoroscopy.

Authors:  S A Sassi; A J Britten
Journal:  Med Phys       Date:  1999-01       Impact factor: 4.071

4.  New light-amplifier-based detector designs for high spatial resolution and high sensitivity CBCT mammography and fluoroscopy.

Authors:  Stephen Rudin; Andrew T Kuhls; Girijesh K Yadava; Gaurav C Josan; Ye Wu; Ravishankar N Chityala; Hussain S Rangwala; N Ciprian Ionita; Kenneth R Hoffmann; Daniel R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2006-12-02

5.  The Micro-Angiographic Fluoroscope (MAF) in High Definition (HD) Mode for Improved Contrast-to-Noise Ratio and Resolution in Fluoroscopy and Roadmapping.

Authors:  Ashish Panse; C N Ionita; W Wang; S K Natarajan; A Jain; D R Bednarek; S Rudin
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2010-10-30

6.  Use of the microangiographic fluoroscope for coiling of intracranial aneurysms.

Authors:  Mandy J Binning; David Orion; Parham Yashar; Sharon Webb; Ciprian N Ionita; Amit Jain; Stephen Rudin; L Nelson Hopkins; Adnan H Siddiqui; Elad I Levy
Journal:  Neurosurgery       Date:  2011-11       Impact factor: 4.654

7.  Spatially different, real-time temporal filtering and dose reduction for dynamic image guidance during neurovascular interventions.

Authors:  S N Swetadri Vasan; P Sharma; C N Ionita; A H Titus; A N Cartwright; D R Bednarek; S Rudin
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2011

8.  The x-ray fovea, a device for reducing x-ray dose in fluoroscopy.

Authors:  M S Labbe; M Y Chiu; M S Rzeszotarski; A R Bani-Hashemi; D L Wilson
Journal:  Med Phys       Date:  1994-03       Impact factor: 4.071

  8 in total
  2 in total

1.  Dynamic electronic collimation method for 3-D catheter tracking on a scanning-beam digital x-ray system.

Authors:  David A P Dunkerley; Jordan M Slagowski; Tobias Funk; Michael A Speidel
Journal:  J Med Imaging (Bellingham)       Date:  2017-04-18

2.  A Patient Dose-Reduction Technique for Neuroendovascular Image-Guided Interventions: Image-Quality Comparison Study.

Authors:  A Sonig; S V Setlur Nagesh; V S Fennell; S Gandhi; L Rangel-Castilla; C N Ionita; K V Snyder; L N Hopkins; D R Bednarek; S Rudin; A H Siddiqui; E I Levy
Journal:  AJNR Am J Neuroradiol       Date:  2018-02-15       Impact factor: 3.825

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.