Improvement of intradialytic hypotension in diabetic hemodialysis patients using vitamin E-bonded polysulfone membrane dialyzers.

Masahide Koremoto1, Noriko Takahara, Masahiro Takahashi, Yasuhiro Okada, Kouhei Satoh, Tomoyoshi Kimura, Tomoyuki Hirai, Itaru Ebihara, Daisuke Nagasaku, Shigeo Miyata, Shunji Maniwa, Takuji Kouzuma, Tetsurou Arimura, Junzou Kamei.   

Abstract

Currently, there are no detailed reports on the effects of vitamin E-bonded polysulfone (PS) membrane dialyzers on intradialytic hypotension (IDH) in diabetic hemodialysis (HD) patients. This study was designed to evaluate changes in intradialytic systolic blood pressure (SBP) using "VPS-HA" vitamin E-bonded super high-flux PS membrane dialyzers. The subjects were 62 diabetic HD patients whose intradialytic SBP fell by more than 20%. Group A comprised patients who required vasopressors to be able to continue treatment or who had to discontinue therapy due to their lowest intradialytic SBP being observed at 210 min (28 patients). Group B comprised patients who showed no symptoms and required no vasopressors but showed a gradual reduction in blood pressure, with the lowest intradialytic SBP seen at the end of dialysis (34 patients). The primary outcome was defined as the lowest intradialytic SBP after 3 months using VPS-HA. Secondary outcomes included changes in the following: lowest intradialytic diastolic blood pressure, pulse pressure, pulse rate, plasma nitric oxide and peroxynitrite, serum albumin, and hemoglobin A1c. Group A's lowest intradialytic SBP had significantly improved at 3 months (128.0 ±  25.1 mm Hg vs. 117.1 ± 29.2 mm Hg; P = 0.017). Group B's lowest intradialytic SBP had significantly improved at 1 month (134.4 ± 13.2 mm Hg vs. 121.5 ± 25.8 mm Hg; P =  0.047) and 3 months (139.1 ± 20.9 mm Hg vs. 121.5 ±  25.8 mm Hg; P = 0.011). We conclude that VPS-HA may improve IDH in diabetic HD patients.
© 2012, Copyright the Authors. Artificial Organs © 2012, International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22845774      PMCID: PMC3500504          DOI: 10.1111/j.1525-1594.2012.01483.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


Intradialytic hypotension (IDH) is a common complication of hemodialysis (HD). Shoji et al. has reported a correlation between IDH and mortality in HD patients (1). According to a report by the Japanese Society for Dialysis Therapy, the following types of HD patients show the lowering systolic blood pressure (SBP) under HD by more than 20% account for 43.7% overall, 40.9% of patients with chronic glomerulonephritis, and 48.1% of patients with diabetic nephropathy (DN) (2). Furthermore, in the same study, HD patients who exhibit decreased SBP under HD by more than 20% account for 43.7% overall and 46.6% of patients aged 75 to 90 years old (2). The number of renal failure patients with diabetes mellitus as a primary diagnosis is increasing every year in Japan, as are their proportion of HD patients: 14.9% in 1990, 26.0% in 2000, and 35.8% in 2010 (3). It thus appears likely that, in Japan at least, the number of aging patients or patients with DN with symptomatic IDH will continue to grow. Several interventional studies on IDH include comparison of albumin concentrations, comparison of the dialysate sodium concentration (4), comparison of the dialysate calcium concentration (5), comparison of acetic acid concentration (6), and comparisons between HD and hemodiafiltration as treatment methods (7). However, few studies include any comparison of dialyzer flux or advance over conventional dialyzers. Matsumura et al. reported that VPS-H vitamin E-bonded high-flux polysulfone (PS) membrane dialyzer (Asahi Kasei Kuraray Medical, Tokyo, Japan) improved the IDH of eight HD patients at a single center (8). However, there are few multicenter studies on diabetic HD patients with symptomatic IDH that investigate the effects of the dialyzers used. We hypothesized that VPS-HA vitamin E-bonded super high-flux PS membrane dialyzer (Asahi Kasei Kuraray Medical), which has a higher permeability than VPS-H, might improve IDH in DN HD patients. In carrying out our study on DN HD patients who have IDH with comorbid symptoms, we focused on the treatment approach.

PATIENTS AND METHODS

Study population

Eligible to enter the study were adult patients who had been undergoing thrice-weekly HD treatments at nine facilities with an arteriovenous fistula as blood access and anuria with no renal function for a minimum of 1 year. All participants had shown decreased SBP under HD by more than 20% when using conventional dialyzers in the 3 months preceding enrollment. Their primary diagnosis was DN. The study was approved by the Ethics Committee at each facility. Subjects who had cardiovascular disease with Class III and IV symptoms, as defined by the New York Heart Association criteria, were excluded. All patients submitted their written informed consent before the start of the study.

Study design

In this study, we defined IDH as intradialytic SBP falling by more than 20%. We recognized two patterns of IDH. In one, blood pressure decreased during HD with visible symptoms. Vasopressors, increased dosage of vasopressors, or saline were administered, and intradialytic blood pressure was raised or HD treatment was interrupted (defined as Group A). The other pattern was that blood pressure gradually decreased during HD without any symptoms, and post-treatment blood pressure fell by more than 20% (defined as Group B). We thus decided to divide the patients into two groups according to the pattern by which the blood pressure fell. To eliminate the influence on the data by confounding factors, all patients maintained the same HD conditions before starting to use VPS-HA as the interventional dialyzer. We regarded the type and dosage of antihypertensive medication, the type and dosage of erythropoiesis-stimulating agents (ESAs), dry weight, HD dosage (except for HD sessions with symptoms), sodium and calcium concentration of the dialysate, ultrafiltration volumes, temperature of the dialysate, environmental temperature, and intradialytic feeding as confounding factors that might affect intradialytic blood pressure. The HD conditions were as follows. HD sessions were performed thrice-weekly for 4 to 5 h at a blood flow rate of approximately 200 mL/min, conforming to the Japanese HD standard conditions. Dialysate with sodium, calcium, and acetate concentrations of approximately 140, 3, and 10 mEq/L, respectively, was supplied by a centrally supplied dialysate system at a flow rate of approximately 500 mL/min. The ultrafiltration rate (UFR) was managed using the UFR controller in the dialysis equipment. Every patient's intradialytic UFR was kept the same both pre- and post-study. The water for the preparation of the dialysate was purified using a reverse osmotic filter, an ultrafiltrate filter, and an endotoxin retention filter.

