An ultra low noise telecom wavelength free running single photon detector using negative feedback avalanche diode.

Zhizhong Yan1, Deny R Hamel, Aimee K Heinrichs, Xudong Jiang, Mark A Itzler, Thomas Jennewein.   

Abstract

It is challenging to implement genuine free running single-photon detectors for the 1550 nm wavelength range with simultaneously high detection efficiency (DE), low dark noise, and good time resolution. We report a novel read out system for the signals from a negative feedback avalanche diode (NFAD) [M. A. Itzler, X. Jiang, B. Nyman, and K. Slomkowski, "Quantum sensing and nanophotonic devices VI," Proc. SPIE 7222, 72221K (2009); X. Jiang, M. A. Itzler, K. ODonnell, M. Entwistle, and K. Slomkowski, "Advanced photon counting techniques V," Proc. SPIE 8033, 80330K (2011); M. A. Itzler, X. Jiang, B. M. Onat, and K. Slomkowski, "Quantum sensing and nanophotonic devices VII," Proc. SPIE 7608, 760829 (2010)], which allows useful operation of these devices at a temperature of 193 K and results in very low darkcounts (∼100 counts per second (CPS)), good time jitter (∼30 ps), and good DE (∼10%). We characterized two NFADs with a time-correlation method using photons generated from weak coherent pulses and photon pairs produced by spontaneous parametric down conversion. The inferred detector efficiencies for both types of photon sources agree with each other. The best noise equivalent power of the device is estimated to be 8.1 × 10(-18) W Hz(-1/2), more than 10 times better than typical InP/InGaAs single photon avalanche diodes (SPADs) show in free running mode. The afterpulsing probability was found to be less than 0.1% per ns at the optimized operating point. In addition, we studied the performance of an entanglement-based quantum key distribution (QKD) using these detectors and develop a model for the quantum bit error rate that incorporates the afterpulsing coefficients. We verified experimentally that using these NFADs it is feasible to implement QKD over 400 km of telecom fiber. Our NFAD photon detector system is very simple, and is well suited for single-photon applications where ultra-low noise and free-running operation is required, and some afterpulsing can be tolerated.

Entities:  

Year:  2012        PMID: 22852669     DOI: 10.1063/1.4732813

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  2 in total

1.  Optimal photon pairs for quantum communication protocols.

Authors:  Mikołaj Lasota; Piotr Kolenderski
Journal:  Sci Rep       Date:  2020-11-30       Impact factor: 4.379

2.  Performing private database queries in a real-world environment using a quantum protocol.

Authors:  Philip Chan; Itzel Lucio-Martinez; Xiaofan Mo; Christoph Simon; Wolfgang Tittel
Journal:  Sci Rep       Date:  2014-06-10       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.