| Literature DB >> 22845877 |
Alexei Kazarine1, Matthew C R Kong, Erin J Templeton, Eric D Salin.
Abstract
In this technical note, a liquid-liquid extraction technique was performed using pneumatic liquid recirculation on a centrifugal microfluidic device. Non-contact pneumatic pumping enabled a multi-cycle liquid-liquid extraction process using aqueous iodine in a potassium iodide solution and hexadecane while requiring a minimal amount of space on the device. The extraction process was completely automated on the device following sample introduction and required only 50 s for each extraction cycle. The pumping rate achieved during liquid recirculation was 120 ± 10 μL/min. A recycling process such as the one demonstrated would be difficult to implement in a conventional centrifugal microfluidic system.Entities:
Year: 2012 PMID: 22845877 DOI: 10.1021/ac301421k
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986