Measurements and analyses

The results reported in our preceding article (8) suggested that the effects of vitamin E-bonded PS membrane dialyzer could be successfully investigated over a research period of 3 months. The observation period for this study was 3 months after switching to VPS-HA, including measurements of some parameters at 0 (before switching to VPS-HA) and at 1 and 3 months. Data taken in response to intradialytic BP variations were recorded at every facility. The sphygmomanometer used was not changed during this study. The SBP, diastolic blood pressure (DBP), and pulse rate (PR) were measured every 30 min during HD. The pulse pressure (PP) was calculated from SBP and DBP every 30 min during HD. Mean BP (MBP) was calculated from the summation value of DBP and one-third PP every 30 min during HD. Blood pressure values were recorded for 300 min. Each patient had their intradialytic BP measured for 2 weeks, and the average value was designated the patient's BP. Next, the BP of the patient of each group was calculated and analyzed from the average value. All blood samples were taken from the HD session at the beginning of the week. Plasma NOx was adjusted using the following process. Two milliliters of blood was collected using evacuated tubes, and the plasma was separated in a refrigerated centrifuge (or a refrigerated sample was centrifuged). Plasma aliquots of 0.5 mL were quickly prepared and immediately frozen to −30°C. Some samples were packed into boxes with dry ice and sent frozen to the Department of Pathophysiology and Therapeutics School of Pharmaceutical Sciences at Hoshi University. Plasma nitric oxide (NO2−) and plasma peroxynitrite (NO3−) were measured by mixing plasma samples with Greiss reagent. The color absorbance of the resulting compound was measured using an ENO-20 HPLC Visible Detector (Eicom, Tokyo, Japan) (9). The combined value of NOx was recorded after measuring both NO2− and NO3−. Plasma glycoalbumin (GA) was adjusted using the following process. Three milliliters of blood was collected using evacuated tubes, and the plasma was separated in a refrigerated centrifuge (or a refrigerated sample was centrifuged). Plasma aliquots of 0.8 mL were quickly prepared and then immediately frozen to −30°C. Some samples were packed into boxes with dry ice and sent frozen to the Diagnostics Department of the Research and Development Group at Asahi Kasei Pharma (Shizuoka, Japan). The concentration of plasma GA was measured using a Lucica GA-L kit (Asahi Kasei Pharma, Tokyo, Japan) (10). The data are expressed as percentage GA, determined from the ratio of the measured GA to the total albumin concentration. Serum albumin, blood glucose level, urea nitrogen, hemoglobin, and hemoglobin A1c (HbA1c) were measured at each facility.

Outcome measures

According to the result of a precedent article (8), the primary outcome measure was defined as the lowest intradialytic SBP after 3 months using VPS-HA: the BP at 210 min in Group A and the BP at the end of HD session in Group B. Secondary outcome measures included changes in the following: lowest intradialytic DBP, intradialytic PP, intradialytic MBP, intradialytic PR, plasma NOx, plasma GA, serum albumin, blood glucose, and HbA1c.

Statistical analyses

In our sample size calculation, we assumed that fewer than 10 patients would be withdrawn and 20 patients would thus remain as analytical subjects. We also assumed that at least 10 patients’ lowest SBP would rise more than 10 mm Hg; consequently, the group's lowest SBP would rise significantly over 3 months, the length of the study term. We therefore decided that each group should consist of 30 patients. Statistical analyses were performed using XLSTAT (ver. 2008, Addinsoft SARL, Paris, France). Baseline characteristics were compared using Student's t-test and the chi-squared test. Primary and secondary outcome measures were compared using Student's t-test, the Wilcoxon signed-ranks test, or repeated-measures analysis of variance followed by a post hoc test. A two-tailed P < 0.05 was regarded as statistically significant.

RESULTS

Patient distribution and baseline characteristics

May to August 2007 served as the entry period. Patients were randomly enrolled until there were more than 30 patients in one group. We aimed to enroll 30 patients; however, one group contained 30 patients, the other group contained 35. Sixty-five HD patients were therefore enrolled at nine facilities. One group started with 30 patients and the other group started with 35 patients. The baseline characteristics of study patients are shown in Table 1. Patient characteristics in Table 1 showed no difference between Group A and Group B, except that Group B's average age was higher than that of Group A (P = 0.090) and Group B's serum albumin tended to be higher than that of Group A (P = 0.089). For intradialytic blood pressure, we identified a different pattern of reduction in blood pressure between Group A and Group B (Table 2): intradialytic SBP and PP from 60 to 240 min in Group A were significantly lower than those in Group B, and intradialytic DBP and MBP at 300 min in Group A were significantly higher than those in Group B. Preparations were made to treat the intradialytic symptoms of more Group A patients than Group B patients (data not shown).
TABLE 1

Baseline characteristics of study patients

Group A (n = 30)Group B (n = 35)P value
Gender (male/female)16/1422/130.437
Age (years)60.6 ± 10.065.0 ± 11.50.090
Dialysis vintage (months)46.3 ± 35.360.8 ± 38.70.240
Weight (kg)54.5 ± 10.256.2 ± 10.20.660
SBP (mm Hg)156.5 ± 31.3155.3 ± 26.20.762
DBP (mm Hg)78.9 ± 16.275.8 ± 11.60.164
PP (mm Hg)77.9 ± 19.879.2 ± 16.50.647
MBP (mm Hg)104.8 ± 21.279.2 ± 16.50.504
PR (beats/min)80.6 ± 14.280.6 ± 4.51.000
GA (%)27.1 ± 8.025.0 ± 6.40.277
sAlb (g/dL)3.4 ± 0.33.5 ± 0.40.089
BG (mg/dL)160.0 ± 68.4158.1 ± 43.30.939
UN (mg/dL)63.8 ± 11.4136.7 ± 11.70.233
Hb (g/dL)10.6 ± 0.910.5 ± 0.90.629
HbA1c (%)6.4 ± 1.06.5 ± 1.00.737
NOx (nmol/L)63.6 ± 35.362.9 ± 74.40.960

Values are mean ± SD.

