An 80-year-old patient having a dual chamber DR pacer connected to a VDD lead presented with chronic lead dislodgment with the atrial ring displaced into the right ventricle. There was no ventricular capture at maximal ventricular output, and given the clinical settings, the condition was managed with a conservative approach, the mode was switched to AAI ensuring a minimal adequate ventricular pacing backup in a non dependent patient.Entities:
Keywords: VDD; dislodgment; elderly; pacing
Year: 2012 PMID: 22844200 PMCID: PMC3403576 DOI: 10.4137/CCRep.S9625
Source DB: PubMed Journal: Clin Med Insights Case Rep ISSN: 1179-5476
Figure 1Bedside antero-posterior Chest X rays showing atrial ring in the right ventricle.
Figure 2VVI-like pacing (paced 6th QRS complex) with AAI mode at nominal setting (3.6V/0.4 ms), the first QRS complex in precordial leads was not captured.
Figure 3Interrogation data and programmed parameters.