Point-of-care monitoring of vitamin K-antagonists: validation of CoaguChek XS test strips with International Standard thromboplastin.

Anton M H P van den Besselaar1, Nathalie C V Péquériaux, Marij Ebben, Joke van der Feest, Kerst de Jong, Maria B J Ganzeboom, Jetty van Ooijen, Ferna Postema, Evelina Witteveen, Felix J M van der Meer.   

Abstract

AIMS: Many patients treated with vitamin K-antagonists (VKA) use point-of-care (POC) whole blood coagulometers for self-testing. The majority of patients in the Netherlands use one type of POC coagulometer, that is, the CoaguChek XS. Each new lot of test strips for the CoaguChek XS is validated by a group of collaborating thrombosis centres. We assessed the International Normalised Ratio (INR) differences between each of 51 new lots of test strips and the International Standard for thromboplastin rTF/95 or its successor rTF/09.
METHODS: Each year, a particular lot of CoaguChek XS test strips was used as reference lot. The reference lot was validated by comparison to the International Standard, yielding a relationship between the reference lot INR and International Standard INR. Successive lots of test strips were compared to the reference lot by three centres using 19-29 capillary blood samples obtained from VKA-treated patients. Each patient provided two blood drops from the same finger prick, one for the reference lot strip and one for the new lot.
RESULTS: The mean INR differences between each lot and the International Standard varied between -8% and +4%. The mean absolute values of the relative differences varied between 2.4% and 8.1%. There were small but clinically unimportant differences in INR between the first and second drop of blood.
CONCLUSIONS: Accuracy of CoaguChek XS INR determinations can be assessed by a group of collaborating centres using a limited number of capillary blood samples. As the mean INR differences with the International Standard were smaller than 10%, the lots were approved for use by the Netherlands Thrombosis Services.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22859395     DOI: 10.1136/jclinpath-2012-200934

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  2 in total

1.  Cost-Utility and Budget Impact Analysis of Implementing Anticoagulation Clinics and Point-of-Care Monitoring Devices in Anticoagulated Patients in Argentina.

Authors:  Osvaldo Ulises Garay; Gonzalo Guiñazú; Yolanda Patricia Adamczuk; Cristina Duboscq
Journal:  Pharmacoecon Open       Date:  2022-07-14

2.  Concept of a point of care test to detect new oral anticoagulants in urine samples.

Authors:  Job Harenberg; Sandra Krämer; Shanshan Du; Christel Weiss; Roland Krämer
Journal:  Thromb J       Date:  2013-08-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.