Statistical testing of the baseline characteristics between Group A and Group B were performed using Student's t-test and the chi-squared test.

SBP, systolic blood pressure; DBP, diastolic blood pressure; PP, pulse pressure; MBP, mean blood pressure; PR, pulse rate; GA, glycoalbumin; sAlb, serum albumin; BG, blood glucose; UN, urea nitrogen; Hb, hemoglobin; HbA1c, hemoglobin A1c; NOx, total of plasma nitric oxide (NO2−) and plasma peroxynitrite (NO3−).

TABLE 2

Baseline characteristics of study patients for each blood pressure

Dialysis time (min)Group AGroup BP value
SBP (mm Hg)0156.5 ± 31.3155.3 ± 26.20.762
30142.2 ± 26.9151.9 ± 33.50.138
60131.5 ± 27.5149.0 ± 25.80.000*
90133.7 ± 27.1148.5 ± 24.90.008*
120125.2 ± 24.6149.3 ± 22.60.000*
150123.2 ± 24.2142.3 ± 20.00.000*
180120.3 ± 29.5145.0 ± 22.80.000*
210117.1 ± 29.2143.2 ± 27.60.000*
240119.6 ± 31.9145.6 ± 32.20.000*
270120.4 ± 41.4128.2 ± 53.70.572
300144.2 ± 36.7121.5 ± 51.10.085
DBP (mm Hg)078.9 ± 16.275.8 ± 11.60.164
3074.2 ± 15.474.7 ± 10.50.872
6070.1 ± 14.674.4 ± 14.10.113
9071.6 ± 15.769.9 ± 11.50.573
12067.4 ± 16.275.3 ± 12.50.003*
15066.1 ± 13.772.5 ± 11.30.020*
18066.1 ± 17.173.7 ± 12.10.003*
21062.3 ± 12.971.6 ± 9.90.000*
24065.0 ± 21.373.7 ± 18.40.020*
27060.2 ± 19.466.3 ± 16.30.389
30082.2 ± 16.566.0 ± 19.80.033*
PP (mm Hg)077.9 ± 19.879.2 ± 16.50.647
3067.8 ± 17.675.2 ± 18.70.105
6061.2 ± 17.573.9 ± 17.60.000*
9061.7 ± 17.577.0 ± 21.10.001*
12057.6 ± 13.973.3 ± 16.10.000*
15056.2 ± 14.969.2 ± 14.50.000*
18055.2 ± 18.270.9 ± 18.40.000*
21055.2 ± 22.271.7 ± 21.80.001*
24060.2 ± 20.572.6 ± 24.00.003*
27063.8 ± 32.864.8 ± 22.00.915
30062.0 ± 20.160.1 ± 20.20.790
MBP (mm Hg)0104.8 ± 21.279.2 ± 16.50.504
3096.9 ± 19.275.2 ± 18.70.875
6090.6 ± 18.973.9 ± 17.60.045*
9092.3 ± 19.577.0 ± 21.10.990
12086.7 ± 19.073.3 ± 16.10.001*
15085.2 ± 17.269.2 ± 14.50.239
18084.2 ± 21.270.9 ± 18.40.001*
21080.6 ± 18.471.7 ± 21.80.000*
24083.2 ± 24.872.6 ± 24.00.000*
27080.3 ± 26.764.8 ± 22.00.920
300102.8 ± 23.260.1 ± 20.20.023*
PR (beats/min)080.6 ± 14.280.6 ± 4.51.000
3079.6 ± 16.778.8 ± 2.00.927
6078.0 ± 11.175.8 ± 4.00.718
9079.8 ± 18.577.6 ± 3.70.821
12079.6 ± 12.078.8 ± 3.40.901
15080.2 ± 10.280.2 ± 4.71.000
18092.8 ± 15.582.0 ± 4.30.218
21094.8 ± 14.481.2 ± 5.90.119
24092.2 ± 13.785.2 ± 8.90.417
27091.7 ± 16.888.0 ± 4.30.780

P < 0.05

Values are mean ± SD.

Statistical testing for baseline characteristics between Group A and Group B was performed using Student's t-test.

SBP, systolic blood pressure; DBP, diastolic blood pressure; PP, pulse pressure; MBP, mean blood pressure; PR, pulse rate.

Baseline characteristics of study patients Values are mean ± SD. Statistical testing of the baseline characteristics between Group A and Group B were performed using Student's t-test and the chi-squared test. SBP, systolic blood pressure; DBP, diastolic blood pressure; PP, pulse pressure; MBP, mean blood pressure; PR, pulse rate; GA, glycoalbumin; sAlb, serum albumin; BG, blood glucose; UN, urea nitrogen; Hb, hemoglobin; HbA1c, hemoglobin A1c; NOx, total of plasma nitric oxide (NO2−) and plasma peroxynitrite (NO3−). Baseline characteristics of study patients for each blood pressure P < 0.05 Values are mean ± SD. Statistical testing for baseline characteristics between Group A and Group B was performed using Student's t-test. SBP, systolic blood pressure; DBP, diastolic blood pressure; PP, pulse pressure; MBP, mean blood pressure; PR, pulse rate. After starting the study, three patients left, having withdrawn their agreement to take part. We thus adopted 62 patients as subjects for the final analysis: 28 in Group A and 34 in Group B. All these 62 patients continued with the study (Fig. 1).
FIG. 1

Study flow diagram. BP, blood pressure; SBP, systolic blood pressure.

