Correlation between temperature and dose rate dependence of semiconductor response; influence of accumulated dose.

J Van Dam1, G Leunens, A Dutreix.   

Abstract

The temperature and dose per pulse dependence of 20 Therados semiconductor detectors have been investigated. The rise in relative response observed when increasing temperature from 20 degree C to 32 degree C ranges between 1.00 and 1.07, that observed when increasing dose per pulse from 9.8 x 10(-6) to 6.5 x 10(-4) Gy/pulse between 1.01 and 1.11. The diodes with the most pronounced dose per pulse dependence have also the most pronounced temperature dependence. Accumulated dose has been found to increase both temperature and dose per pulse dependence until a level off dose (less than 4 kGy for one of the diodes) is reached. However, no level off has been observed concerning sensitivity decrease with dose: after an accumulated dose of 29 kGy a sensitivity decrease of 65% was still observed when giving another 25 kGy to one of the diodes.

Mesh:

Year:  1990        PMID: 2284444     DOI: 10.1016/0167-8140(90)90035-u

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  5 in total

Review 1.  In vivo dosimetry: trends and prospects for brachytherapy.

Authors:  G Kertzscher; A Rosenfeld; S Beddar; K Tanderup; J E Cygler
Journal:  Br J Radiol       Date:  2014-07-08       Impact factor: 3.039

2.  In vivo dosimetry with semiconductors in medium dose rate (MDR) brachytherapy for cervical cancer.

Authors:  Mahmoud Allahverdi; Ramin Jaberi; Mehdi Aghili; Fatemeh Ghahremani; Ghazale Geraily
Journal:  Jpn J Radiol       Date:  2012-12-05       Impact factor: 2.374

3.  The consistency of Fletcher-Suit applicator geometry and of the rectal probe's position in high dose rate brachytherapy treatment fraction of cervix carcinoma.

Authors:  Jenő Pálvölgyi
Journal:  J Contemp Brachytherapy       Date:  2009-10-08

4.  In vivo dosimetry using a single diode for megavoltage photon beam radiotherapy: implementation and response characterization.

Authors:  V C Colussi; A S Beddar; T J Kinsella; C H Sibata
Journal:  J Appl Clin Med Phys       Date:  2001       Impact factor: 2.102

5.  Entrance dose measurements for in-vivo diode dosimetry: Comparison of correction factors for two types of commercial silicon diode detectors.

Authors:  X R Zhu
Journal:  J Appl Clin Med Phys       Date:  2000       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.