Effect of angle of incidence on macular thickness and volume measurements obtained by spectral-domain optical coherence tomography.

Amirhossein Hariri1, Sun Young Lee, Humberto Ruiz-Garcia, Muneeswar Gupta Nittala, Florian M Heussen, Srinivas R Sadda.   

Abstract

PURPOSE: Evaluation of the effect of angle of incidence on macular thickness and volume measurements obtained by spectral-domain optical coherence tomography (OCT).
METHODS: A total of 30 eyes from 15 healthy young subjects underwent macular cube volume scans (512 × 128 protocol) following dilation using the Cirrus spectral domain OCT. For each eye, scans were obtained by positioning the scanning beam in the center of the dilated pupil, as well as in four eccentric positions (approximately 3 mm from the center), superior, inferior, nasal, and temporal to the pupillary center, to create oblique angles of incidence between the light beam and retina. In all cases, the region scanned by the volume cube was centered on the fovea. Macular thickness and volume measurements were computed for volume scan acquisitions, and differences in values between eccentric scans and the central scan were analyzed.
RESULTS: Retinal thickness and volume values were observed to increase significantly in all subfields for all eccentrically-obtained scans compared to scans obtained through the center of the pupil. The mean increase in thickness for the various scan positions and subfields ranged from 3.76 to 11.38. Scans that were displaced temporally consistently showed the greatest increase in thickness and volume, whereas nasally positioned scans showed the least increase. The increase in retinal thickness for all subfields correlated significantly with angle of inclination or tilting of the retina.
CONCLUSIONS: Macular thickness and volume measurement results may be affected significantly by positioning of the scanning beam in the pupil and resultant angle of incidence on the retina. These findings suggest that care should be taken to position the scanning beam consistently in the center of the pupil to achieve reliable measurements.

Mesh:

Year:  2012        PMID: 22859741     DOI: 10.1167/iovs.12-9767

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  11 in total

1.  Adjustment of the retinal angle in SD-OCT of glaucomatous eyes provides better intervisit reproducibility of peripapillary RNFL thickness.

Authors:  Kyungmoo Lee; Milan Sonka; Young H Kwon; Mona K Garvin; Michael D Abràmoff
Journal:  Invest Ophthalmol Vis Sci       Date:  2013-07-18       Impact factor: 4.799

2.  Quantitative assessment of outer retinal layers and ellipsoid zone mapping in hydroxychloroquine retinopathy.

Authors:  Obinna Ugwuegbu; Atsuro Uchida; Rishi P Singh; Lucas Beven; Ming Hu; Stephanie Kaiser; Sunil K Srivastava; Justis P Ehlers
Journal:  Br J Ophthalmol       Date:  2018-09-06       Impact factor: 4.638

3.  Tissue thickness calculation in ocular optical coherence tomography.

Authors:  David Alonso-Caneiro; Scott A Read; Stephen J Vincent; Michael J Collins; Maciej Wojtkowski
Journal:  Biomed Opt Express       Date:  2016-01-21       Impact factor: 3.732

4.  Higher-Order Assessment of OCT in Diabetic Macular Edema from the VISTA Study: Ellipsoid Zone Dynamics and the Retinal Fluid Index.

Authors:  Justis P Ehlers; Atsuro Uchida; Ming Hu; Natalia Figueiredo; Peter K Kaiser; Jeffrey S Heier; David M Brown; David S Boyer; Diana V Do; Andrea Gibson; Namrata Saroj; Sunil K Srivastava
Journal:  Ophthalmol Retina       Date:  2019-07-06

5.  Characterizing the Impact of Off-Axis Scan Acquisition on the Reproducibility of Total Retinal Thickness Measurements in SDOCT Volumes.

Authors:  Bhavna J Antony; Paul F Stetson; Michael D Abramoff; Kyungmoo Lee; Johanna M Colijn; Gabriëlle H S Buitendijk; Caroline C W Klaver; Austin Roorda; Brandon J Lujan
Journal:  Transl Vis Sci Technol       Date:  2015-07-31       Impact factor: 3.283

6.  Retinal Fluid Volatility Associated With Interval Tolerance and Visual Outcomes in Diabetic Macular Edema in the VISTA Phase III Trial.

Authors:  Justis P Ehlers; Atsuro Uchida; Duriye Damla Sevgi; Ming Hu; Kim Reed; Alyson Berliner; Robert Vitti; Karen Chu; Sunil K Srivastava
Journal:  Am J Ophthalmol       Date:  2020-11-28       Impact factor: 5.488

7.  Evaluation of macular retinal ganglion cell-inner plexiform layer thickness after vitrectomy with internal limiting membrane peeling for idiopathic macular holes.

Authors:  Alfonso L Sabater; Álvaro Velázquez-Villoria; Miguel A Zapata; Marta S Figueroa; Marta Suárez-Leoz; Luis Arrevola; María-Ángeles Teijeiro; Alfredo García-Layana
Journal:  Biomed Res Int       Date:  2014-07-07       Impact factor: 3.411

8.  Macular Ganglion Cell Imaging Study: Covariate Effects on the Spectral Domain Optical Coherence Tomography for Glaucoma Diagnosis.

Authors:  Jae Hoon Jeong; Yun Jeong Choi; Ki Ho Park; Dong Myung Kim; Jin Wook Jeoung
Journal:  PLoS One       Date:  2016-08-04       Impact factor: 3.240

9.  Variability of Retinal Thickness Measurements in Tilted or Stretched Optical Coherence Tomography Images.

Authors:  Akihito Uji; Nizar Saleh Abdelfattah; David S Boyer; Siva Balasubramanian; Jianqin Lei; SriniVas R Sadda
Journal:  Transl Vis Sci Technol       Date:  2017-03-01       Impact factor: 3.283

Review 10.  Detection of Glaucoma Deterioration in the Macular Region with Optical Coherence Tomography: Challenges and Solutions.

Authors:  Kouros Nouri-Mahdavi; Robert E Weiss
Journal:  Am J Ophthalmol       Date:  2020-09-18       Impact factor: 5.258

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.