Study flow diagram. BP, blood pressure; SBP, systolic blood pressure.

Pharmaceutical agents as confounding factors

Prior to the study, all the subjects were being administered one or more of the following antihypertensive medications: calcium channel blockers, angiotensin-converting enzyme inhibitors, alpha-blockers, beta-blockers, angiotensin receptor blockers, and other types. All patients maintained the same dosages and types of antihypertensive medications that they were being prescribed before the study began. As ESAs, either epoetin alpha or epoetin beta was used, and the same dosages and types of ESAs (alpha or beta) were used as before the start of the study.

Conventional dialyzers

All the subjects had been undergoing HD using the following conventional dialyzers: APS-SA (PS, Asahi Kasei Kuraray Medical), APS-S (PS, Asahi Kasei Kuraray Medical), APS-MD (PS, Asahi Kasei Kuraray Medical), TS-U (PS, Toray, Tokyo, Japan), TS-UL (PS, Toray), CS-U (PS, Toray), PS-UW (PS, Kawasumi Laboratories, Inc., Tokyo, Japan), PES-S alpha (polyethersulfone, Nipro, Osaka, Japan), BP-H (polyethersulfone, JMS, Hiroshima, Japan), FDY-GW (polyethersulfone polymer alloy, Nikkiso, Tokyo, Japan), BG-PQ (polymethyl methacrylate, Toray), EK (ethylene vinyl alcohol, Kawasumi Laboratories, Inc.), kf-m (ethylene vinyl alcohol, Kawasumi Laboratories, Inc.), FB-U beta (cellulose triacetate, Nipro), FB-F (cellulose triacetate, Nipro), FB-E (cellulose triacetate, Nipro), AM-BC-P (cellulose, Asahi Kasei Kuraray Medical). Conventional PS membrane dialyzers were being used by 75% of the subjects in Group A and 65% in Group B.

Blood pressure analysis

All the planned data were obtained from the analysis subjects, except for the PR of patients in Group A at 300 min, which was not recorded. As for the intradialytic blood pressure of Group A patients (Table 3), some blood pressures at 3 months had risen significantly: SBP at 210 min: 117.1 ± 29.2 mm Hg at 0 months, 128.0 ± 25.1 mm Hg at 3 months (P = 0.017); DBP at 210 min: 62.3.1 ± 12.9 mm Hg at 0 months, 68.4 ± 15.0 mm Hg at 3 months (P = 0.012); MBP at 210 min: 77.8 ± 21.0 mm Hg at 0 months and 85.6 ± 21.2 mm Hg at 3 months (P = 0.014). Additionally, PR at 90 min at 1 month briefly but significantly rose (79.8 ± 18.5 beats/min at 0 months, 97.0 ± 12.8 beats/min at 1 month, P = 0.039).
TABLE 3

Variation in intradialytic blood pressure of Group A at pre- and post-intervention using VPS-HA

013

Dialysis time (min)P valueP value
SBP (mm Hg)0156.5 ± 31.3159.5 ± 24.60.299157.5 ± 27.10.264
30142.2 ± 26.9134.0 ± 29.20.287142.0 ± 28.50.589
60131.5 ± 27.5132.0 ± 26.20.711135.3 ± 30.10.376
90133.7 ± 27.1134.3 ± 28.00.941136.0 ± 32.20.530
120125.2 ± 24.6126.7 ± 24.60.627126.9 ± 26.60.898
150123.2 ± 24.2122.1 ± 24.90.679126.6 ± 27.20.408
180120.3 ± 29.5119.8 ± 26.10.636126.9 ± 28.30.142
210117.1 ± 29.2115.9 ± 28.90.566128.0 ± 25.10.017*
240119.6 ± 31.9115.4 ± 29.10.278135.6 ± 28.60.023*
270120.4 ± 41.4121.9 ± 27.90.846134.0 ± 33.40.100
300144.2 ± 36.7132.1 ± 38.00.551136.5 ± 33.30.880
DBP (mm Hg)078.9 ± 16.281.4 ± 14.90.14681.5 ± 15.00.139
3074.2 ± 15.470.7 ± 14.50.67775.6 ± 16.40.112
6070.1 ± 14.668.4 ± 14.10.98572.8 ± 15.70.343
9071.6 ± 15.774.1 ± 14.40.16870.5 ± 15.70.971
12067.4 ± 16.267.9 ± 15.10.55067.0 ± 14.40.701
15066.1 ± 13.767.0 ± 13.60.80865.4 ± 15.50.312
18066.1 ± 17.164.7 ± 13.90.59668.3 ± 14.60.432
21062.3 ± 12.968.1 ± 20.80.26268.4 ± 15.00.012*
24065.0 ± 21.363.3 ± 16.00.38472.1 ± 17.40.136
27060.2 ± 19.461.3 ± 15.60.54264.0 ± 15.60.154
30082.2 ± 16.574.0 ± 14.40.52575.8 ± 19.50.537
PP (mm Hg)077.9 ± 19.877.9 ± 18.60.99678.5 ± 19.40.826
3067.8 ± 17.664.8 ± 23.10.64466.3 ± 20.30.698
6061.2 ± 17.564.1 ± 20.00.37862.3 ± 18.60.580
9061.7 ± 17.561.3 ± 20.70.78265.0 ± 20.40.354
12057.6 ± 13.960.3 ± 17.50.17859.3 ± 18.00.734
15056.2 ± 14.956.0 ± 17.80.94357.7 ± 17.70.822
18055.2 ± 18.255.4 ± 20.80.68559.6 ± 18.80.134
21055.2 ± 22.252.0 ± 21.70.47458.8 ± 18.10.360
24060.2 ± 20.555.7 ± 19.50.25162.6 ± 19.80.759
27063.8 ± 32.862.9 ± 25.20.71566.6 ± 20.20.365
30062.0 ± 20.155.7 ± 21.90.74858.6 ± 20.40.896
MBP (mm Hg)0101.1 ± 26.0103.5 ± 23.80.205103.7 ± 25.10.205
3089.9 ± 27.984.6 ± 27.00.22591.3 ± 28.00.241
6085.5 ± 25.384.3 ± 24.40.84089.0 ± 26.30.338
9087.5 ± 25.588.1 ± 26.50.82287.6 ± 26.80.739
12082.3 ± 24.482.6 ± 24.10.81882.8 ± 23.80.850
15080.9 ± 22.680.3 ± 23.70.76980.7 ± 23.90.357
18080.5 ± 24.578.5 ± 23.00.29682.7 ± 25.90.236
21077.8 ± 21.082.6 ± 23.80.81585.6 ± 21.20.014*
24082.3 ± 24.477.4 ± 23.50.014*90.3 ± 23.70.167
27087.8 ± 19.682.1 ± 16.70.51385.9 ± 19.30.391
300104.8 ± 21.093.4 ± 21.80.52995.4 ± 24.30.490
PR (beats/min)080.6 ± 14.289.0 ± 11.90.43887.0 ± 8.90.719
3079.6 ± 16.788.0 ± 11.60.24982.4 ± 11.40.590
6078.0 ± 11.193.2 ± 14.20.19283.4 ± 8.70.396
9079.8 ± 18.597.0 ± 12.80.039*85.2 ± 12.20.303
12079.6 ± 12.097.4 ± 11.20.13187.2 ± 13.00.361
15080.2 ± 10.297.4 ± 10.80.11588.2 ± 11.90.310
18092.8 ± 15.5100.4 ± 16.20.10787.8 ± 9.30.213
21094.8 ± 14.497.5 ± 10.20.95491.2 ± 10.70.505
24092.2 ± 13.788.0 ± 6.70.73888.6 ± 11.60.499
27091.7 ± 16.8107.5 ± 10.50.35695.7 ± 12.80.751

P < 0.05

Values are mean ± SD.

Statistical testing of either 0 month versus 1 month or 0 month versus 3 months was performed using repeated-measures ANOVA followed by a post hoc test.

SBP, systolic blood pressure; DBP, diastolic blood pressure; PP, pulse pressure; MBP, mean blood pressure; PR, pulse rate.

Variation in intradialytic blood pressure of Group A at pre- and post-intervention using VPS-HA P < 0.05 Values are mean ± SD. Statistical testing of either 0 month versus 1 month or 0 month versus 3 months was performed using repeated-measures ANOVA followed by a post hoc test. SBP, systolic blood pressure; DBP, diastolic blood pressure; PP, pulse pressure; MBP, mean blood pressure; PR, pulse rate. As for the intradialytic blood pressure of Group B patients (Table 4), some blood pressures at 3 months had risen significantly: SBP at 300 min: 121.5 ± 25.8 mm Hg at 0 months, 134.4 ± 13.2 mm Hg at 1 month (P = 0.047), and 139.1 ± 20.9 mm Hg at 3 months (P = 0.011); DBP at 300 min: 66.0 ± 19.8 mm Hg at 0 months, 67.7 ± 19.3 mm Hg at 1 month (P = 0.034), and 73.1 ± 19.3 mm Hg at 3 months (P = 0.016); MBP at 300 min: 84.5 ± 20.9 mm Hg at 0 months, 89.9 ± 16.2 mm Hg at 1 month (P = 0.022), and 95.1 ± 18.8 mm Hg at 3 months (P = 0.003).
Table 4

Variation in intradialytic blood pressure of Group B at pre- and post-intervention using VPS-HA

013

Dialysis time (min)P valueP value
SBP (mm Hg)0155.3 ± 20.1157.7 ± 19.80.305161.3 ± 24.30.031*
30151.9 ± 22.9151.7 ± 22.90.978157.0 ± 27.80.256
60149.0 ± 23.5150.4 ± 22.70.612156.6 ± 21.80.013*
90148.5 ± 23.0153.2 ± 26.70.249153.7 ± 23.80.266
120149.3 ± 22.5148.5 ± 22.00.774146.2 ± 22.00.460
150142.3 ± 20.3139.8 ± 24.90.802140.3 ± 24.70.667
180145.0 ± 22.6144.3 ± 24.50.861145.3 ± 26.60.888
210143.2 ± 25.5135.0 ± 28.20.225138.1 ± 26.80.306
240145.6 ± 30.2142.0 ± 27.70.362140.6 ± 26.70.148
270128.2 ± 31.3133.9 ± 22.10.978136.8 ± 18.80.261
300121.5 ± 25.8134.4 ± 13.20.047*139.1 ± 20.90.011*
DBP (mm Hg)075.8 ± 11.676.6 ± 15.90.62278.1 ± 17.00.186
3074.7 ± 10.572.9 ± 9.40.57072.9 ± 13.80.882
6074.4 ± 14.175.6 ± 14.50.74778.7 ± 12.40.026*
9069.9 ± 11.574.4 ± 13.10.07173.4 ± 12.70.108
12075.3 ± 12.574.2 ± 11.80.56372.9 ± 11.90.230
15072.5 ± 11.368.3 ± 13.30.59169.2 ± 14.00.458
18073.7 ± 12.171.9 ± 11.30.18973.8 ± 12.80.662
21071.6 ± 9.969.6 ± 12.90.22672.5 ± 10.80.948
24073.7 ± 18.472.9 ± 13.40.64373.0 ± 12.50.538
27066.3 ± 16.378.1 ± 13.00.05476.8 ± 7.80.044*
30066.0 ± 19.867.7 ± 19.30.034*73.1 ± 19.30.016*
PP (mm Hg)079.2 ± 16.580.8 ± 16.90.43082.2 ± 22.30.191
3075.2 ± 18.778.3 ± 17.60.52382.1 ± 23.60.199
6073.9 ± 17.674.6 ± 17.30.48477.3 ± 17.10.056
9077.0 ± 21.176.7 ± 18.30.99478.0 ± 17.40.814
12073.3 ± 16.175.2 ± 18.30.30172.4 ± 17.90.942
15069.2 ± 14.570.4 ± 17.90.40269.7 ± 16.60.914
18070.9 ± 18.473.1 ± 18.90.41070.7 ± 19.90.885
21071.7 ± 21.867.3 ± 23.80.88465.4 ± 20.80.257
24072.6 ± 24.069.7 ± 21.20.35567.2 ± 20.10.081
27064.8 ± 22.053.1 ± 11.00.10560.6 ± 16.30.375
30060.1 ± 20.260.6 ± 14.00.94263.7 ± 16.10.612
MBP (mm Hg)098.6 ± 20.7100.0 ± 22.50.420101.9 ± 24.40.068
3088.7 ± 31.388.9 ± 28.50.99190.2 ± 31.00.468
6094.6 ± 24.295.8 ± 23.90.55499.6 ± 24.10.008*
9087.6 ± 26.892.3 ± 28.90.07991.9 ± 28.10.114
12095.9 ± 22.795.5 ± 21.20.97193.2 ± 21.60.293
15087.2 ± 27.084.4 ± 26.50.87585.9 ± 26.60.499
18093.4 ± 21.892.4 ± 21.80.41593.4 ± 23.70.687
21093.4 ± 17.785.1 ± 25.70.22389.4 ± 23.60.558
24097.5 ± 19.692.9 ± 21.90.51293.4 ± 20.40.282
27087.0 ± 20.996.7 ± 15.70.13196.8 ± 9.10.071
30084.5 ± 20.989.9 ± 16.20.022*95.1 ± 18.80.003*
PR (beats/min)080.6 ± 4.573.6 ± 21.00.50081.6 ± 3.90.600
3078.8 ± 2.071.1 ± 22.00.40076.4 ± 8.10.497
6075.8 ± 4.069.5 ± 20.00.69377.6 ± 7.30.508
9077.6 ± 3.771.1 ± 21.00.66579.0 ± 9.70.703
12078.8 ± 3.471.9 ± 21.00.61480.6 ± 9.70.609
15080.2 ± 4.774.7 ± 22.00.40483.6 ± 8.80.447
18082.0 ± 4.375.0 ± 22.00.56983.4 ± 10.70.725
21081.2 ± 5.976.2 ± 22.00.21384.0 ± 10.90.452
24085.2 ± 8.978.7 ± 23.00.82584.6 ± 11.60.854
27088.0 ± 4.390.3 ± 8.00.60689.0 ± 9.60.874
30087.7 ± 4.588.7 ± 7.00.58091.7 ± 8.70.314

P < 0.05

Values are mean ± SD.

Statistical testing of either 0 month versus 1 month or 0 month versus 3 months was performed using repeated-measures ANOVA followed by a post hoc test.

SBP, systolic blood pressure; DBP, diastolic blood pressure; PP, pulse pressure; MBP, mean blood pressure; PR, pulse rate.

Variation in intradialytic blood pressure of Group B at pre- and post-intervention using VPS-HA P < 0.05 Values are mean ± SD. Statistical testing of either 0 month versus 1 month or 0 month versus 3 months was performed using repeated-measures ANOVA followed by a post hoc test. SBP, systolic blood pressure; DBP, diastolic blood pressure; PP, pulse pressure; MBP, mean blood pressure; PR, pulse rate. As for symptomatic treatment of Group A patients, we monitored the number of times that treatment was required using saline infusion, glucose infusion, norepinephrine hydrochloride administration, or amezinium methylsulfate administration as the required vasopressor (data not shown). Symptomatic treatments per HD session tended to show a decreasing tendency (not significant). Saline: 0.08 ± 0.16 times at 0 months, 0.04 ± 0.08 times at 1 month, and 0.00 ± 0.00 times at 3 months; glucose: 1.58 ± 1.94 times at 0 months, 0.00 ± 0.00 times at 1 month, and 0.00 ± 0.00 times at 3 months; norepinephrine hydrochloride: 1.39 ± 1.15 times at 0 months, 1.00 ± 1.11 times at 1 month, and 0.89 ± 1.28 times at 3 months; amezinium methylsulfate: 1.20 ± 0.76 times at 0 months, 1.04 ± 0.70 times at 1 month, and 0.76 ± 0.69 times at 3 months. As for Group B patients’ symptomatic treatment, no difference was seen between pre- and post-intervention using VPS-HA.

Blood sample analysis

For some secondary outcome measurements except for blood pressure, no significant differences between Group A and Group B (Table 5A) were seen. However, serum albumin in Group A had significantly risen from 3.4 ± 0.3 g/dL at 0 months to 3.6 ± 0.3 g/dL at 3 months (P = 0.006 in Table 5B); that of Group B had risen significantly from 3.5 ± 0.4 g/dL at 0 months to 3.6 ± 0.3 g/dL at 1 month (P = 0.042); and plasma NOx in Group A had risen significantly from 63.6 ± 35.3 nm at 0 months to 112.4 ± 85.4 nm at 3 months (P = 0.006).
TABLE 5A

Clinical characteristics of each group using VPS-HA (between groups)

VPS-HA treatment period (months)Group AGroup BP value
Glycoalbumin (%)027.1 ± 8.025.0 ± 6.40.277
126.9 ± 6.725.1 ± 5.70.271
326.8 ± 5.625.0 ± 7.00.291
Serum albumin (g/dL)03.4 ± 0.33.5 ± 0.40.089
13.5 ± 0.43.6 ± 0.30.059
33.6 ± 0.33.6 ± 0.40.846
Blood glucose (mg/dL)0160.0 ± 68.4158.1 ± 43.30.939
1152.4 ± 39.5165.6 ± 52.50.456
3131.3 ± 37.3152.0 ± 46.90.214
Urea nitrogen (mg/dL)063.8 ± 11.4136.7 ± 11.70.233
164.7 ± 20.0215.5 ± 14.70.204
366.6 ± 21.9175.1 ± 13.20.263
Hemoglobin (g/dL)010.6 ± 0.910.5 ± 0.90.629
110.4 ± 1.110.6 ± 0.90.743
311.0 ± 1.110.6 ± 1.00.450
HbA1c (%)06.4 ± 1.06.5 ± 1.00.737
16.2 ± 1.36.4 ± 0.80.713
36.2 ± 1.56.5 ± 0.90.543
Plasma NOx (nmol/L)063.6 ± 35.362.9 ± 74.40.960
178.7 ± 61.667.0 ± 60.10.463
3112.4 ± 85.485.5 ± 81.20.220

Values are mean ± SD.

Statistical testing between Group A and Group B was performed using Student's t-test.

TABLE 5B

Statistical differences in the P value of each group using VPS-HA (continuous variation)

Group AGroup B


0 versus 10 versus 30 versus 10 versus 3
Glycoalbumin (%)0.7660.7940.9510.993
Serum albumin (g/dL)0.1460.006*0.042*0.333
Blood glucose (mg/dL)0.7510.2800.5020.635
Urea nitrogen (mg/dL)0.8790.6090.1960.749
Hemoglobin (g/dL)0.6570.3790.4360.466
HbA1c (%)0.5640.5790.3030.788
Plasma NOx (nmol/L)0.2320.006*0.6980.097

P < 0.05

“0 versus 1” means the P value between each data item at 0 months and each data item at 1 month.

“0 versus 3” means the P value between each data item at 0 months and each data item at 3 months.

Values are mean ± SD.

Statistical testing of each group was performed using repeated-measures ANOVA followed by a post hoc test.

Clinical characteristics of each group using VPS-HA (between groups) Values are mean ± SD. Statistical testing between Group A and Group B was performed using Student's t-test. Statistical differences in the P value of each group using VPS-HA (continuous variation) P < 0.05 “0 versus 1” means the P value between each data item at 0 months and each data item at 1 month. “0 versus 3” means the P value between each data item at 0 months and each data item at 3 months. Values are mean ± SD. Statistical testing of each group was performed using repeated-measures ANOVA followed by a post hoc test.

DISCUSSION

The results of our present study, lasting 3 months, which included 62 Japanese HD patients with DN as their primary diagnosis, indicated that using VPS-HA vitamin E-bonded super high-flux PS membrane dialyzer significantly increased the lowest intradialytic SBP and DBP, and also improved IDH. Specifically, the lowest intradialytic SBP in Group A rose without the need for additional dosages of vasopressor or symptomatic treatment. Although the rate of decrease of SBP was significant, the rate of decrease improved after the start of the study. The rate of decrease of DBP also improved. When Group A's patients showed the lowest BP and the condition of the patients at 210 min was serious before using VPS-HA, vasopressor was administered and BP rose as a result at the end of dialysis. A phenomenon in Group A was also observed in which BP rose rapidly at 300 min before using VPS-HA. Similarly, in Group B, when there was no rapid fall in BP, there was no need to use vasopressors, etc., and the rise of the rapid blood pressure at the end of the session was not seen. The significant improvement in BP after switching to the vitamin E-bonded PS membrane dialyzers resulted in a reduction in the use of vasopressors or similar medications during dialysis. Our present study showed that VPS-HA has the potential to improve IDH in diabetes dialysis patients. In our present study, dialyzer flux, which influences the removal performance of free immunoglobulin light chains (11) was almost the same before and after intervention. As for the absorbance potential of protein-bound solutes by the inner surfaces of the membrane, conventional polymethyl methacrylate dialyzers were used in only four patients and polyacrylonitrile in none; we therefore concluded that the adsorbance potential of the membrane would have little influence on our present study data (12,13). In addition, we concluded that the influence of IDH caused by bradykinins would be minimal, due to the use of conventional dialyzers. Our present study showed that VPS-HA significantly improved the lowest intradialytic blood pressure and serum albumin concentration in the same patients. The membrane flux for each group during the prestudy, and that between pre- and post-intervention in both groups, was not significantly different. We thus concluded that the membrane flux of VPS-HA had no influence on the results of the present study. Daugirdas reported that NO influences the vasodilatory effect and is one cause of IDH (14). Nitric oxide is believed to have two different physiological effects on the blood vessels. One is the relaxation and expansion of blood vessels when exposed to high levels of NO, resulting in a rapid fall in BP, and the other is contraction/expansion mediated by the release of NO from vascular endothelial cells. Because we judged that it was not practical to collect blood every 30 min, we decided to evaluate the contraction/expansion of the vascular endothelium. We therefore collected blood before the start of dialysis and measured NOx as reactants of NO. NOx levels were measured in our present study, and proved to be significantly elevated in Group A at 3 months and in Group B at 1 month. As for the observations of significant increases in the lowest intradialytic blood pressure and in NOx, it is possible that both vasodilation and vasoconstriction potency were improved, leading to alleviation of IDH. Miyazaki et al. reported their evaluation of endothelial function, estimated by flow-mediated vasodilation during reactive hyperemia using high-resolution ultrasound Doppler echocardiography, before and after a single session in patients on maintenance HD, using a vitamin E-bonded low-flux cellulose membrane dialyzer and a non-vitamin E-bonded PS membrane dialyzer. Their results indicated that HD using the low-flux cellulose type vitamin E-bonded membrane prevented HD-induced endothelial dysfunction and increases in oxidized LDL (15). In our present study, a transient increase in PR was observed in Group A at 1 month, possibly due to the influence of improved contractile/diastolic capacity of the blood vessels. We did not measure increased amount of food ingested or any inflammation markers. In light of the increasing improvement of albumin concentration as a result of using vitamin E-bonded membrane dialyzers, further research on serum albumin as a factor in colloid osmotic pressure would be worthwhile. Davenport et al. reported that DN HD patients tended to be prescribed higher dosages of antihypertensive medications, and showed worse compliance with blood pressure guidelines, higher weight gain in interdialysis, and more episodes of IDH than non-DN HD patients (16). The serious nature of these problems prompted us to investigate the effect on IDH in DN HD patients by testing a single variable, namely the use of vitamin E-bonded super high-flux PS membrane dialyzers. We performed a subanalysis of dialyzers pre- and post-study, which showed that APS-SA was the most commonly used. Thirteen patients in Group A used APS-SA. In our comparative review of pre- and post-study results in this patient group, no significant difference due to the use of different dialyzers was observed. The number of DN HD patients has been markedly increasing in Japan, and the importance of medical management of such patients has grown accordingly. Our group reported, in a previous study, improved IDH symptoms in eight HD patients in a single facility (8). The results of the present study, showing improved IDH in DN HD patients and a decreased management burden on medical staff, will be useful in HD therapy.

CONCLUSION

The VPS-HA vitamin E-bonded super high-flux PS membrane dialyzer appears to improve the intradialytic hypertension of diabetic nephropathy patients receiving hemodialysis. Randomized controlled trials on other potential effects of vitamin E-bonded membrane dialyzers will now be needed.
  14 in total

1.  Study of glycated amino acid elimination reaction for an improved enzymatic glycated albumin measurement method.

Authors:  Takuji Kouzuma; Yumiko Uemastu; Tomomi Usami; Shigeyuki Imamura
Journal:  Clin Chim Acta       Date:  2004-08-16       Impact factor: 3.786

2.  Hemodialysis-associated hypotension as an independent risk factor for two-year mortality in hemodialysis patients.

Authors:  Tatsuya Shoji; Yoshiharu Tsubakihara; Masamitsu Fujii; Enyu Imai
Journal:  Kidney Int       Date:  2004-09       Impact factor: 10.612

3.  Improved management of intradialytic hypotension (IDH) using vitamin E-bonded polysulfone membrane dialyzer.

Authors:  Masaki Matsumura; Hisako Sasaki; Kumiko Sekizuka; Hiroyuki Sano; Kouji Ogawa; Chihiro Shimizu; Hiroaki Yoshida; Satsuki Kobayashi; Masahide Koremoto; Masaharu Aritomi; Kazue Ueki
Journal:  Int J Artif Organs       Date:  2010-03       Impact factor: 1.595

4.  Dialysis hypotension: a hemodynamic analysis.

Authors:  J T Daugirdas
Journal:  Kidney Int       Date:  1991-02       Impact factor: 10.612

5.  Hemodialysis impairs endothelial function via oxidative stress: effects of vitamin E-coated dialyzer.

Authors:  H Miyazaki; H Matsuoka; H Itabe; M Usui; S Ueda; S Okuda; T Imaizumi
Journal:  Circulation       Date:  2000-03-07       Impact factor: 29.690

6.  Online hemodiafiltration: 4 years of clinical experience.

Authors:  Rosa Muñoz; Isabel Gallardo; Evandro Valladares; Ramón Saracho; Isabel Martínez; Julen Ocharan; Jesús Montenegro
Journal:  Hemodial Int       Date:  2006-01       Impact factor: 1.812

7.  Can modification of renal replacement therapy improve the outcome of patients with systemic inflammatory response syndrome?

Authors:  Andrew Davenport
Journal:  Blood Purif       Date:  2006-03-03       Impact factor: 2.614

8.  Treatment of Severe Intradialytic Hypotension With the Addition of High Dialysate Calcium Concentration to Midodrine and/or Cool Dialysate.

Authors:  R Alappan; D Cruz; A K Abu-Alfa; R Mahnensmith; M A Perazella
Journal:  Am J Kidney Dis       Date:  2001-02       Impact factor: 8.860

9.  Continuous computerized monitoring of hemodynamic parameters during acetate dialysis, bicarbonate dialysis, and acetate-free biofiltration.

Authors:  M Spongano; A Santoro; G Ferrari; F Badiali; M Rossi; A Parrino; C Lamberti; E Sarti; P Zucchelli
Journal:  Artif Organs       Date:  1988-12       Impact factor: 3.094

10.  Impact of sodium and ultrafiltration profiling on haemodialysis-related hypotension.

Authors:  Yi Lun Zhou; Hui Lan Liu; Xiao Feng Duan; Ying Yao; Yi Sun; Qun Liu
Journal:  Nephrol Dial Transplant       Date:  2006-09-05       Impact factor: 5.992

View more
  4 in total

1.  Randomized controlled open-label trial of vitamin E-bonded polysulfone dialyzer and erythropoiesis-stimulating agent response.

Authors:  Tsutomu Sanaka; Takahiro Mochizuki; Eriko Kinugasa; Eiji Kusano; Shigeru Ohwada; Tsutomu Kuno; Kenichiro Kojima; Shuzo Kobayashi; Minoru Satoh; Noriaki Shimada; Kazushi Nakao; Ryoichi Nakazawa; Hideki Nishimura; Eisei Noiri; Takashi Shigematsu; Tadashi Tomo; Teiryo Maeda
Journal:  Clin J Am Soc Nephrol       Date:  2013-04-18       Impact factor: 8.237

Review 2.  Intradialytic hypotension and cardiac remodeling: a vicious cycle.

Authors:  Chia-Ter Chao; Jenq-Wen Huang; Chung-Jen Yen
Journal:  Biomed Res Int       Date:  2015-01-14       Impact factor: 3.411

3.  Clinical evaluation of performance, biocompatibility, and safety of vitamin E-bonded polysulfone membrane hemodialyzer compared to non-vitamin E-bonded hemodialyzer.

Authors:  Mercedeh Kiaii; Masaharu Aritomi; Mitsuyo Nagase; Myriam Farah; Beverly Jung
Journal:  J Artif Organs       Date:  2019-06-26       Impact factor: 1.731

4.  The effect of vitamin E-bonded polysulfone membrane dialyzer on a new oxidative lipid marker.

Authors:  Yuki Kitamura; Kumi Kamimura; Noriko Yoshioka; Yoko Hosotani; Kenji Tsuchida; Masahide Koremoto; Jun Minakuchi
Journal:  J Artif Organs       Date:  2013-02-10       Impact factor: 1.731

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.